The Recall Desk

Manufacturer

Ortho-Clinical Diagnostics, INc.

90 recalls in our database name Ortho-Clinical Diagnostics, INc. as the manufacturing or recalling firm.

What the numbers show

Total
90
Critical
0
Severe
1
Most recent
2025-09-24

Of the 90 recalls from Ortho-Clinical Diagnostics, INc. we have scored against our rubric, 0 (0%) are Critical — deaths or serious injuries reported, or the agency’s own highest classification — and 1 (1%) are Severe.

About this listing

The manufacturer string comes from the agency’s own source notice — for FDA recalls, the “recalling firm” field; for NHTSA recalls, the vehicle manufacturer; for CPSC recalls, the firm-of-record on the joint press release. A single corporate family can appear under multiple manufacturer strings as agency notices aren’t standardized on a canonical name. If you don’t see recalls you expect, try a parent-company search or a broader full-text search.

Severity scores below are computed from the agency’s classification and the recall text, never from manufacturer behaviour or history. Many manufacturers issue routine corrective actions (software updates, labelling corrections) that are correctly scored low — having recalls on this list isn’t inherently a quality signal.

1–25 of 90

  • ModerateFDA (Devices)·Z-2559-2025·2025-09-24

    Lactate test slides recalled over calibration failures delaying patient results

    VITROS Chemistry Products LAC Slides from coating 0130 and above are recalled because of increased calibration failures on VITROS XT systems, delaying lactate results. 2,908 units are affected.

    Product
    Chemistry Products LAC Slides. Catalog Number: 8433880. 1 Sales Unit = contains 5 cartridges of 60 slides, for a total of 300 tests.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-2560-2025·2025-09-24

    Lactate slides in 18-slide cartridges recalled over calibration failures delaying results

    VITROS Chemistry Products LAC Slides, catalog 8150112, from coating 0130 and above are recalled because of increased calibration failures on VITROS XT systems, delaying lactate results. 2,320 units are affected.

    Product
    VITROS Chemistry Products LAC Slides. Catalog Number: 8150112. 1 Sales Unit = contains 5 cartridges of 18 slides, for a total of 90 tests.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-2335-2025·2025-08-27

    Calcium test slides recalled after calibration failures delayed patient results

    VITROS Chemistry Products Ca Slides, catalog number 145 0261, generations 67 and above, are recalled because calibration failures on VITROS XT systems delay calcium test results. 174,215 units are affected.

    Product
    VITROS Chemistry Products Ca Slides; Catalog Number: 145 0261;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2121-2025·2025-07-23

    VITROS Performance Verifier II lots can cause quality control drift

    VITROS Chemistry Products Performance Verifier II is being recalled because quality control drift can occur with certain cholinesterase slide lots, which could produce erroneous patient results.

    Product
    Brand Name: VITROS Chemistry Products Product Name: VITROS Chemistry Products Performance Verifier II Model/Catalog Number: 8231474 Software Version: N/A Product Description: VITROS Chemistry Products Performance Verifier is an assayed control used to monitor performance on…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2120-2025·2025-07-23

    VITROS Performance Verifier I lots can cause quality control drift

    VITROS Chemistry Products Performance Verifier I is being recalled because quality control drift can occur with certain cholinesterase slide lots, which could produce erroneous patient results.

    Product
    Brand Name: VITROS Chemistry Products Product Name: VITROS Chemistry Products Performance Verifier I Model/Catalog Number: 806 7324 Software Version: Not Applicable Product Description: VITROS Chemistry Products Performance Verifier is an assayed control used to monitor…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2123-2025·2025-07-23

    VITROS CHE slides sold in Japan can cause quality control drift

    VITROS Chemistry Products CHE Slides supplied in Japan are being recalled because quality control drift can occur with certain Performance Verifier lots, which could produce erroneous patient results.

    Product
    VITROS Chemistry Products CHE Slides (Japan); Catalog Numbers: (1) 684 4074 (90 slides), (2) 684 4138 (300 slides);
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2122-2025·2025-07-23

    VITROS CHE slides can cause quality control drift with performance verifiers

    VITROS Chemistry Products CHE Slides are being recalled because quality control drift can occur with certain Performance Verifier lots, which could produce erroneous patient results.

    Product
    VITROS Chemistry Products CHE Slides; Catalog Numbers: (1) 191 4605 (300 slides), (2) 800 4707 (90 slides);
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2039-2025·2025-07-09

    VITROS XT 7600 Chemistry Analyzer Software Anomaly Causing Erroneous Test Results

    Ortho-Clinical Diagnostics is recalling VITROS XT 7600 Integrated System analyzers due to a software anomaly that can report test results as expired when reagent shelf expiration dates are omitted during manual loading, potentially causing erroneous or delayed test results.

    Product
    Brand Name: VITROS; Product Name: VITROS XT 7600 Integrated System; Model/Catalog Number: (1) 6844461, (2) 6272222; Software Version: version 3.2 to version 3.8.3 (all available software versions); Product Description: In vitro diagnostic chemistry analyzer.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2040-2025·2025-07-09

    VITROS 4600 chemistry systems may report results using expired diluent packs

    VITROS 4600 Chemistry Systems are being recalled because a software anomaly can allow results to be reported using diluent packs the system treats as expired, which can produce erroneous results.

