The Recall Desk

Manufacturer

CooperSurgical, Inc.

84 recalls in our database name CooperSurgical, Inc. as the manufacturing or recalling firm.

What the numbers show

Total
84
Critical
0
Severe
4
Most recent
2025-07-30

Of the 84 recalls from CooperSurgical, Inc. we have scored against our rubric, 0 (0%) are Critical — deaths or serious injuries reported, or the agency’s own highest classification — and 4 (5%) are Severe.

Which agencies issued them

About this listing

The manufacturer string comes from the agency’s own source notice — for FDA recalls, the “recalling firm” field; for NHTSA recalls, the vehicle manufacturer; for CPSC recalls, the firm-of-record on the joint press release. A single corporate family can appear under multiple manufacturer strings as agency notices aren’t standardized on a canonical name. If you don’t see recalls you expect, try a parent-company search or a broader full-text search.

Severity scores below are computed from the agency’s classification and the recall text, never from manufacturer behaviour or history. Many manufacturers issue routine corrective actions (software updates, labelling corrections) that are correctly scored low — having recalls on this list isn’t inherently a quality signal.

1–25 of 84

  • HighFDA (Devices)·Z-2136-2025·2025-07-30

    Endosee system convenience trays label non-sterile iodine pouches as sterile

    Endosee System Convenience Kits are being recalled because they contain non-sterile iodine pouches while the peel away lid labels incorrectly identify them as sterile. 991 units are affected.

    Product
    Endosee System Convenience Kit. Model Number: ES-TRAY. Endosee System Convenience Kit
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2137-2025·2025-07-30

    Endosee convenience kits label non-sterile iodine pouches as sterile

    Endosee System Convenience Kits with IV Tube are being recalled because they contain non-sterile iodine pouches while the peel away lid labels incorrectly identify them as sterile. 475 units affected.

    Product
    Endosee System Convenience Kit with IV Tube. Model Number: ES-TRAY-IV. Endosee System Convenience Kit with IV Tube.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2135-2025·2025-07-30

    HSG procedure trays label non-sterile iodine pouches as sterile

    HSG Procedure Trays are being recalled because they contain non-sterile iodine pouches while the peel away lid labels incorrectly identify them as sterile. 171 units are affected.

    Product
    HSG Procedure Tray. Model Number: 6050T. The HSG Procedure Tray is for use in hysterosalpingography and sonohysterography procedures.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0689-2025·2024-12-25

    K-Systems T47 Warming Plate may exceed desired temperatures

    The K-Systems T47 Warming Plate has a temperature control defect where warming surfaces may exceed desired set temperatures. CooperSurgical is recalling 22 affected units distributed to fertility clinics.

    Product
    Brand Name: K-Systems Product Name: T47 Warming Plate Model/Catalog Number: K23054 Software Version: N/A Product Description: An electrically-powered device intended to provide suitable warming and/or incubation conditions for reproductive biological materials (e.g., gametes
    Category
    Medical Device
    Distribution
    14 states
  • HighFDA (Devices)·Z-0690-2025·2024-12-25

    K-Systems T47 Warming Plate May Exceed Set Temperatures

    CooperSurgical recalls K-Systems T47 Warming Plates because temperature may exceed the desired set value, potentially affecting biological materials. The device alarms appropriately when this occurs.

    Product
    Brand Name: K-Systems Product Name: T47 Warming Plate Model/Catalog Number: K23055 Software Version: N/A Product Description: An electrically-powered device intended to provide suitable warming and/or incubation conditions for reproductive biological materials (e.g., gametes
    Category
    Medical Device
    Distribution
    14 states
  • HighFDA (Devices)·Z-0687-2025·2024-12-25

    K-Systems G73 Dry Bath Incubator may exceed target temperature

    CooperSurgical is recalling K-Systems G73 Dry Bath Incubators because warming surfaces may exceed the desired set temperature. The device alarms when this occurs.

    Product
    Brand Name: K-Systems Product Name: G73 Dry Bath Incubator Model/Catalog Number: K22065 Product Description: An electrically-powered device intended to provide suitable warming and/or incubation conditions for reproductive biological materials (e.g., gametes and embryos), at c
    Category
    Medical Device
    Distribution
    14 states
  • HighFDA (Devices)·Z-0692-2025·2024-12-25

    K-Systems Incubator Device Recall: Temperature May Exceed Desired Setting

    CooperSurgical is recalling 34 K-Systems incubator devices used in assisted reproduction procedures. The warming surfaces may exceed the desired set temperature, though the device will alarm if this occurs.

