Maquet HCU 30 Heater-Cooler Units Recalled for Bacterial Contamination Risk
Maquet Cardiovascular is recalling 305 HCU 30 units because contaminated water, including Mycobacteria, may enter the device circuits.
What this means for you
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall involves a high-risk pathogen (Mycobacteria) in a medical device where injury has not yet been reported, fitting the 'High' severity criteria.
Plain-English summary
Maquet Cardiovascular US Sales, LLC is recalling 305 units of the HCU 30 Heater-Cooler Unit. These devices are used to supply temperature-controlled water for cardioplegia heat exchangers, blood heat exchangers in extra-corporeal circulation, and patient warming or cooling blankets during short-duration cardiopulmonary bypass procedures.
The recall is issued because there is a potential for contaminated water to enter parts of the Heater-Cooler Unit circuits. This creates a potential risk of bacterial contamination, including Mycobacteria species.
The affected units are the HCU 30 Base Unit 100-120 V (Model 704629), HCU 30 Base Unit Canadian/US (Model 705501), and HCU 30 Base Unit 200-240 V (Model 939001). All serial numbers for these models are affected. The units were distributed worldwide and nationwide in the US.
The recalled product
- Product
- Heater Cooler Unit Usage: The Heater-Cooler Unit (HCU 30) supplies temperature-controlled water for cardioplegia heat exchangers and for blood heat exchangers in extra corporeal circulation and for blankets with which patients can be warmed or cooled. The HCU 30 is intended t
- Manufacturer
- Maquet Cardiovascular Us Sales, Llc
- Category
- Medical Device — Cardiovascular
- Affected units
- 305
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (4)
- HCU 30 Base Unit 100-120 V
- Model Number 704629 I 70103.4371 HCU 30 Base Unit Canadian/US
- Model Number 705501 I 70103.4653 HCU 30 Base Unit 200-240 V
- Model Number 939001 I 70102.8718 All serial numbers are affected.
Distribution
Distributed nationwide across the United States.
Related recalls
Same manufacturer · Maquet Cardiovascular Us Sales, Llc
See all →- HighMaquet Cardiovascular Recalls HLS Set Advanced Due to Sterility Breaches
FDA (Devices) · 2020-12-23
- HighMaquet Cardiovascular Recalls HLS Set Advanced Due to Sterility Breach
FDA (Devices) · 2020-12-23
- HighMaquet BMU 40 Blood Monitoring Units Recalled for Electrical Safety Non-Compliance
FDA (Devices) · 2020-11-18
- HighMaquet Yuno Mobile Operating Table Recalled for Electric Shock Risk
FDA (Devices) · 2020-10-28
- HighMaquet Yuno II Mobile Operating Table Recalled for Electric Shock Risk
FDA (Devices) · 2020-10-28
Same hazard · bacterial contamination
See all →- HighClorox Recalls Scented Mistolin and Lestoil Cleaners in Puerto Rico
CPSC · 2026-09-03
- HighCurry chicken salad flagged for possible Listeria contamination in two states
USDA FSIS · 2026-07-30
- ModerateMedline Medical Convenience Kits Recalled Due to Potential Bacillus Contamination
FDA (Devices) · 2026-07-29
- HighMedline Convenience Kits Recalled Over Bacillus Contamination
FDA (Devices) · 2026-07-29
- SevereSurgical Convenience Kit Recalled for Bacterial Contaminant in Alcohol Prep Pads
FDA (Devices) · 2026-07-08