Abbott ARCHITECT C System Mixer Recall Due to Blade Separation Risk
Abbott is recalling ARCHITECT C System Mixers because the blade may separate from the mixer due to screw and nut failure, leading to inadequate mixing of samples and reagents.
What this means for you
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall is classified as FDA Class II. The hazard involves a mechanical failure leading to inadequate sample mixing, which is a risk-of-harm product issue where injury has not yet been reported, fitting the 'High' severity criteria.
Plain-English summary
Abbott Laboratories, Inc. is recalling the ARCHITECT C System Mixer. This component is used on all ARCHITECT Clinical Chemistry systems to mix samples with reagents inside cuvettes. The recall involves lot numbers 09D59-01, 09D59-02, and 09D59-03.
The recall was initiated because the mixer blade may separate from the mixer due to screw and nut failure and detachment. If the mixer blade is missing, the sample and reagent are inadequately mixed, which can affect the performance of the automated chemistry tests.
The affected products were distributed worldwide and nationwide in the United States, including specific states and territories, as well as numerous international countries. Users of the ARCHITECT Clinical Chemistry System should check their equipment for the affected lot numbers and contact Abbott Laboratories for further instructions.
The recalled product
- Product
- ARCHITECT C System Mixer - Product Usage: The ARCHITECT Clinical Chemistry System is designed to perform automated chemistry tests, utilizing photometry and potentiometric technology. The mixer is a component and it is used on all ARCHITECT Clinical Chemistry systems to mix the s
- Manufacturer
- Abbott Laboratories, Inc
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (3)
- LN 09D59-01
- LN 09D59-02
- LN 09D59-03
Distribution
Related recalls
Same manufacturer · Abbott Laboratories, Inc
See all →- HighAbbott Alinity s System software error affects diagnostic analyzer wash cycle
FDA (Devices) · 2022-03-23
- HighAutomated diagnostic analyzer from Abbott recalled for waste pressure buildup
FDA (Devices) · 2021-08-18
- ModerateAbbott APS Accelerator Modules Recalled for Missing Pacemaker Safety Label
FDA (Devices) · 2019-10-02
- ModerateAbbott Transferrin Assay Recalled for Instruction Manual Updates
FDA (Devices) · 2019-07-31
- ModerateAbbott Immunoglobulin G Diagnostic Assay Recall for IFU Updates
FDA (Devices) · 2019-07-31
Same hazard · mechanical failure
See all →- HighIMRIS ORT300 Operating Room Table Recalled for Mechanical Failure Risk
FDA (Devices) · 2026-09-02
- HighIMRIS ORT200 Operating Room Table Recalled for Mechanical Failure Risk
FDA (Devices) · 2026-09-02
- HighForest River Transit Buses Recalled for Wheelchair Lift Control Failure
NHTSA · 2026-09-01
- HighBlue Bird Recalls Vision School Buses Due to Wheelchair Lift Defect
NHTSA · 2026-08-31
- SevereMagic Tilt Watercraft Trailers Recalled Due to Winch Weld Failure
NHTSA · 2026-08-28