    Product
    Brand Name: K-Systems Product Name: Multiple/unknown Model/Catalog Number: Multiple/unknown Software Version: N/A Product Description: An electrically-powered device intended to provide suitable warming and/or incubation conditions for reproductive biological materials (e.g.,
    Category
    Medical Device
    Distribution
    14 states
  • HighFDA (Devices)·Z-0688-2025·2024-12-25

    K-Systems G85 Mini Incubator May Exceed Desired Temperature

    CooperSurgical recalls 37 K-Systems G85 Mini Incubators because the device warming surfaces may exceed the desired set temperature. The unit alarms when this occurs, but affected serial numbers require verification.

    Product
    Brand Name: K-Systems Product Name: G85 Mini Incubator Model/Catalog Number: K22074 Software Version: N/A Product Description: An electrically-powered device intended to provide suitable warming and/or incubation conditions for reproductive biological materials (e.g., gametes
    Category
    Medical Device
    Distribution
    14 states
  • ModerateFDA (Devices)·Z-0691-2025·2024-12-25

    K-Systems R65 embryo incubator temperature may exceed setpoint

    CooperSurgical's K-Systems R65 Trolley reproductive incubators may experience temperature excursions above desired setpoints, though devices will alarm when this occurs. The FDA issued a Class II recall of 43 units across eight U.S. states.

    Product
    Brand Name: K-Systems Product Name: R65 Trolley Model/Catalog Number: K24012 Software Version: N/A Product Description: An electrically-powered device intended to provide suitable warming and/or incubation conditions for reproductive biological materials (e.g., gametes and em
    Category
    Medical Device
    Distribution
    14 states
  • HighFDA (Devices)·Z-3297-2024·2024-10-02

    CooperSurgical Recalls Origio Sperm Wash Medium Due to Low Sodium Pyruvate

    CooperSurgical is recalling Origio Sperm Wash Medium because low sodium pyruvate levels could compromise gamete development.

    Product
    Origio Sperm Wash Medium Pack of Five 60ml bottles - Intended for washing of spermatozoa. Part Number: 84055060A
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3298-2024·2024-10-02

    Origio Sperm Wash Medium Recalled for Low Sodium Pyruvate Concentration

    CooperSurgical is recalling Origio Sperm Wash Medium due to low sodium pyruvate concentration. The deficiency could compromise gamete development in assisted reproductive procedures.

    Product
    Origio Sperm Wash Medium Pack of Five 60ml bottles - Intended for washing of spermatozoa. Part Number: 84055060D
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3296-2024·2024-10-02

    Origio Sperm Wash Medium Recalled Due to Low Sodium Pyruvate

    CooperSurgical is recalling Origio Sperm Wash Medium due to low sodium pyruvate concentration that could compromise sperm cell development. This affects 980 units distributed worldwide, including throughout the US.

    Product
    Origio Sperm Wash Medium Pack of Ten 10ml bottles - Intended for washing of spermatozoa Part Number: 84051010A
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-3188-2024·2024-10-02

    CooperSurgical Recalls Silver-Plated Copper Sims Uterine Sound for Incorrect Marking

    CooperSurgical is recalling 468 units of Silver-Plated Copper Sims Uterine Sounds because they were laser-marked with the incorrect part number during manufacturing.

    Product
    Silver-Plated Copper Sims Uterine Sound, Part Number 64-600
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-3189-2024·2024-10-02

    Stainless Steel Sims Uterine Sound Recalled for Incorrect Part Number

    CooperSurgical, Inc. is recalling the Stainless Steel Sims Uterine Sound (Part Number 64-601) because some units were laser-marked with incorrect part numbers during manufacturing. No injuries have been reported.

    Product
    Stainless Steel Sims Uterine Sound, Part Number 64-601
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1089-2024·2024-02-21

    global Medium embryo culture media recalled for potential development impairment

    CooperSurgical is recalling three specific lots of global Medium embryo culture media distributed to fertility clinics worldwide due to performance issues that may impair embryo development during culture.

    Product
    global Medium, Model Numbers LGGG-020 (20mL bottle), LGGG-050 (50mL bottle), LGGG-100 (100mL bottle). Intended use: Culture of human embryos from zygote to blastocyst, embryo transfer.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0985-2024·2024-02-14

    Carter-Thomason CloseSure Surgical Kit Recall: Mislabeled Component Descriptions

    CooperSurgical is recalling Carter-Thomason CloseSure System XL surgical kits due to incorrect component descriptions on packaging. The affected batch includes 127 boxes (381 total kits) with lot number 326728.

    Product
    Box as labeled: Regard, Carter-Thomason CloseSure System XL, REF 733776, Quantity 1 Box of 3 Systems. Kit as labeled: Regard, Carter-Thomason CloseSure System, REF 733776, Contents: One (1) Carter-Thomason Suture Passer, One (1) 5 mm Pilot Guide, One (1) 10/12 mm Pilot Guide.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0504-2024·2023-12-13

    Medical Catheter Sets Recalled Due to Bonding Material Defect

    CooperSurgical H/S Catheter Sets may have defective bonding material causing component detachment or leakage. The sets are used in hysterosalpingography procedures to detect uterine abnormalities.

