Ortho-Clinical Recalls Deformed VITROS Chemistry VersaTips
Ortho-Clinical Diagnostics is recalling deformed VITROS Chemistry VersaTips that may cause assay results to be biased by up to +/-16%.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall involves a risk-of-harm product (medical device) where the defect leads to inaccurate test results, but no specific injuries or illnesses have been reported in the source text.
Plain-English summary
Ortho-Clinical Diagnostics, Inc. is recalling 4,062 cartons of VITROS Chemistry Products VersaTips (Lot 413420). The recall involves a total of 4,062,000 tips. These products are used to dispense samples and reagent fluids for various VITROS chemistry assays.
The recall is due to a deformity affecting the specific lot. The use of deformed VersaTips may result in assay results that are biased by up to +/-16% for both immunometric and competitive assays. This inaccuracy could affect the reliability of laboratory test results.
The affected products were distributed worldwide, including nationwide in the United States and in Canada. The tips are compatible with VITROS 250, 350, ECi, 5,1 FS, 3600, 4600, 5600, and XT7600 models. Users should check their inventory for Lot 413420 and stop using the affected tips.
The recalled product
- Product
- VITROS Chemistry Products VersaTips - Product Usage: used to dispense samples to MicroSlide assays and sample and reagent fluid for MicroWell assays. In addition, VersaTips are used to dispense R1 Reagent Fluid into cuvettes in combination with sample fluid for Micro Tips…
- Manufacturer
- Ortho-Clinical Diagnostics, Inc.
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (9)
- Lot 413420 Model(s): VITROS 250
- VITROS 350
- VITROS ECi
- VITROS 5
- 1 FS
- VITROS 3600
- VITROS 4600
- VITROS 5600
- VITROS XT7600.
Distribution
Related recalls
Same manufacturer · Ortho-Clinical Diagnostics, Inc.
See all →- ModerateLactate test slides recalled over calibration failures delaying patient results
FDA (Devices) · 2025-09-24
- ModerateLactate slides in 18-slide cartridges recalled over calibration failures delaying results
FDA (Devices) · 2025-09-24
- ModerateCalcium test slides recalled after calibration failures delayed patient results
FDA (Devices) · 2025-08-27
- HighVITROS CHE slides can cause quality control drift with performance verifiers
FDA (Devices) · 2025-07-23
- HighVITROS Performance Verifier II lots can cause quality control drift
FDA (Devices) · 2025-07-23
Same hazard · inaccurate results
See all →- HighHealthmark Protein Test Recalled Due to Expired Ingredient
FDA (Devices) · 2026-09-02
- HighGEM Premier 7000 with iQM3 Blood Analysis Systems Recalled for Solution Concentration Issues
FDA (Devices) · 2026-08-12
- HighHardyDisk AST Cefiderocol 30µg for In Vitro Testing Recalled Due to Incorrect Cartridges
FDA (Devices) · 2026-08-12
- HighAbbott i-STAT EG7+ Blood Gas Cartridges Reporting Inaccurate Results
FDA (Devices) · 2026-05-06
- HighAbbott i-STAT Blood Gas Cartridges Reporting Inaccurate pH and PCO2 Results
FDA (Devices) · 2026-05-06
Similar recalls
- HighLaboratory diagnostic tips recalled for potential dispensing errors
FDA (Devices) · 2022-03-16
- HighOrtho-Clinical Diagnostics Recalls VITROS VALP Reagent Packs Due to Negative Bias
FDA (Devices) · 2019-02-27
- ModerateOrtho-Clinical VITROS 5,1 FS Chemistry System Recalled for Biased Results
FDA (Devices) · 2014-06-18
- ModerateOrtho-Clinical Diagnostics Recalls VITROS Chemistry Products PHYT Slides
FDA (Devices) · 2018-01-03
- SevereOrtho-Clinical Diagnostics Recalls VITROS TRIG Slides Due to Biased Results
FDA (Devices) · 2014-12-31
- ModerateOrtho-Clinical Diagnostics Recalls VITROS Chemistry Products Calibrator Kit 17
FDA (Devices) · 2014-06-25