The Recall Desk
SevereFDA (Devices)·Z-1602-2020·Announced 2020-04-08

Stryker Recalls Ceiling Mounted Surgical Light Systems Due to Fall Risk

Stryker is recalling 1,415 Berchtold Chromophare Ceiling Mounted Surgical Light Systems because a missing snap ring or spanner nut may cause the light to fall.

What this means for you

Illness, injury, or structural failure has been reported. Stop using the product immediately and contact the manufacturer for a refund or repair.

Our severity reasoning: The source text explicitly states that the falling light may potentially cause serious injuries, which aligns with the rubric criterion for a score of 4 (Severe) involving significant injury reports or potential.

Plain-English summary

Stryker Communications is recalling 1,415 Berchtold Chromophare Ceiling Mounted Surgical Light Systems, Catalog #CH13000000. The recall affects units with serial numbers ranging from 091316 70016 through 083117-73252. These products were distributed nationwide in the United States and in Canada.

The recall was initiated because the firm became aware that a missing snap ring or spanner nut may result in insufficient mount force. This defect could cause the surgical light to fall from its suspension. If the light falls, it may potentially cause serious injuries to health care providers and/or patients.

Health care providers should stop using the affected surgical lights and contact Stryker Communications for instructions on how to resolve the issue.

The recalled product

Product
Berchtold Chromophare Ceiling Mounted Surgical Light System, Catalog #CH13000000
Affected units
1,415

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (1)

  • Serial Number Range: 091316 70016 through 083117-73252

Distribution

Distributed nationwide across the United States.