Spectral CT scanners recalled over rotor fasteners that may be undertorqued
Philips is recalling 283 Spectral CT units because fasteners attaching the rotor to the gantry bearing may not be torqued to specification.
What this means for you
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: Scored 3 under the risk-of-harm criterion: the firm states unsecured rotor parts may potentially be expelled during gantry rotation, and no such event or injury has been reported.
Plain-English summary
The Philips Spectral CT, product code (REF) 728333, is being recalled by Philips North America Llc.
The record states that as part of the CT system sub assembly, 32 fasteners are used to attach the rotor to the bearing within the gantry, and that one or more of these fasteners may not be torqued to specification. If multiple fasteners are not torqued to specification, the rotor or rotor parts may become unsecured or displaced. The record states that while there has been no reported or observed event of expelled parts during gantry rotation, unsecured rotor parts may potentially be expelled.
The record lists 283 units under UDI-DI 00884838101111, with serial numbers recorded in the enforcement record. Distribution was worldwide, including nationwide in the United States across the states listed in the record. The FDA has classified the action as Class II.
The enforcement record reports no injuries. Facilities should follow Philips' instructions.
The recalled product
- Product
- Spectral CT; Product Code (REF): 728333;
- Manufacturer
- Philips North America Llc
- Category
- Medical Device — CT scanner
- Affected units
- 283
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (20)
- Product Code (REF): 728333
- UDI-DI: 00884838101111
- Serial Numbers: 10055
- 10075
- 10086
- 10096
- 10106
- 10116
- 10126
- 10136
- 10146
- 10156
- 10166
- 10176
- 10186
- 10196
- 10206
- 10217
- 10227
- 10237
Distribution
Part of a larger recall action
This product is one of 4 recalled by Philips North America Llc under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 4 products in this recall →Related recalls
Same manufacturer · Philips North America Llc
See all →- HighPhilips Recalls IQon Spectral CT Systems Due to Radiation Issue
FDA (Devices) · 2026-09-09
- HighPhilips CT Systems Recalled for Unintended Radiation Performance Issues
FDA (Devices) · 2026-09-09
- HighPhilips Brilliance CT Systems Recalled for Unintended Radiation Issue
FDA (Devices) · 2026-09-09
- HighPhilips CT Systems Recalled for Unintended Radiation Performance Issues
FDA (Devices) · 2026-09-09
- HighPhilips Recalls Cardiac Workstation 5000 Units for Inaccurate Heart Rate Data
FDA (Devices) · 2026-08-26
Same hazard · fastener torque
See all →- HighSpectral CT 7500 scanners recalled over rotor fasteners that may loosen
FDA (Devices) · 2026-02-18
- HighBrilliance iCT scanners recalled over rotor fasteners that may be undertorqued
FDA (Devices) · 2026-02-18
- HighIQon Spectral CT recalled over rotor fasteners that may be undertorqued
FDA (Devices) · 2026-02-18
- Severe
- Severe