Siemens Artis X-Ray Systems Recalled for Ceiling Mount Screw Failure
Siemens is recalling 120 Artis X-Ray systems because ceiling-mounted support screws may loosen or snap, posing a risk of severe injury to patients and staff.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 4/5 (Severe).
The hazard poses a serious risk of illness, injury, or structural failure. Stop using the product immediately and contact the manufacturer for a refund or repair.
Our severity reasoning: The source text explicitly states that the defect 'can result in severe injuries to patients and operating personnel,' which aligns with the rubric criterion for 'significant injury/property-damage reports' or risk of severe harm.
Plain-English summary
Siemens Medical Solutions USA, Inc. is recalling 120 units of the Siemens Artis zee biplane-Interventional Fluoroscopic X-Ray System (Model Number: 10094141). The recall also covers Artis Q, Artis Pheno, and Artis Icono systems equipped with a DCS Extended or DCS Movable. The affected units were distributed nationwide.
The recall is due to a defect in the screwed connection joining the pivot bearing to the ceiling-mounted support. Individual screws in this connection have the potential to work loose or snap off. If this occurs, it can result in severe injuries to patients and operating personnel.
Healthcare facilities should contact Siemens Medical Solutions USA, Inc. for instructions on how to address the defect. The source text does not specify a specific action for consumers, as these are professional medical devices.
The recalled product
- Product
- Siemens Artis zee biplane-Interventional Fluoroscopic X-Ray System Model Number: 10094141
- Manufacturer
- Siemens Medical Solutions USA, Inc
- Affected units
- 120
Distribution
Distributed nationwide across the United States.
Part of a larger recall action
This product is one of 12 recalled by Siemens Medical Solutions USA, Inc under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 12 products in this recall →Related recalls
Same manufacturer · Siemens Medical Solutions USA, Inc
See all →- ModerateARTIS icono biplane medical imaging system detector lift malfunction
FDA (Devices) · 2026-07-08
- HighARTIS icono floor Medical Imaging System Detector Movement Loss Recall
FDA (Devices) · 2026-07-08
- ModerateARTIS icono ceiling medical imaging monitor suspension system
FDA (Devices) · 2026-07-08
- HighARTIS Icono Ceiling Model 11328100 Loss of Movement Capability Recall
FDA (Devices) · 2026-07-08
- HighARTIS pheno systems may display zero dose while dose is applied
FDA (Devices) · 2026-06-10
Same hazard · structural failure
See all →- High
- SevereMedline Bariatric Rollator Recalled Due to Front Wheel Detachment Risk
FDA (Devices) · 2026-10-14
- SevereMedline Bariatric Rollator Recalled Due to Front Wheel Assembly Disconnect Risk
FDA (Devices) · 2026-10-14
- ModerateQuantum Ops Recalls ADAPTABLE Beach Chair Surgical Patient Positioners
FDA (Devices) · 2026-10-14
- SeverePacearth Recalls Swing Seats Due to Serious Injury Risk from Falling
CPSC · 2026-10-08
Similar recalls
- ModerateSiemens Artis zee X-Ray System Recalled for Cabling Abrasion
FDA (Devices) · 2021-05-05
- SevereSiemens Artis Biplane Systems Recalled for Ceiling Mount Defect
FDA (Devices) · 2018-02-14
- SevereSiemens Artis X-Ray Systems Recalled for Patient Table Cracks
FDA (Devices) · 2019-09-04
- HighSiemens Artis X-Ray Systems Recalled for Electrical Contact Issues
FDA (Devices) · 2020-06-17
- ModerateSiemens Artis zee Biplane Systems Recalled for C-Arm Vibration Issues
FDA (Devices) · 2018-11-21
- HighSiemens Artis zee Ceiling Recalled for Display Malfunction
FDA (Devices) · 2018-07-04