Thermo Scientific General Purpose Dynamic Bath Recalled for Assembly Error
Thermo Fisher Scientific is recalling 12 General Purpose Dynamic Bath units due to an assembly error that may cause the fluid temperature to exceed the user's set point.
What this means for you
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall involves a risk-of-harm product where the malfunction could cause fluid temperatures to exceed set points, but the source text does not report any specific injuries or illnesses.
Plain-English summary
Thermo Fisher Scientific (Asheville) LLC is recalling 12 units of the Thermo Scientific General Purpose Dynamic Bath Precision 10L GP Bath, Model Number TSGP10. The recall affects specific serial numbers: 300246811, 300246812, 300246814, 300246815, 300247403, 300247404, 300247406, 300247407, 300247409, 300247789, 300247791, and 300247792.
The recall was initiated because the product may have been assembled incorrectly. This assembly error can cause the bath to malfunction, potentially raising the temperature of the fluid in the reservoir above the temperature set point selected by the user.
The affected units were distributed in California, Georgia, Kentucky, and Pennsylvania, as well as in Canada, China, Germany, Japan, New Zealand, and Australia. Users should check their serial numbers against the list provided and contact the manufacturer for instructions on how to proceed.
The recalled product
- Product
- Thermo Scientific General Purpose Dynamic Bath Precision 10L GP Bath Model Number: TSGP10
- Manufacturer
- Thermo Fisher Scientific (Asheville) LLC
- Hazard
- Affected units
- 12
Distribution
Part of a larger recall action
This product is one of 9 recalled by Thermo Fisher Scientific (Asheville) LLC under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 9 products in this recall →Related recalls
Same manufacturer · Thermo Fisher Scientific (Asheville) LLC
See all →- HighThermo Scientific Dual GP Bath Recalled for Assembly Error
FDA (Devices) · 2019-07-03
- HighThermo Scientific General Purpose Dynamic Bath Recalled for Assembly Error
FDA (Devices) · 2019-07-03
- HighThermo Fisher Water Bath Recall Due to Assembly Error
FDA (Devices) · 2019-07-03
- HighThermo Scientific General Purpose Dynamic Bath Recalled for Assembly Error
FDA (Devices) · 2019-07-03
- HighThermo Fisher Isotemp Water Bath Recalled for Assembly Error
FDA (Devices) · 2019-07-03
Same hazard · overheating
See all →- ModerateGentherm Medical Recalls Blanketrol II Thermal Regulating Systems
FDA (Devices) · 2026-08-26
- HighGentherm Medical Recalls Blanketrol III Thermal Regulating Systems
FDA (Devices) · 2026-08-26
- HighGentherm Medical Recalls Innercool Coolblue Console Due to Overheating Risk
FDA (Devices) · 2026-08-26
- Severe
- SevereTherabody Theragun G4 PRO Charger Recalled for Fire Risk
FDA (Devices) · 2026-08-12