24 recalls in our database name FUJIFILM Healthcare Americas Corporation as the manufacturing or recalling firm.

What the numbers show

Total
24
Critical
0
Severe
0
Most recent
2026-09-02

Of the 24 recalls from FUJIFILM Healthcare Americas Corporation we have scored against our rubric, 0 (0%) are Critical — deaths or serious injuries reported, or the agency’s own highest classification — and 0 (0%) are Severe.

About this listing

The manufacturer string comes from the agency’s own source notice — for FDA recalls, the “recalling firm” field; for NHTSA recalls, the vehicle manufacturer; for CPSC recalls, the firm-of-record on the joint press release. A single corporate family can appear under multiple manufacturer strings as agency notices aren’t standardized on a canonical name. If you don’t see recalls you expect, try a parent-company search or a broader full-text search.

Severity scores below are computed from the agency’s classification and the recall text, never from manufacturer behaviour or history. Many manufacturers issue routine corrective actions (software updates, labelling corrections) that are correctly scored low — having recalls on this list isn’t inherently a quality signal.

1–24 of 24

  • ModerateFDA (Devices)·Z-3030-2026·2026-09-02

    Fujifilm Synapse PACS Software Recalled for Inaccurate Ruler Measurements

    Fujifilm is recalling Synapse PACS software v7.4.310 because a software issue may cause ruler measurements to be inaccurate.

    Product
    FUJIFILM Synapse PACS software v7.4.310; radiological image processing system
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1299-2026·2026-02-11

    X-ray systems recalled over welded contactor contacts that may overheat

    FUJIFILM Healthcare Americas is recalling 16 FDR Visionary Suite systems because welded contactor contacts may keep supplying current after the device is switched off.

    Product
    FDR Visionary Suite; Model Number: CH-200; Version: (1) 566-16130-23, (2) 566-16130-31, (3)
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0044-2026·2025-10-15

    Radiography suites recalled because X-ray tube mounting shaft may break

    FUJIFILM is recalling 16 FDR Visionary Suite systems because the shaft mounting the X-ray tube unit on the support may break, with protective parts possibly not working.

    Product
    FDR Visionary Suite, CH-200 Model/Version Numbers: (1) 566-16130-23, (2) 566-16130-31, (3) 566-16130-33
    Category
    Medical Device
    Distribution
    8 states
  • ModerateFDA (Devices)·Z-2219-2025·2025-08-06

    ASPIRE Cristalle mammography systems shipped with unapproved slabbing software enabled

    ASPIRE Cristalle Mammography Systems with the Digital Breast Tomosynthesis option are being recalled because devices had an unapproved slabbing software function enabled for use. 293 units affected.

    Product
    ASPRIRE Cristalle Mammography System (cleared under K212873) installed with the ASPIRE Cristalle Digital Breast Tomosynthesis (DBT) Option (approved under P160031).
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1407-2025·2025-03-26

    Synapse PACS Software incorrect patient age calculation for infants

    FUJIFILM's Synapse PACS Software versions 7.4.x incorrectly calculates patient age for infants under 3 months, potentially affecting clinical decisions. 104 units recalled in US and internationally.

    Product
    Synapse PACS Software Version 7.4.x; Software Versions: 7.4.000, 7.4.001, 7.4.010, 7.4.100, 7.4.110, 7.4.200.
    Category
    Medical Device
    Distribution
    30 states
  • HighFDA (Devices)·Z-3158-2024·2024-10-02

    Mobile X-Ray System Exposure Settings May Not Apply Correctly

    Fujifilm mobile x-ray systems may fail to apply configured exposure settings or default to adult chest imaging parameters. This could result in inappropriate radiation exposure levels.

    Product
    The device is a mobile x-ray system designed to work with Fujifilm's GOS and CsI scintillator FDR D-EVO2 (DR-ID) 12XXSE) family of digital x-ray detectors coupled.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1898-2024·2024-06-05

    Radiographic imaging system tubes may fall due to bolt rupture

    FUJIFILM's FDR Visionary Suite radiographic imaging system has bolts that may rupture due to fatigue, potentially causing the tube support to fall and injure patients. The recall affects 18 units distributed across nine states.

    Product
    FDR Visionary Suite - Intended to generate digital or conventional radiographic images of the skull, spinal column, chest, abdomen, extremities, and other body parts of human anatomies in all routine radiography examinations. Model/Catalog Number: 566-16130-23…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1884-2024·2024-05-29

    Noblus AC Adapter for Ultrasound System Recalled for Smoke Hazard

    FUJIFILM is recalling Noblus AC Adapters nationwide due to fire and smoke hazard. The adapters can overheat and generate smoke when the power cable is repeatedly inserted and removed.

    Product
    Noblus AC Adapter used with the Noblus Ultrasound Imaging System Model : AHM250PS24
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1271-2024·2024-03-13

    FUJIFILM Synapse PACS Software Produces Incorrect Breast Imaging Measurements

    FUJIFILM is recalling Synapse PACS Version 7.2.100 because the medical imaging software produces incorrect measurements when analyzing certain breast images. Affected healthcare facilities using this software version should verify measurement accuracy in their systems.

    Product
    Synapse PACS - Version 7.2.100
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1272-2024·2024-03-13

    Medical imaging software measurements incorrect in specific cases

    FUJIFILM's Synapse PACS Version 7.2.200 produces incorrect measurements when analyzing certain breast imaging scans without pixel spacing data. This measurement error could affect diagnostic accuracy.