    Product
    CooperSurgical H/S CATHETER SET FR 5 with Integated Stylet Model: 61-4005S For administering contrast media during Hysterosalpingography or Hysterosonography procedures to detect uterine pathology such as polyps, fibroids, adhesions or endometrial thickening, and/or patency of
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0502-2024·2023-12-13

    Cooper Surgical catheter recall due to component detachment risk

    The Cooper Surgical H/S ELLIPTOSPHERE CATH 5 Fr catheter may detach or leak due to defective bonding material used in assembly. The 73 affected units were distributed nationwide and internationally.

    Product
    Cooper Surgical H/S ELLIPTOSPHERE CATH 5 Fr -For administering contrast media during Hysterosalpingography or Hysterosonography procedures to detect uterine pathology such as polyps, fibroids, adhesions or endometrial thickening, and/or patency of fallopian tubes. Model: 61-3005
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0503-2024·2023-12-13

    Medical Catheter Set Recalled for Faulty Bonding Material

    CooperSurgical is recalling H/S ELLIPTOSPHERE CATH 5 FR Sets (169 units) due to incorrect bonding material that may cause component detachment or leakage, potentially delaying diagnostic procedures.

    Product
    Cooper Surgical H/S ELLIPTOSPHERE CATH 5 FR Set-For administering contrast media during Hysterosalpingography or Hysterosonography procedures to detect uterine pathology such as polyps, fibroids, adhesions or endometrial thickening, and/or patency of fallopian tubes Model: 61-40
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1348-2023·2023-04-12

    CooperSurgical embryo culture medium product substitution recall

    CooperSurgical is recalling Global Total LP embryo culture medium after discovering affected units may contain a different medium lacking the required protein source for embryo development.

    Product
    Global Total LP single step medium, 60mL, bicarbonate-buffered media for D1-5 embryo culture and transfer, Reference Number H5GT-060
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1044-2023·2023-02-08

    Biopsy Medium for In Vitro Fertilization Recalled Due to Product Contamination

    CooperSurgical is recalling a lot of Biopsy Medium used in IVF procedures because the product may contain an incorrect medium that could degrade embryos. No illnesses have been reported.

    Product
    BIOPSY MEDIUM 10ML, Reference Number 10620010 A, used for blastomere biopsy of cleavage stage embryos for Pre-implantation Genetic Diagnosis (PGD)
    Category
    Medical Device
    Distribution
    0 states
  • SevereFDA (Devices)·Z-0223-2023·2022-11-16

    Neonatal Endotracheal Tube Grip with Detachable Metal Clips Recalled

    CooperSurgical is recalling NEO-fit Neonatal Endotracheal Tube Grips due to metal clips that can become loose or detach. Eleven complaints include three cases of clip ingestion, three removals from the mouth, and one throat laceration.

    Product
    NEO-fit Neonatal Endotracheal Tube Grip, Part Number 42-2540
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1822-2022·2022-10-05

    Humidifier Bottle and Syringe Filter Kit Sterile Barrier Breach Recall

    CooperSurgical is recalling BT37 benchtop incubator humidifier and filter kits due to a potential sterile barrier breach. The defect may allow unsterilized devices to contaminate and degrade embryos during incubation.

    Product
    Humidifier bottle and syringe filter kit (6 count of each) for the BT37 Mark I Benchtop Incubator (AY102295). The Mark I model is no longer produced, but the legacy humidifier bottle component design can also be used with the BT37 Mark II Benchtop Incubator.
    Category
    Medical Device
    Distribution
    25 states
  • HighFDA (Devices)·Z-1823-2022·2022-10-05

    CooperSurgical incubator humidifier kit recall due to sterile barrier breach

    CooperSurgical is recalling 5,182 humidifier bottle and filter kits for BT37 Mark II incubators due to potential sterile barrier damage from packaging defects. This could result in use of unsterilized components, risking embryo contamination.

    Product
    Humidifier bottle and syringe filter kit (6 count of each) for the BT37 Mark II Benchtop Incubator (AY200246)
    Category
    Medical Device
    Distribution
    25 states
  • HighFDA (Devices)·Z-1421-2022·2022-07-27

    Medical Device Thawing Kit Recalls Due to Spanish Instruction Translation Error

    CooperSurgical recalls its LifeGlobal Fast Freeze Thawing Kit due to a Spanish instruction translation error that specifies 5 minutes instead of 5 seconds for blastocyte air thawing, affecting 403 units.

    Product
    LifeGlobal Fast Freeze Thawing Kit, Model No. GFT5-055
    Category
    Medical Device
    Distribution
    31 states