Intuitive Recalls 17 da Vinci SP Surgical Systems Due to Software Anomaly
Intuitive Surgical is recalling 17 da Vinci SP surgical systems because a software anomaly may cause mechanical vibration of instrument tips and the endoscope under rare conditions.
What this means for you
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall involves a medical device with a potential for mechanical vibration during surgery, which is a risk-of-harm product where injury has not yet been reported, fitting the High severity criteria.
Plain-English summary
Intuitive Surgical, Inc. is recalling 17 units of the da Vinci SP surgical system. The recall is due to a software anomaly that may trigger mechanical vibration of the instrument tips and the endoscope. This issue has the potential to occur only under extremely rare specific conditions: the user is activating Adjust Mode, the instrument Arm is near its vertical position limit, and the user is applying sustained force against the hand control haptic feedback.
The affected units are distributed in the United States in Ohio, Pennsylvania, New York, Minnesota, Tennessee, Texas, Oregon, Illinois, Maryland, New Jersey, and Alabama, as well as in South Korea. The specific serial numbers for the recalled units are SP0022, SP0031, SP0023, SP0032, SP0024, SP0034, SP0025, SP0035, SP0026, SP0036, SP0027, SP0037, SP0028, SP0038, SP0029, SP0039, and SP0030.
The source text does not specify the corrective action or instructions for consumers or healthcare providers. Users should contact Intuitive Surgical for further information regarding the recall and any necessary remediation.
The recalled product
- Product
- da Vinci SP surgical system
- Manufacturer
- Intuitive Surgical, Inc.
- Category
- Medical Device — Surgical Robot
- Hazard
- Affected units
- 17
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (17)
- Serial Numbers: SP0022
- SP0031
- SP0023
- SP0032
- SP0024
- SP0034
- SP0025
- SP0035
- SP0026
- SP0036
- SP0027
- SP0037
- SP0028
- SP0038
- SP0029
- SP0039
- SP0030
Distribution
Related recalls
Same manufacturer · Intuitive Surgical, Inc.
See all →- Moderateda Vinci SP Instrument Arm Drape may detach during surgical procedures
FDA (Devices) · 2026-07-22
- SevereDa Vinci surgical system operative components with E-brake retainer pin defect
FDA (Devices) · 2026-07-01
- HighDa Vinci 5 tower may unintentionally activate the infrared handheld camera
FDA (Devices) · 2026-06-10
- Highda Vinci X and Xi Surgical Systems Recalled for Manipulator Arm Screw Breakage
FDA (Devices) · 2026-05-13
- SevereIntuitive 8 mm SureForm 30 Gray Reloads Incomplete Staple Defect
FDA (Devices) · 2026-05-13
Same hazard · mechanical vibration
See all →- ModerateSiemens Artis zee Biplane Systems Recalled for C-Arm Vibration Issues
FDA (Devices) · 2018-11-21