GE Healthcare Recalls Infinia Nuclear Medicine Systems for Detector Motion Risk
GE Healthcare is recalling Infinia and Infinia Hawkeye Nuclear Medicine Systems due to a potential safety issue where the detector may drift toward the patient.
What this means for you
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall is classified as FDA Class II. The hazard involves a mechanical defect (detector drifting toward the patient) that poses a risk of harm, but the source text does not report any injuries or illnesses, fitting the criteria for a High severity score (risk-of-harm products where injury has not yet been reported).
Plain-English summary
GE Healthcare, LLC is recalling GE Healthcare Infinia and Infinia Hawkeye Nuclear Medicine Systems, including the Hawkeye Option for Dual-Head Variable Angle Gamma Camera and GE Quasar Nuclear Medicine Systems. The recall involves 3,062 units distributed worldwide, including 1,182 units in the United States and 1,880 units outside the US.
The recall is issued due to a potential safety issue related to unintended radial detector motion. This failure mode may occur during patient setup or at the end of a patient scan. The issue has been observed on a single system where the detector moved to the home position and then slowly started drifting inward toward the patient. This movement generated both visual and audio alert conditions on both the console and the gantry.
Affected devices are used for general Nuclear Medicine imaging procedures to detect radioisotope tracer uptake in the patient body. The systems are used in Oncology, Cardiology, Neurology, and other clinical diagnostic imaging applications. Users should check their device serial numbers against the list provided in the official recall notice to determine if their unit is affected.
The recalled product
- Product
- GE Healthcare Infinia and Infinia Hawkeye Nuclear Medicine Systems. For Dual-Head Variable Angle Gamma Camera, Hawkeye Option for Dual-Head Variable Angle Gamma Camera (this includes GE Quasar Nuclear Medicine Systems). Nuclear Medicine Systems Infinia devices Product Usage: T
- Manufacturer
- GE Healthcare, LLC
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (20)
- Serial Numbers for all devices: 22
- 343
- 1002
- 1003
- 1006
- 1007
- 1008
- 1010
- 1012
- 1013
- 1016
- 1018
- 1019
- 1021
- 1025
- 1026
- 1027
- 1034
- 1035
- 1038
Distribution
Part of a larger recall action
This product is one of 3 recalled by GE Healthcare, LLC under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 3 products in this recall →Related recalls
Same manufacturer · GE Healthcare, LLC
See all →- HighGE SIGNA Premier MRI magnets may have ferrous fittings fitted
FDA (Devices) · 2026-04-08
- HighGE Medical Imaging Software: Inaccurate Measurements in Global Stack Viewport
FDA (Devices) · 2022-08-31
- HighGE Healthcare CT, PET, and NM Patient Tables May Drop During Service
FDA (Devices) · 2022-08-17
- HighGE Centricity PACS Software fails to notify of exam study modifications
FDA (Devices) · 2022-08-03
- HighSIGNA Premier MRI scanner recalled for elevated acoustic noise hazard
FDA (Devices) · 2022-07-27
Same hazard · mechanical failure
See all →- HighIMRIS ORT300 Operating Room Table Recalled for Mechanical Failure Risk
FDA (Devices) · 2026-09-02
- HighIMRIS ORT200 Operating Room Table Recalled for Mechanical Failure Risk
FDA (Devices) · 2026-09-02
- HighForest River Transit Buses Recalled for Wheelchair Lift Control Failure
NHTSA · 2026-09-01
- HighBlue Bird Recalls Vision School Buses Due to Wheelchair Lift Defect
NHTSA · 2026-08-31
- SevereMagic Tilt Watercraft Trailers Recalled Due to Winch Weld Failure
NHTSA · 2026-08-28