Siemens Medical Solutions USA, Inc recalled 15 products on 2025-10-08

The FDA (Devices) publishes a recall like this one as 15 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 15 pages.

Products
15
Announced
2025-10-08
Critical
15
Units
1,796

Why these products were recalled

There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–15 of 15

  • CriticalFDA (Devices)··2025-10-08

    MAGNETOM Vida MRI systems recalled over ice blockage in magnet venting

    Siemens is recalling 498 MAGNETOM Vida (DE) MRI systems because an ice blockage in the magnet venting system could stop helium escaping during a quench.

    Product
    MAGNETOM Vida (DE). Model Number: 11060815. 510(k): K170396, K173617, K181433, K183254, K192924, K203443, K213693, K213805, K220425, K231560.
    Category
    Distribution
    0 states
  • CriticalFDA (Devices)··2025-10-08

    MAGNETOM Skyra MRI systems recalled over ice blockage in venting system

    Siemens is recalling 586 MAGNETOM Skyra (DE) MRI systems because an ice blockage in the magnet venting system could prevent helium escaping during a quench.

    Product
    MAGNETOM Skyra (DE). Model Number: 10432915. 510(k) Numbers: K101347, K111242, K123510, K133435, K140253, K141977, K142515, K153343, K173592, K202014, K231560.
    Category
    Distribution
    0 states
  • CriticalFDA (Devices)··2025-10-08

    MAGNETOM Spectra MRI systems recalled over helium venting ice blockage

    Siemens is recalling three MAGNETOM Spectra MRI systems because an ice blockage in the magnet venting system could prevent helium escaping during a quench.

    Product
    MAGNETOM Spectra as the following: 1. MAGNETOM Spectra. Model Number: 10837643. 2. MAGNETOM Spectra (CN). Model Number: 10655588.
    Category
    Distribution
    0 states