The Recall Desk
HighFDA (Devices)·Z-3132-2026·Announced 2026-09-23

Nova Biomedical Recalls StatStrip Glucose Meters Due to Software Defect

Nova Biomedical is recalling StatStrip Glucose and B-Ketone meters because a software defect may cause patient data to be assigned to the wrong medical record.

What this means for you

Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.

Our severity reasoning: The recall involves a risk-of-harm product where a software defect can lead to inappropriate clinical intervention, but no specific injuries or illnesses have been reported in the source text.

Plain-English summary

Nova Biomedical Corporation is recalling 2,044 units of the StatStrip Glucose and B-Ketone meter, Model No. 66429. The recall covers software versions 0.00.13.10 through 1.5.14.25. The devices were distributed nationwide in the United States and in 22 foreign countries, including Canada, the United Kingdom, and Germany.

The recall is due to a software synchronization and timing issue in the barcode scanning and patient data processing component. This defect can cause the device to intermittently retain a previously scanned barcode, such as a patient ID or quality control solution ID, when a new barcode is scanned. If an operator accepts the incorrectly displayed patient information, the current patient's glucose result may be assigned to the previous patient's medical record. Consequently, the current patient's record will not receive the result, leaving their glucose status unrecorded. The previous patient may receive inappropriate clinical intervention based on the erroneous result, particularly in critically ill patients, neonates, and infants.

The defect does not affect the analytical accuracy of the glucose measurements themselves. The issue also affects quality control barcode scanning, potentially causing QC results to be attributed to incorrect identifiers or QC tests to fail without the user recognizing the cause. Consumers and healthcare providers should stop using the affected devices and contact Nova Biomedical Corporation for further instructions.

The recalled product

Product
StatStrip Glucose and B- Ketone meter. Model No: 66429. Glucose test system.
Affected units
2,044

Distribution

Distributed nationwide across the United States.

Similar recalls