Medical Device Thawing Kit Recalls Due to Spanish Instruction Translation Error
CooperSurgical recalls its LifeGlobal Fast Freeze Thawing Kit due to a Spanish instruction translation error that specifies 5 minutes instead of 5 seconds for blastocyte air thawing, affecting 403 units.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: FDA Class II recall involving a critical instruction error that could affect specimen viability and medical procedure outcomes. No illnesses, injuries, or deaths have been reported, limiting the score per the rubric's guidance that theoretical risks without reported harm score at most 3.
Plain-English summary
CooperSurgical, Inc. is recalling the LifeGlobal Fast Freeze Thawing Kit, Model No. GFT5-055, due to a translation error in the Spanish-language Instructions for Use (IFU). The Spanish version of the IFU incorrectly instructs users to allow blastocytes to air thaw for 5 minutes, while the correct procedure specified in the English version requires only 5 seconds of air thawing.
This discrepancy could result in improper thawing of embryos used in assisted reproductive procedures, potentially affecting the viability and quality of the specimens. The error affects 403 units distributed across multiple U.S. states (including California, Florida, New York, Texas, and others) and internationally in Belgium, Canada, Croatia, Netherlands, Portugal, and Spain.
Lot numbers affected by this recall include 21335060, GFT5-200218U, and GFT5-201026U. Users of this device should immediately verify that they are following the English version of the IFU or contact CooperSurgical for a corrected Spanish translation. No illnesses or injuries have been reported in connection with this recall.
The recalled product
- Product
- LifeGlobal Fast Freeze Thawing Kit, Model No. GFT5-055
- Manufacturer
- CooperSurgical, Inc.
- Category
- Medical Device — Embryo Thawing
- Affected units
- 403
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (2)
- GFT5-200218U
- GFT5-201026U
Distribution
Related recalls
Same manufacturer · CooperSurgical, Inc.
See all →- SevereCooperSurgical Recalls Neonatal Nasal CPAP Replacement Sets Due to Loose Tubing
FDA (Devices) · 2026-09-30
- SevereCooperSurgical Recalls Infant Nasal CPAP Tubing Due to Loose Connection Risk
FDA (Devices) · 2026-09-30
- HighEndosee system convenience trays label non-sterile iodine pouches as sterile
FDA (Devices) · 2025-07-30
- HighHSG procedure trays label non-sterile iodine pouches as sterile
FDA (Devices) · 2025-07-30
- HighEndosee convenience kits label non-sterile iodine pouches as sterile
FDA (Devices) · 2025-07-30
Same hazard · translation error
See all →- HighTandem insulin pump Czech guide inverted a key connection instruction
FDA (Devices) · 2026-04-01
- HighOlympus Soltive Premium SuperPulsed Laser Recalled for Translation Labeling Error
FDA (Devices) · 2024-10-30
- HighVentilator Instructions for Use Contain Translation Error in Korean Materials
FDA (Devices) · 2023-07-12
- HighHeartWare HVAD Controller: Translation Errors in Critical Documentation
FDA (Devices) · 2022-05-25
- HighHeartWare HVAD Ventricular Assist Device Recalled for Translation Errors
FDA (Devices) · 2022-05-25
Similar recalls
- ModerateMedical device recall: sterilization validator with incorrect incubation time
FDA (Devices) · 2025-03-05
- LowCooperSurgical Recalls Sage CSC Thawing Medium Due to Incorrect Package Insert
FDA (Devices) · 2016-09-28
- ModerateIntegra LifeSciences Recalls C7000 CUSA Clarity Ultrasonic Tissue Ablation Systems
FDA (Devices) · 2019-12-18
- ModerateIntegra C7000 CUSA Clarity System Recalled for Inadequate Sterilization Tray Drying Time
FDA (Devices) · 2019-12-18
- ModeratePrismaflex HF20 Medical Device Recall Due to Mistranslated Instructions
FDA (Devices) · 2022-10-19
- HighInstrumentation Laboratory Recalls 722 ROTEM ex-tem Kits Due to Prolonged Clotting Times
FDA (Devices) · 2019-09-18