Siemens Recalls Linear Accelerator Tabletops Due to Screw Breakage Risk
Siemens is recalling specific linear accelerator tabletops because manufacturing errors may cause screws to break off, potentially affecting patient safety.
What this means for you
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall involves a risk-of-harm product (medical device) where a mechanical failure could lead to injury, but no specific injuries or illnesses have been reported in the source text.
Plain-English summary
Siemens Medical Solutions USA, Inc. is recalling specific PRIMUS, ONCOR, and ARTISTE linear accelerator systems equipped with 550TxT treatment tabletops. The recall involves 179 active devices distributed worldwide, including in the United States and numerous international locations.
The recall is due to a manufacturing error in the rail threaded holes of the 550TxT treatment tabletop. This error may cause the screws holding the side rails in position to break off. The affected components are identified by serial number ranges 1001-1233 for the TT-A model and 1001-1083 for the TT-S model.
These devices are intended to deliver X-ray photon and electron radiation for the therapeutic treatment of cancer. Siemens has identified the issue through a customer complaint. Users of these devices should contact Siemens Medical Solutions USA, Inc. for further instructions on how to address the potential defect.
The recalled product
- Product
- PRIMUS, ONCOR, ARTISTE Linac systems with component Tabletop, 550TxT, TT-A s/n 1001-1233 and Tabletop 550TxT, TT-S s/n 1001-1083. Product Usage: The intended use of the SIEMENS branded, ARTISTE, ONCOR and PRIMUS family of linear accelerator systems is to deliver X-ray ph
- Manufacturer
- Siemens Medical Solutions USA, Inc
- Affected units
- 179
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (5)
- ONCOR
- PRIMUS and ARTISTE Lican systems with tabletop component: Tabletop
- 550TxT
- TT-A 08632296* Tabletop
- TT-S 10046140
Distribution
Related recalls
Same manufacturer · Siemens Medical Solutions USA, Inc
See all →- HighARTIS icono floor Medical Imaging System Detector Movement Loss Recall
FDA (Devices) · 2026-07-08
- HighARTIS Icono Ceiling Model 11328100 Loss of Movement Capability Recall
FDA (Devices) · 2026-07-08
- ModerateARTIS icono biplane medical imaging system detector lift malfunction
FDA (Devices) · 2026-07-08
- ModerateARTIS icono ceiling medical imaging monitor suspension system
FDA (Devices) · 2026-07-08
- HighARTIS pheno systems may display zero dose while dose is applied
FDA (Devices) · 2026-06-10
Same hazard · mechanical failure
See all →- HighIMRIS ORT300 Operating Room Table Recalled for Mechanical Failure Risk
FDA (Devices) · 2026-09-02
- HighIMRIS ORT200 Operating Room Table Recalled for Mechanical Failure Risk
FDA (Devices) · 2026-09-02
- HighForest River Transit Buses Recalled for Wheelchair Lift Control Failure
NHTSA · 2026-09-01
- HighBlue Bird Recalls Vision School Buses Due to Wheelchair Lift Defect
NHTSA · 2026-08-31
- SevereMagic Tilt Watercraft Trailers Recalled Due to Winch Weld Failure
NHTSA · 2026-08-28