The Recall Desk
Agency classClass IOur severity:SevereFDA (Devices)·Z-3139-2026·Announced 2026-09-30

CooperSurgical Recalls Infant Nasal CPAP Tubing Due to Loose Connection Risk

CooperSurgical is recalling specific lots of INCA Infant Nasal CPAP tubing because a loose connection may cause loss of pressure and reduced oxygenation.

The agency’s classification

FDA (Devices): Class I

Class I is a situation in which there is a reasonable probability that the use of or exposure to a violative product will cause serious adverse health consequences or death.

This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work

What this means for you

Our assessment, not the agency’s: severity 4/5 (Severe).

The hazard poses a serious risk of illness, injury, or structural failure. Stop using the product immediately and contact the manufacturer for a refund or repair.

Our severity reasoning: The agency classified this recall as Class I — a situation in which there is a reasonable probability that the use of or exposure to a violative product will cause serious adverse health consequences or death. On that classification alone this is scored Severe; a score of Critical is reserved for recalls where a death or serious injury was reported. The agency gives the reason as: "Potential to exhibit a loose connection between the flexible tubing and the nasal cannula, during assembly or while in use, which may result in loss of CPAP pressure, possibly reducing the effectiveness of treatment and lead to decreased…"

Plain-English summary

CooperSurgical, Inc. is recalling 2,643 boxes (13,215 units) of INCA (Infant Nasal Cannulae Assy) Infant Nasal CPAP, Complete Set, specifically the Stay-Flex tubing component. The affected model numbers are 44-2707, 44-2709, 44-2710, 44-2712, and 44-2715. The recall covers specific lot codes and UDI-DIs listed in the agency's structured data.

The recall was initiated because the flexible tubing may exhibit a loose connection with the nasal cannula during assembly or use. This defect may result in a loss of CPAP pressure, which can reduce the effectiveness of treatment and lead to decreased oxygenation until the connection is restored.

The products were distributed nationwide to Connecticut, Florida, Georgia, Iowa, Illinois, Indiana, Maryland, Michigan, Mississippi, New Jersey, New York, Ohio, Pennsylvania, and Virginia, as well as internationally to Brazil and Peru. Consumers and healthcare providers should check their inventory against the specific model numbers and lot codes provided by the FDA and contact CooperSurgical for instructions on how to return the affected units.

The recalled product

Product
INCA (Infant Nasal Cannulae Assy) Infant Nasal CPAP, Complete Set. Stay-Flex tubing component. 5 units per box. Model Numbers: 44-2707, 44-2709, 44-2710, 44-2712, 44-2715.

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (7)

  • 550033781
  • 550031616
  • 550031620
  • 550031614
  • 550033779
  • 550033780
  • 550034763.

Distribution

Distributed nationwide across the United States.

The notice also names 14 states specifically: