Philips Ingenia Blue Seal MRI Systems Recalled for DTI Export Errors
Philips is recalling six Ingenia Blue Seal MRI systems because diffusion tensor imaging exports contain incorrect data, preventing accurate tractography reconstruction.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 2/5 (Moderate).
Minor labeling or packaging issue, or a voluntary precautionary recall. Risk of harm is low but real — follow the manufacturer's instructions.
Our severity reasoning: The recall is classified as FDA Class II. The hazard involves incorrect data export affecting specialized analysis, with no reported injuries, illnesses, or deaths in the source text, fitting the criteria for a moderate severity software defect.
Plain-English summary
Philips North America is recalling six Ingenia Blue Seal SE and Blue Seal XE MRI systems. The affected units are identified by model numbers 782185 (SE) and 782192 (XE) and software versions R11, R12, and R13. The recall covers specific serial numbers: 53002 for the SE model, and 56000, 56001, 56005, 56010, and 56012 for the XE model.
The recall is due to a software defect where exporting diffusion tensor imaging (DTI) scans using the Enhanced MR DICOM format results in incorrect diffusion direction data. If this exported data is used with third-party software for tractography, such as mapping nerve fiber pathways in the brain, the software cannot correctly process the data for reconstruction.
The affected units were distributed worldwide, including in the United States and numerous international locations. The source text does not specify actions for consumers or patients, as the issue pertains to data export functionality for specialized imaging analysis.
The recalled product
- Product
- Ingenia Blue Seal SE and Blue Seal XE; Model number (REF): 1. 782185 (SE), 2. 782192 (XE); Software Versions: R11, R12 and R13;
- Manufacturer
- Philips North America
- Category
- Medical Device — MRI System
- Affected units
- 6
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (10)
- 1. Model number (REF): 1. 782185 (SE)
- UDI-DI: 00884838125698
- Serial Number: 53002
- 2. Model Number (REF): 782192 (XE)
- UDI-DI: 00884838128422
- Serial Numbers: 56000
- 56001
- 56005
- 56010
- 56012
Distribution
Distributed nationwide across the United States.
Related recalls
Same manufacturer · Philips North America
See all →- HighPhilips Model 712010 Medical Device Recalled for Potential Immediate Failure
FDA (Devices) · 2026-11-04
- ModeratePhilips Smart-hopping 2.0 AP Recall for Wireless Roaming Issues
FDA (Devices) · 2026-10-28
- ModeratePhilips SmartPath to MR 7700 Recalled for Incorrect DTI Export Data
FDA (Devices) · 2026-10-07
- ModeratePhilips MR 5300 Recalled for Incorrect Diffusion Tensor Imaging Data
FDA (Devices) · 2026-10-07
- ModeratePhilips MR 7700 Software Upgrade Recalled for Incorrect DTI Data
FDA (Devices) · 2026-10-07
Same hazard · software defect
See all →- High
- SevereWaymo Recalls Automated Driving Systems for Closed Freeway Lane Risk
NHTSA · 2026-11-20
- SevereRam Recalls 2022-2024 1500 Classic Trucks for ABS Software Defect
NHTSA · 2026-11-19
- ModerateSiemens ACUSON Orgin ICE Ultrasound System Software Recall
FDA (Devices) · 2026-11-18
- ModerateSiemens ACUSON Sequoia Ultrasound System Recalled for Software Issue
FDA (Devices) · 2026-11-18
Similar recalls
- ModeratePhilips Recalls Ingenia MRI Systems Due to Incorrect DTI Export Data
FDA (Devices) · 2026-10-07
- ModeratePhilips Ingenia MRI Systems Recalled for Incorrect DTI Data Exports
FDA (Devices) · 2026-10-07
- ModeratePhilips Ingenia MRI Systems Recalled for Incorrect DTI Export Data
FDA (Devices) · 2026-10-07
- ModeratePhilips Ingenia Ambition MRI Systems Recalled for Incorrect DTI Data Exports
FDA (Devices) · 2026-10-07
- ModeratePhilips MR 7700 MRI Systems Recalled for Incorrect DTI Data Exports
FDA (Devices) · 2026-10-07
- ModeratePhilips MRI Systems Recalled for Incorrect Diffusion Tensor Imaging Data Exports
FDA (Devices) · 2026-10-07