Synapse PACS Software incorrect patient age calculation for infants
FUJIFILM's Synapse PACS Software versions 7.4.x incorrectly calculates patient age for infants under 3 months, potentially affecting clinical decisions. 104 units recalled in US and internationally.
What this means for you
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: FDA Class II recall involving a software error in medical imaging systems that could affect clinical decision-making for vulnerable pediatric patients under 3 months. No reported illnesses or injuries have been documented, classifying this as a risk-of-harm product without reported harm.
Plain-English summary
Synapse PACS Software, version 7.4.x, manufactured by FUJIFILM Healthcare Americas Corporation, is being recalled. The software displays incorrect computed patient age in the VX viewer for patients less than 3 months old.
This error could affect clinical decision-making and patient safety, as accurate patient age is essential for pediatric care, including age-based diagnostic protocols and medication dosing. Incorrect age data may be stored in patient medical records.
The recall affects approximately 104 units (69 in the US, 35 internationally) distributed across 30 US states and countries including Australia, Belgium, Brazil, France, Germany, Portugal, and the UK. Affected software versions are 7.4.000, 7.4.001, 7.4.010, 7.4.100, 7.4.110, and 7.4.200. Healthcare facilities should contact FUJIFILM Healthcare Americas Corporation for guidance on corrective measures.
The recalled product
- Product
- Synapse PACS Software Version 7.4.x; Software Versions: 7.4.000, 7.4.001, 7.4.010, 7.4.100, 7.4.110, 7.4.200.
- Manufacturer
- FUJIFILM Healthcare Americas Corporation
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (17)
- Model numbers: Synapse PACS 7.4.000
- 7.4.001
- 7.4.010
- 7.4.100
- 7.4.110
- 7.4.200 Software Version Numbers: Synapse PACS 7.4.000
- UDI: (01)00854904006008(10)0704000
- Synapse PACS 7.4.001
- UDI: (01)00854904006008(10)0704001
- Synapse PACS 7.4.010
- UDI: (01)00854904006008(10)0704010
- Synapse PACS 7.4.100
- UDI: (01)00854904006008(10)0704100
- Synapse PACS 7.4.110
- UDI: (01)00854904006008(10)0704110
- Synapse PACS 7.4.200
- UDI: (01)00854904006008(10)0704200.
Distribution
Related recalls
Same manufacturer · FUJIFILM Healthcare Americas Corporation
See all →- ModerateFujifilm Synapse PACS Software Recalled for Inaccurate Ruler Measurements
FDA (Devices) · 2026-09-02
- HighX-ray systems recalled over welded contactor contacts that may overheat
FDA (Devices) · 2026-02-11
- HighRadiography suites recalled because X-ray tube mounting shaft may break
FDA (Devices) · 2025-10-15
- ModerateASPIRE Cristalle mammography systems shipped with unapproved slabbing software enabled
FDA (Devices) · 2025-08-06
- HighMobile X-Ray System Exposure Settings May Not Apply Correctly