The Recall Desk

FDA (Devices) recall action 71593

Sorin Group USA, Inc. recalled 6 products on 2015-07-22

The FDA (Devices) publishes a recall like this one as 6 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 6 pages.

Products
6
Announced
2015-07-22
Critical
0
Units
5,055

Why these products were recalled

Potential colonization of organisms, including Mycobacteria, in Sorin Heater Cooler Devices, if proper disinfection and maintenance is not performed per Instructions for Use.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–6 of 6

  • HighFDA (Devices)··2015-07-22

    Sorin Stockert Heater-Cooler 3T Recalled for Potential Organism Colonization

    Sorin Group USA is recalling 21 Sorin Stockert Heater-Cooler 3T units due to the potential for organism colonization, including Mycobacteria, if proper disinfection and maintenance are not performed.

    Product
    Sorin Stockert Heater-Cooler 3T, 240 V / 60 Hz Temperature control for extracorporeal perfusion of durations up to 6 hours.
    Category
    Distribution
    1 state
  • HighFDA (Devices)··2015-07-22

    Sorin Stockert Heater-Cooler 3T Recalled for Potential Organism Colonization

    Sorin Group USA is recalling 131 Stockert Heater-Cooler 3T units due to potential colonization of organisms, including Mycobacteria, if proper disinfection and maintenance are not performed.

    Product
    Sorin Stockert Heater-Cooler 3T, 200 V / 50Hz / 60 Hz Temperature control for extracorporeal perfusion of durations up to 6 hours.
    Category
    Distribution
    1 state
  • HighFDA (Devices)··2015-07-22

    Sorin Stockert Heater-Cooler 3T Recalled for Potential Organism Colonization

    Sorin Group USA is recalling 28 Sorin Stockert Heater-Cooler 3T devices due to potential colonization of organisms, including Mycobacteria, if proper disinfection and maintenance are not performed.

    Product
    Sorin Stockert Heater-Cooler 3T, 127 V / 60 Hz Temperature control for extracorporeal perfusion of durations up to 6 hours.
    Category
    Distribution
    1 state
  • HighFDA (Devices)··2015-07-22

    Sorin Stockert Heater-Cooler 3T Recalled for Potential Organism Colonization

    Sorin Group USA, Inc. is recalling 2,837 Sorin Stockert Heater-Cooler 3T units due to potential colonization of organisms, including Mycobacteria, if proper disinfection and maintenance are not performed.

    Product
    Sorin Stockert Heater-Cooler 3T, 230 V Temperature control for extracorporeal perfusion of durations up to 6 hours.
    Category
    Distribution
    1 state
  • HighFDA (Devices)··2015-07-22

    Sorin Stockert Heater-Cooler 3T Recalled for Potential Organism Colonization

    Sorin Group USA is recalling 283 Sorin Stockert Heater-Cooler 3T units due to potential colonization of organisms, including Mycobacteria, if proper disinfection is not performed.

    Product
    Sorin Stockert Heater-Cooler 3T, 208 V / 60 Hz Temperature control for extracorporeal perfusion of durations up to 6 hours.
    Category
    Distribution
    1 state
  • HighFDA (Devices)··2015-07-22

    Sorin Heater-Cooler Devices Recalled for Potential Microbial Colonization

    Sorin Group USA is recalling Stockert Heater-Cooler 3T devices due to the potential for microbial colonization, including Mycobacteria, if proper disinfection and maintenance are not performed.

    Product
    Sorin Stockert Heater-Cooler 3T, 120 V / 60 Hz Temperature control for extracorporeal perfusion of durations up to 6 hours.
    Category
    Distribution
    1 state