The Recall Desk
Agency classClass IIOur severity:HighFDA (Devices)·Z-2077-2015·Announced 2015-07-22

Sorin Stockert Heater-Cooler 3T Recalled for Potential Organism Colonization

Sorin Group USA is recalling 21 Sorin Stockert Heater-Cooler 3T units due to the potential for organism colonization, including Mycobacteria, if proper disinfection and maintenance are not performed.

The agency’s classification

FDA (Devices): Class II

Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.

This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work

What this means for you

Our assessment, not the agency’s: severity 3/5 (High).

Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.

Our severity reasoning: The recall involves a high-risk pathogen (Mycobacteria) and a medical device where injury has not yet been reported, fitting the criteria for a High severity score.

Plain-English summary

Sorin Group USA, Inc. is recalling 21 units of the Sorin Stockert Heater-Cooler 3T, 240 V / 60 Hz. This device is used for temperature control during extracorporeal perfusion of durations up to 6 hours. The affected units have serial numbers ranging from 16S10743 to 16S11708 and product code 16-02-81.

The recall is issued because of the potential colonization of organisms, including Mycobacteria, in the devices. This risk exists if proper disinfection and maintenance are not performed according to the Instructions for Use. The agency classifies this recall as Class II.

The devices were distributed nationwide in the United States and internationally to numerous countries. Users should refer to the Instructions for Use to ensure proper disinfection and maintenance protocols are followed to prevent potential colonization.

The recalled product

Product
Sorin Stockert Heater-Cooler 3T, 240 V / 60 Hz Temperature control for extracorporeal perfusion of durations up to 6 hours.
Affected units
21

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (1)

  • Product code 16-02-81 Serial number 16S10743-16S11708

Distribution

Distributed nationwide across the United States.

The notice also names 1 state specifically:

Part of a larger recall action

This product is one of 6 recalled by Sorin Group USA, Inc. under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.

See all 6 products in this recall →