Siemens Artis zee Biplane Systems Recalled for C-Arm Vibration Issues
Siemens is recalling 14 Artis zee biplane systems because C-arm vibrations may reduce image quality due to a ceiling support screw connection.
What this means for you
Minor labeling or packaging issue, or a voluntary precautionary recall. Risk of harm is low but real — follow the manufacturer's instructions.
Our severity reasoning: The recall is classified as FDA Class II, and the hazard involves reduced image quality rather than direct physical injury or death, fitting the criteria for moderate severity.
Plain-English summary
Siemens Medical Solutions USA, Inc. is recalling 14 units of the Artis zee biplane system, specifically those with Material 10094141. These are dedicated angiography systems used for diagnostic imaging and interventional procedures, including cardiac, neuro, and general angiography.
The recall is issued because vibrations of the C-arm on plane B may result in reduced image quality. This issue is caused by the screw connection to the C-arm in the ceiling support. The affected units were distributed nationwide in the United States.
Healthcare facilities should contact Siemens Medical Solutions USA, Inc. for instructions on how to address the defect. The specific serial numbers for the affected units are 155122, 155149, 155151, 155105, 155148, 154671, 155126, 155131, 155132, 155119, 155121, 155118, and 155147.
The recalled product
- Product
- ARTIS zee biplane systems with Material 10094141- a Angiographic x-ray system Product Usage: Artis zee is a dedicated angiography system developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patien
- Manufacturer
- Siemens Medical Solutions USA, Inc
- Affected units
- 14
Distribution
Distributed nationwide across the United States.
Part of a larger recall action
This product is one of 3 recalled by Siemens Medical Solutions USA, Inc under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 3 products in this recall →Related recalls
Same manufacturer · Siemens Medical Solutions USA, Inc
See all →- HighARTIS icono floor Medical Imaging System Detector Movement Loss Recall
FDA (Devices) · 2026-07-08
- HighARTIS Icono Ceiling Model 11328100 Loss of Movement Capability Recall
FDA (Devices) · 2026-07-08
- ModerateARTIS icono biplane medical imaging system detector lift malfunction
FDA (Devices) · 2026-07-08
- ModerateARTIS icono ceiling medical imaging monitor suspension system
FDA (Devices) · 2026-07-08
- HighARTIS pheno systems may display zero dose while dose is applied
FDA (Devices) · 2026-06-10
Same hazard · reduced image quality
See all →- ModerateSiemens ARTIS Q Biplane Recalled for C-Arm Vibration Issues
FDA (Devices) · 2018-11-21
- ModerateSiemens Artis Q.zen Biplane X-Ray System Recalled for C-Arm Vibration
FDA (Devices) · 2018-11-21