CooperSurgical Recalls Infant Nasal CPAP Tubing Due to Loose Connection Risk
CooperSurgical is recalling specific lots of INCA Infant Nasal CPAP tubing because a loose connection may cause loss of pressure and reduced oxygenation.
The agency’s classification
FDA (Devices): Class I
Class I is a situation in which there is a reasonable probability that the use of or exposure to a violative product will cause serious adverse health consequences or death.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 4/5 (Severe).
The hazard poses a serious risk of illness, injury, or structural failure. Stop using the product immediately and contact the manufacturer for a refund or repair.
Our severity reasoning: The agency classified this recall as Class I — a situation in which there is a reasonable probability that the use of or exposure to a violative product will cause serious adverse health consequences or death. On that classification alone this is scored Severe; a score of Critical is reserved for recalls where a death or serious injury was reported. The agency gives the reason as: "Potential to exhibit a loose connection between the flexible tubing and the nasal cannula, during assembly or while in use, which may result in loss of CPAP pressure, possibly reducing the effectiveness of treatment and lead to decreased…"
Plain-English summary
CooperSurgical, Inc. is recalling 2,643 boxes (13,215 units) of INCA (Infant Nasal Cannulae Assy) Infant Nasal CPAP, Complete Set, specifically the Stay-Flex tubing component. The affected model numbers are 44-2707, 44-2709, 44-2710, 44-2712, and 44-2715. The recall covers specific lot codes and UDI-DIs listed in the agency's structured data.
The recall was initiated because the flexible tubing may exhibit a loose connection with the nasal cannula during assembly or use. This defect may result in a loss of CPAP pressure, which can reduce the effectiveness of treatment and lead to decreased oxygenation until the connection is restored.
The products were distributed nationwide to Connecticut, Florida, Georgia, Iowa, Illinois, Indiana, Maryland, Michigan, Mississippi, New Jersey, New York, Ohio, Pennsylvania, and Virginia, as well as internationally to Brazil and Peru. Consumers and healthcare providers should check their inventory against the specific model numbers and lot codes provided by the FDA and contact CooperSurgical for instructions on how to return the affected units.
The recalled product
- Product
- INCA (Infant Nasal Cannulae Assy) Infant Nasal CPAP, Complete Set. Stay-Flex tubing component. 5 units per box. Model Numbers: 44-2707, 44-2709, 44-2710, 44-2712, 44-2715.
- Manufacturer
- CooperSurgical, Inc.
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (7)
- 550033781
- 550031616
- 550031620
- 550031614
- 550033779
- 550033780
- 550034763.
Distribution
Related recalls
Same manufacturer · CooperSurgical, Inc.
See all →- SevereCooperSurgical Recalls Neonatal Nasal CPAP Replacement Sets Due to Loose Tubing
FDA (Devices) · 2026-09-30
- HighHSG procedure trays label non-sterile iodine pouches as sterile
FDA (Devices) · 2025-07-30
- HighEndosee system convenience trays label non-sterile iodine pouches as sterile
FDA (Devices) · 2025-07-30
- HighEndosee convenience kits label non-sterile iodine pouches as sterile
FDA (Devices) · 2025-07-30
- HighK-Systems G85 Mini Incubator May Exceed Desired Temperature
FDA (Devices) · 2024-12-25
Same hazard · loss of pressure
See all →- HighBoston Scientific Hurricane RX Dilation Balloon Recalled for Pressure Loss Defect
FDA (Devices) · 2021-12-08
- HighFord Recalls 2020 F-150 Trucks Due to Damaged Spare Tires
NHTSA · 2019-12-18
- SevereToyo Recalls Tires Due to Tread and Belt Separation Risk
NHTSA · 2014-11-04
- SevereFederal Couragia A/T Tires Recalled for Steel Belt Separation
NHTSA · 2014-04-18
- HighDePuy Spine Recalls Confidence Spinal Cement System 5cc Kits
FDA (Devices) · 2013-09-11
Similar recalls
- SevereCooperSurgical Recalls Neonatal Nasal CPAP Replacement Sets Due to Loose Tubing
FDA (Devices) · 2026-09-30
- CriticalFisher & Paykel Recalls Infant Nasal CPAP Prongs for Detachment Risk
FDA (Devices) · 2014-05-28
- HighIntersurgical Solus Flexible Laryngeal Mask Airways Recalled for Occlusion Risk
FDA (Devices) · 2017-12-13
- LowCooperSurgical Recalls Carter-Thomason II Port Closure System Due to Printing Defect
FDA (Devices) · 2015-05-27
- HighCooper Surgical Tampa Catheter Recalled for Potential Sterility Compromise
FDA (Devices) · 2014-06-18
- HighCooperSurgical Recalls Colpo-Pneumo Occluder Due to Compromised Sterile Pouch Seal
FDA (Devices) · 2018-10-17