GE Healthcare Recalls Vivid Ultrasound Systems Due to Patient Selection Error
GE Healthcare is recalling Vivid E95, E90, and E80 ultrasound systems because slow search times can cause operators to select the wrong patient, potentially leading to misdiagnosis.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall is classified as FDA Class II, and the hazard involves a risk of misdiagnosis (a risk-of-harm product) where no injuries have yet been reported, fitting the criteria for a score of 3.
Plain-English summary
GE Healthcare, LLC is recalling specific Vivid E95, E90, and E80 ultrasound systems (models H45581DC, H45581LB, and H45581DA). The recall involves 971 units, with 412 located in the United States and 559 outside the US. The affected systems were distributed to numerous countries, including the United Kingdom, Canada, Australia, and various nations in Europe, Asia, and the Middle East.
The recall is due to a software issue where a patient other than the intended one may be incorrectly selected by the operator when the DICOM Worklist search response time is slow. This can result in incorrect patient information appearing on the screen during an exam. If the operator does not detect the error, images from the actual patient may be stored under the wrong patient's record after the exam, which could lead to misdiagnosis.
GE Healthcare reports that there have been no injuries reported as a result of this issue. Healthcare facilities should contact GE Healthcare for instructions on how to resolve the issue and ensure patient data integrity.
The recalled product
- Product
- Vivid E95/E90/E80, H45581DC, H45581LB and H45581DA.
- Manufacturer
- GE Healthcare, LLC
Distribution
Distribution scope not specified by the agency.
Part of a larger recall action
This product is one of 2 recalled by GE Healthcare, LLC under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 2 products in this recall →Related recalls
Same manufacturer · GE Healthcare, LLC
See all →- HighGE SIGNA Premier MRI magnets may have ferrous fittings fitted
FDA (Devices) · 2026-04-08
- HighGE Medical Imaging Software: Inaccurate Measurements in Global Stack Viewport
FDA (Devices) · 2022-08-31
- HighGE Healthcare CT, PET, and NM Patient Tables May Drop During Service
FDA (Devices) · 2022-08-17
- HighGE Centricity PACS Software fails to notify of exam study modifications
FDA (Devices) · 2022-08-03
- SevereGE Healthcare Anesthesia Systems Recalled for Potential Transducer Connection Reversal
FDA (Devices) · 2022-07-27
Same hazard · misdiagnosis
See all →- HighPhilips EPIQ Elite Ultrasound System Recalled for Software Error
FDA (Devices) · 2026-09-30
- HighGE Healthcare Recalls Centricity PACS Software Due to Data Association Errors
FDA (Devices) · 2026-09-23
- HighGE Healthcare Recalls Centricity PACS Server V7 Due to Data Association Error
FDA (Devices) · 2026-09-23
- SevereNova Biomedical Recalls StatStrip Glucose Meters Due to Software Defect
FDA (Devices) · 2026-09-23
- HighGE Healthcare Enterprise Archive V8 Software Recalled for Data Association Errors
FDA (Devices) · 2026-09-23
Similar recalls
- HighGE Vivid Ultrasound Systems Display Incorrect Patient Names
FDA (Devices) · 2024-07-17
- ModerateGE Healthcare Recalls LOGIQ e Ultrasound Systems for Patient Data Linking Error
FDA (Devices) · 2013-08-07
- HighGE Vivid E9 Ultrasound System Recalled for Probe Overheating Risk
FDA (Devices) · 2014-04-30
- HighGE Vivid E90 Ultrasound System Recalled for Electrical Shock Risk
FDA (Devices) · 2018-09-12
- HighGE Vivid Ultrasound Scanner Displays Incorrect Patient Data
FDA (Devices) · 2024-07-17
- HighGE Vivid E80 Ultrasound System Recalled for Power Cord Shock Risk
FDA (Devices) · 2018-09-12