Calcium test slides recalled after calibration failures delayed patient results
VITROS Chemistry Products Ca Slides, catalog number 145 0261, generations 67 and above, are recalled because calibration failures on VITROS XT systems delay calcium test results. 174,215 units are affected.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 2/5 (Moderate).
Minor labeling or packaging issue, or a voluntary precautionary recall. Risk of harm is low but real — follow the manufacturer's instructions.
Our severity reasoning: The consequence stated is a delay in obtaining calcium results rather than erroneous results, which places this at rubric level 2.
Plain-English summary
VITROS Chemistry Products Ca Slides, catalog number 145 0261, UDI-DI 10758750009114, are being recalled for generations 67 and above. The recall covers 174,215 units distributed in the United States and in Australia, Belgium, Bermuda, Brazil, Canada, Chile and other countries.
Affected generations are experiencing an increased occurrence of Condition Code TH4-63J during calibration on VITROS XT 3400 and VITROS XT 7600 Systems, causing calibration failures that delay calcium test results and potentially impact patient management. The imaging reflectometer algorithm detects excessive spikes, or imperfections, in slide images and generates the condition code. The FDA classified the action as Class II.
Laboratories should check the generation numbers of their slide stock against the recall notice and follow the manufacturer's instructions.
The recalled product
- Product
- VITROS Chemistry Products Ca Slides; Catalog Number: 145 0261;
- Manufacturer
- Ortho-Clinical Diagnostics, Inc.
- Affected units
- 174,215
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (3)
- Catalog Number: 145 0261
- UDI-DI: 10758750009114
- Affected GENs: 67 and above
Distribution
Distributed nationwide across the United States.
Related recalls
Same manufacturer · Ortho-Clinical Diagnostics, Inc.
See all →- ModerateLactate test slides recalled over calibration failures delaying patient results
FDA (Devices) · 2025-09-24
- ModerateLactate slides in 18-slide cartridges recalled over calibration failures delaying results
FDA (Devices) · 2025-09-24
- HighVITROS CHE slides can cause quality control drift with performance verifiers
FDA (Devices) · 2025-07-23
- HighVITROS Performance Verifier II lots can cause quality control drift
FDA (Devices) · 2025-07-23
- HighVITROS CHE slides sold in Japan can cause quality control drift
FDA (Devices) · 2025-07-23
Same hazard · calibration failure
See all →- ModeratePhosphorus reagent lots recalled because calibration may fail before expiration date
FDA (Devices) · 2025-10-15
- ModerateBeckman Coulter DxC 500 AU Chemistry Analyzer Software Calibration Error
FDA (Devices) · 2025-05-21
- ModerateBeckman Coulter Access PCT Reagent Pack Recalled for Calibration Failures
FDA (Devices) · 2025-02-12
- ModerateDRG:HYBRiD-XL Insulin Device Recalled for Calibration Failures
FDA (Devices) · 2025-01-22
- HighMedical Device Homocysteine Test Assay Recalled for Calibration Defects
FDA (Devices) · 2024-11-27
Similar recalls
- ModerateLactate test slides recalled over calibration failures delaying patient results
FDA (Devices) · 2025-09-24
- ModerateLactate slides in 18-slide cartridges recalled over calibration failures delaying results
FDA (Devices) · 2025-09-24
- HighOrtho-Clinical VITROS Calcium Slides Recalled for Biased Results
FDA (Devices) · 2016-08-03
- ModerateOrtho-Clinical Diagnostics Recalls VITROS Chemistry Ca Slides Due to Dispense Issues
FDA (Devices) · 2013-11-27
- ModerateDiagnostic Calibrator Kit Recalled for Software Defect Affecting System Calibration
FDA (Devices) · 2022-11-16
- ModerateOrtho-Clinical VITROS Chemistry Calibration Data Module Recalled for Calibration Failure
FDA (Devices) · 2014-04-30