Siemens Medical Linear Accelerators Recalled for Defective Weld Seams
Siemens is recalling 35 MEVATRON M 6730/6740 linear accelerators because defective weld seams on overhead suspensions could break and fall, posing a risk of severe injury.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 4/5 (Severe).
The hazard poses a serious risk of illness, injury, or structural failure. Stop using the product immediately and contact the manufacturer for a refund or repair.
Our severity reasoning: The source text explicitly states that the defect can lead to a 'severe injury' to a patient or other person, which aligns with the rubric criterion for 'significant injury' reports or risks, warranting a score of 4 (Severe).
Plain-English summary
Siemens Medical Solutions USA, Inc. is recalling 35 units of the MEVATRON M 6730/6740 (MEX) linear accelerator systems. These devices are intended to deliver x-ray radiation for the therapeutic treatment of cancer. The recall is due to a defective weld seam located at the overhead suspensions.
The defect creates a risk that the weld seam may break, causing the overhead suspension to fall. If this occurs, it could result in severe injury to a patient or any other person in the vicinity. The affected units were distributed nationwide.
Healthcare facilities should contact Siemens Medical Solutions USA, Inc. for instructions on how to address the defect. Consumers and patients should not use the affected equipment until the issue is resolved by the manufacturer.
The recalled product
- Product
- MEVATRON M 6730/6740 (MEX) Accelerator, Linear, Medical The intended use of the SIEMENS branded PRIMUS family of linear accelerator systems is to deliver x-ray radiation for therapeutic treatment of cancer
- Manufacturer
- Siemens Medical Solutions USA, Inc
- Affected units
- 35
Distribution
Distributed nationwide across the United States.
Part of a larger recall action
This product is one of 6 recalled by Siemens Medical Solutions USA, Inc under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 6 products in this recall →Related recalls
Same manufacturer · Siemens Medical Solutions USA, Inc
See all →- ModerateARTIS icono biplane medical imaging system detector lift malfunction
FDA (Devices) · 2026-07-08
- HighARTIS icono floor Medical Imaging System Detector Movement Loss Recall
FDA (Devices) · 2026-07-08
- ModerateARTIS icono ceiling medical imaging monitor suspension system
FDA (Devices) · 2026-07-08
- HighARTIS Icono Ceiling Model 11328100 Loss of Movement Capability Recall
FDA (Devices) · 2026-07-08
- HighARTIS pheno systems may display zero dose while dose is applied
FDA (Devices) · 2026-06-10
Same hazard · falling object
See all →- SevereLithonia Lighting Recalls RSX Luminaires with Adjustable Tilt Arms
CPSC · 2026-10-22
- HighMaquet PowerLED 300 Concealed Lighting System Recalled for Detachment Risk
FDA (Devices) · 2026-09-30
- HighMirion Captus 4000e Thyroid Uptake System Recalled for Spring Arm Failure
FDA (Devices) · 2026-09-23
- HighFord F-650 SD Recall: Liftgate Platform Deck Pins May Break
NHTSA · 2026-08-20
- SevereTerex Digger Derrick Trucks Recalled for Weld Defect
NHTSA · 2026-08-17
Similar recalls
- HighSiemens PRIMUS Linear Accelerators Recalled for Software Safety Issues
FDA (Devices) · 2016-10-05
- HighSiemens MEVATRON M2 Linear Accelerator Recalled for Collision Risk
FDA (Devices) · 2018-04-25
- SevereSiemens Linear Accelerator Recall Due to Potential Patient Mistreatment
FDA (Devices) · 2013-07-10
- SevereSiemens PRIMUS Radiation Therapy Systems Recalled for Collimator and Weld Defects
FDA (Devices) · 2017-02-22
- HighSiemens Linear Accelerators Recalled for Potential Patient Collision Risk
FDA (Devices) · 2014-08-20
- HighSiemens Oncor Avant-garde Radiation Therapy Systems Recalled for Sensor Defects
FDA (Devices) · 2017-02-22