Maquet HCU 30 Heater-Cooler Unit Recalled for Power Supply Board Failure
Maquet is recalling HCU 30 Heater-Cooler Units due to a power supply board defect that can cause overheating and unit shutdown during cardiopulmonary bypass procedures.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall is classified as FDA Class II, and the hazard involves a risk-of-harm product (medical device failure during surgery) where specific injury reports are not detailed in the source text, fitting the 'High' severity criteria for risk-of-harm products without reported injury.
Plain-English summary
Maquet Cardiovascular US Sales, LLC is recalling the HCU 30 Heater-Cooler Unit (HCU) Type 2. This medical device is intended to circulate water through heat exchange circuits to warm or cool a patient during short-duration cardiopulmonary bypass procedures lasting six hours or less. The recall involves 100 units distributed nationwide in the United States.
The recall is due to a failure of the Type 2 power supply board. A voltage drop of more than 10% can cause the R56 thermistor on the power supply board to overheat and fail. This failure typically occurs when the compressor is switched on prior to patient use, but it can also occur during patient use. The failure can result in the shutdown of the compressor or the entire HCU 30 Type 2 unit.
Affected units include the HCU 30 Base Unit 100 (120 V), HCU 30 Base Unit 200 (240 V), and HCU 30 Base Unit Canadian/US. Specifically, the recall covers units with serial numbers less than 93101480 and power supply boards with serial numbers less than 107006001. Healthcare facilities should contact Maquet Cardiovascular US Sales, LLC for instructions on how to proceed with the recall.
The recalled product
- Product
- Maquet HCU 30 Coolant Type R134A Filling weight 250 g 120 V 50-60 Hz 1920 W Product Usage: The Heater Cooler Unit (HCU) 30 is intended to circulate water through heat exchange circuits to warm or cool a patient during short duration cardiopulmonary bypass procedures lasting 6…
- Manufacturer
- Maquet Cardiovascular Us Sales, Llc
- Affected units
- 100
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (3)
- HCU 30 Base Unit 100 - 120 V
- Model Number 704629/70103.4371 HCU 30 Base Unit 200 - 240 V
- Model Number 704631/70103.4642 HCU 30 Base Unit Canadian/US
Distribution
Distributed nationwide across the United States.
Related recalls
Same manufacturer · Maquet Cardiovascular Us Sales, Llc
See all →- HighMaquet Cardiovascular Recalls HLS Set Advanced Due to Sterility Breaches
FDA (Devices) · 2020-12-23
- HighMaquet Cardiovascular Recalls HLS Set Advanced Due to Sterility Breach
FDA (Devices) · 2020-12-23
- HighMaquet BMU 40 Blood Monitoring Units Recalled for Electrical Safety Non-Compliance
FDA (Devices) · 2020-11-18
- HighMaquet Yuno II Mobile Operating Table Recalled for Electric Shock Risk
FDA (Devices) · 2020-10-28
- HighMaquet Yuno Mobile Operating Table Recalled for Electric Shock Risk
FDA (Devices) · 2020-10-28
Same hazard · overheating
See all →- Severe3M Recalls Scotch Thermal Laminators Due to Fire Hazard
CPSC · 2026-10-15
- SevereIntellijoint Navigation Systems Recalled for Potential Fire Risk
FDA (Devices) · 2026-10-14
- CriticalINMO Air3 Smart Glasses Recalled for Overheating Burn Hazard
CPSC · 2026-09-24
- HighGentherm Medical Recalls Blanketrol III Thermal Regulating Systems
FDA (Devices) · 2026-08-26
- ModerateGentherm Medical Recalls Blanketrol II Thermal Regulating Systems
FDA (Devices) · 2026-08-26
Similar recalls
- HighMaquet HCU 30 Heater-Cooler Units Recalled for Bacterial Contamination Risk
FDA (Devices) · 2016-11-09
- HighHeater Cooler Unit Recall Due to Potential Bacterial Contamination Risk
FDA (Devices) · 2021-08-04
- ModerateMaquet VH-3010 Power Supply Recalled for Inadequate Heating Function
FDA (Devices) · 2025-04-23
- HighMaquet BMU 40 Blood Monitoring Units Recalled for Electrical Safety Non-Compliance
FDA (Devices) · 2020-11-18
- ModerateMaquet Compressor Mini Recalled for Excessive Compressed Air Temperature
FDA (Devices) · 2014-02-05
- HighCardioQuip Modular Cooler-Heater Units Recalled for Heating Chamber Malfunction
FDA (Devices) · 2023-12-13