    Product
    Synapse PACS - Version 7.2.200
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1270-2024·2024-03-13

    Synapse PACS Software Version 7.2.000 Measurement Error in Breast Imaging

    FUJIFILM's Synapse PACS Version 7.2.000 produces incorrect measurements in breast imaging when analyzing Secondary Capture 2D images lacking pixel spacing information. Affected healthcare facilities should discontinue use for breast imaging measurements.

    Product
    Synapse PACS - Version 7.2.000
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1269-2024·2024-03-13

    Medical Device Imaging Software May Calculate Incorrect Breast Measurements

    Synapse PACS Version 7.1.000US may produce incorrect measurements when Secondary Capture 2D images lack pixel spacing information in the DICOM header. The affected software is installed in healthcare facilities worldwide, including the United States and Puerto Rico.

    Product
    Synapse PACS - Version 7.1.000US
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1273-2024·2024-03-13

    Synapse PACS version 7.3.000 produces incorrect breast imaging measurements

    FUJIFILM's Synapse PACS version 7.3.000 software produces incorrect measurements when secondary capture images without pixel spacing are combined with breast tomosynthesis series. No illnesses or injuries reported.

    Product
    Synapse PACS - Version 7.3.000
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1171-2024·2024-02-28

    Cardiology imaging software may calculate heart chamber size incorrectly

    Synapse Cardiology PACS software versions 7.0–7.3.0 may calculate left ventricle mass incorrectly, potentially leading to misdiagnosis or wrong treatment plans. Thirteen units have been distributed in the US and Israel.

    Product
    Synapse Cardiology PACS V7.3.0, V7.2, V7.1, V7.0. A web-based application as the primary user interface for the processing of medical images.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1172-2024·2024-02-28

    Synapse CV 6 Imaging Software Calculation Error May Cause Misdiagnosis

    FUJIFILM Synapse CV 6 medical imaging software may incorrectly calculate left ventricle mass measurements. If used for diagnosis, this calculation error could lead to patient misdiagnosis and incorrect treatment, potentially causing long-term health consequences.

    Product
    Synapse CV 6. with AR. A web-based application as the primary user interface for the processing of medical images.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1470-2023·2023-05-03

    Digital X-ray System Arm May Break, Posing Injury Risk

    FUJIFILM's FDR AQRO digital X-ray system arm may break during use. The tube head support is susceptible to failure, which could cause serious harm to patients or operators.

    Product
    FujiFilm FDR AQRO (Model DR-XD1000)- A digital mobile X-ray system intended for use in general purpose radiography for generating radiographic images of human anatomy, including adult, pediatric, and neonatal exams
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1748-2022·2022-09-21

    SCENARIA View software error may require additional patient radiation exposure

    SCENARIA View imaging software errors prevent image reconstruction, which may require additional scans and expose patients to unnecessary radiation.

    Product
    SCENARIA View: Software Version V1.08, V1.09B, V1.09C, V1.09D, V1.0B, V1.0C
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1749-2022·2022-09-21

    FUJIFILM Surpria Imaging Software May Fail During Scanning

    Surpria software versions V3.11 and V3.22 may encounter processing errors that prevent image reconstruction, potentially requiring additional scans and exposing patients to unnecessary radiation.

    Product
    Surpria: Software Version V3.11, V3.22
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1750-2022·2022-09-21

    Surpria 64 Software Error May Prevent Image Reconstruction During Scanning

    Surpria 64 CT scanner software versions V3.11 and V3.22 may encounter errors that prevent image reconstruction, requiring repeat scans that expose patients to additional radiation. 16 units nationwide are affected.

    Product
    Surpria 64: Software Version V3.11, V3.22
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0191-2022·2021-11-10

    Lisendo 880 Ultrasound Software displays incorrect cardiac valve measurements

    Lisendo 880 Ultrasound Software versions V1.0–V4.1.3 may not display cardiac measurement results correctly for mitral valve assessments, affecting diagnostic accuracy.

    Product
    Lisendo 880 Ultrasound Software Version: V1.0 through V4.1.3
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0190-2022·2021-11-10

    Arietta 850 Ultrasound Software Displays Cardiology Measurements Incorrectly

    FujiFilm Arietta 850 ultrasound systems with affected software versions display cardiology measurement results incorrectly. The recall affects 38 units distributed nationwide.

    Product
    Arietta 850 Ultrasound- Software Version: V1.0.0 through V4.1.3
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0189-2022·2021-11-10

    Arietta 750 Ultrasound Software Versions Affected by Incorrect Measurement Display

    FujiFilm Healthcare is recalling Arietta 750 Ultrasound Software versions V1.0.0 through V2.1.3 due to measurement display errors in critical cardiology functions that could lead to incorrect clinical diagnoses. The instruction manual also contains an error.

    Product
    Arietta 750 Ultrasound Software Version: V1.0.0 through V2.1.3
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0192-2022·2021-11-10

    Arietta 65 Ultrasound Software: Calculation Error in EROA Diagnostic Measurements

    Arietta 65 Ultrasound systems with software versions 1.0 through 4.0.0 contain a calculation error that produces incorrect measurements of cardiac valve function during diagnostic assessments.

    Product
    Arietta 65 Ultrasound Software Versions: V1.0 through V4.0.0
    Category
    Medical Device
    Distribution
    Distributed nationwide