CooperSurgical Recalls Infant Nasal CPAP Tubing Due to Loose Connection Risk
CooperSurgical is recalling specific lots of INCA Infant Nasal CPAP tubing because a loose connection may cause loss of pressure and reduced oxygenation.
What this means for you
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall involves a medical device for infants where a defect could lead to decreased oxygenation, representing a risk-of-harm product where injury has not yet been reported, which aligns with the 'High' severity criteria.
Plain-English summary
CooperSurgical, Inc. is recalling 2,643 boxes (13,215 units) of INCA (Infant Nasal Cannulae Assy) Infant Nasal CPAP, Complete Set, specifically the Stay-Flex tubing component. The affected model numbers are 44-2707, 44-2709, 44-2710, 44-2712, and 44-2715. The recall covers specific lot codes and UDI-DIs listed in the agency's structured data.
The recall was initiated because the flexible tubing may exhibit a loose connection with the nasal cannula during assembly or use. This defect may result in a loss of CPAP pressure, which can reduce the effectiveness of treatment and lead to decreased oxygenation until the connection is restored.
The products were distributed nationwide to Connecticut, Florida, Georgia, Iowa, Illinois, Indiana, Maryland, Michigan, Mississippi, New Jersey, New York, Ohio, Pennsylvania, and Virginia, as well as internationally to Brazil and Peru. Consumers and healthcare providers should check their inventory against the specific model numbers and lot codes provided by the FDA and contact CooperSurgical for instructions on how to return the affected units.
The recalled product
- Product
- INCA (Infant Nasal Cannulae Assy) Infant Nasal CPAP, Complete Set. Stay-Flex tubing component. 5 units per box. Model Numbers: 44-2707, 44-2709, 44-2710, 44-2712, 44-2715.
- Manufacturer
- CooperSurgical, Inc.
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (7)
- 550033781
- 550031616
- 550031620
- 550031614
- 550033779
- 550033780
- 550034763.
Distribution
Related recalls
Same manufacturer · CooperSurgical, Inc.
See all →- HighCooperSurgical Recalls Neonatal Nasal CPAP Replacement Sets Due to Loose Tubing
FDA (Devices) · 2026-09-30
- HighEndosee system convenience trays label non-sterile iodine pouches as sterile
FDA (Devices) · 2025-07-30
- HighEndosee convenience kits label non-sterile iodine pouches as sterile
FDA (Devices) · 2025-07-30
- HighHSG procedure trays label non-sterile iodine pouches as sterile
FDA (Devices) · 2025-07-30
- HighK-Systems G85 Mini Incubator May Exceed Desired Temperature
FDA (Devices) · 2024-12-25
Same hazard · loss of pressure
See all →- HighBoston Scientific Hurricane RX Dilation Balloon Recalled for Pressure Loss Defect
FDA (Devices) · 2021-12-08
- HighFord Recalls 2020 F-150 Trucks Due to Damaged Spare Tires
NHTSA · 2019-12-18
- SevereToyo Recalls Tires Due to Tread and Belt Separation Risk
NHTSA · 2014-11-04
- SevereFederal Couragia A/T Tires Recalled for Steel Belt Separation
NHTSA · 2014-04-18
- HighDePuy Spine Recalls Confidence Spinal Cement System 5cc Kits
FDA (Devices) · 2013-09-11