    Product
    Brand Name: VITROS Product Name: VITROS 4600 Chemistry System Model/Catalog Number: (1) 6802445, (2) 6900440 Software Version: version 3.2 to version 3.8.3 (all available software versions) Product Description: In vitro diagnostic chemistry analyzer.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-2073-2025·2025-07-09

    VITROS opiate reagent lots may give low quality control results

    VITROS Chemistry Products OP Reagent Gen 15 lots are being recalled because they may generate lower than expected quality control results. 2,017 units are affected.

    Product
    VITROS Chemistry Products OP Reagent Gen 15. Model Number: 680 1997. The VITROS OP Reagent is a dual-chambered package containing ready-to-use liquid reagents that are used to detect opiates in urine. One sales unit contains 6 Microtip packs of reagent. Each pack has 50 OP…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-2038-2025·2025-07-09

    VITROS 5600 Integrated System software anomaly reported

    Ortho-Clinical Diagnostics is recalling the VITROS 5600 Integrated System due to a software anomaly that can report test results using expired reagents even when the use-expired-reagents setting is disabled, potentially causing erroneous or delayed test results.

    Product
    Brand Name: VITROS Product Name: VITROS 5600 Integrated System Model/Catalog Number: (1) 6802413, (2) 6802915
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1948-2025·2025-06-18

    Solana Bordetella Complete Assay Kit Recalled for False Positive Results

    Ortho-Clinical Diagnostics is recalling the Solana Bordetella Complete Assay Kit (Model M308) because it may produce false positive results for Bordetella parapertussis at a rate of up to 16.7%.

    Product
    Brand Name: Solana¿ Bordetella Complete Assay Product Name: Solana Bordetella kit Model/Catalog Number: M308 Software Version: not applicable Product Description: One sales unit (kit) contains materials for 48 tests. One kit contains 48 tubes of 1.45 mL Process Buffer and…
    Category
    Medical Device
    Distribution
    0 states
  • ModerateFDA (Devices)·Z-1876-2025·2025-06-04

    VITROS Chemistry Products CRP Slides Incorrectly Labeled Coating Generation

    Ortho-Clinical Diagnostics is recalling VITROS Chemistry Products CRP Slides because the product carton, foil wrapper, and cartridge label were incorrectly marked with "GEN 88" instead of the correct "GEN 83" designation.

    Product
    VITROS Chemistry Products CRP Slides. One sales unit contains 5 cartridges, of 18 slides per cartridge, for a total of 90 slides.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1007-2025·2025-01-29

    VITROS 3600 Diagnostic System May Fail to Alert Users to Expired Reagents

    The VITROS 3600 Immunodiagnostic System may not properly alert users when Signal Reagent Packs expire, risking use of expired products and potentially inaccurate test results. Affected systems worldwide should be checked for this software issue.

    Product
    Brand Name: VITROS Product Name: VITROS 3600 Immunodiagnostic System Model/Catalog Number: 6802783 (New), 6802914 (Certified/Refurbished), 6904000 (China) Software Version: version 3.8.0 or higher Product Description: In vitro diagnostic chemistry analyzer. FDA 3 letter…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1006-2025·2025-01-29

    Diagnostic analyzer software may fail to alert users to expired reagents

    VITROS XT 7600 Integrated System software may not warn users when Signal Reagent Packs expire, risking inaccurate or missing test results.

    Product
    Brand Name: VITROS Product Name: VITROS XT 7600 Integrated System Model/Catalog Number: 6844461 (New), 6272222 (Certified), 6904003 (China) Software Version: version 3.8.0 or higher Product Description: In vitro diagnostic chemistry analyzer. FDA 3 letter product code…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1005-2025·2025-01-29

    Medical Analyzer Software May Fail to Alert of Expired Reagent Packs

    The VITROS 5600 Integrated System's software (version 3.8.0 and higher) may fail to notify users when Signal Reagent Packs expire. This could result in use of expired reagents, causing no results or erroneous results from tests.

    Product
    Brand Name: VITROS Product Name: VITROS 5600 Integrated System Model/Catalog Number: 6802413 (New), 6802415 (Certified/Refurbished), 6904001 (China) Software Version: version 3.8.0 or higher Product Description: In vitro diagnostic chemistry analyzer. FDA 3 letter product…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0535-2025·2024-12-04

    VITROS Chemistry CRP Diagnostic Slides May Underreport Elevated Results

    Ortho-Clinical Diagnostics is recalling 7,717 units of VITROS CRP diagnostic slides that may produce falsely low results for high C-reactive protein samples, risking missed diagnosis of infection or inflammation.

    Product
    VITROS Chemistry Products CRP Slides- In vitro diagnostic quantitatively measures C-reactive protein (CRP) concentration in serum and plasma Model/Catalog Number: 1926740
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2684-2024·2024-08-28

    VITROS XT Chemistry Slides Recalled for Dust Contamination of Diagnostic Results

    Ortho-Clinical Diagnostics is recalling 88,253 units of VITROS XT Chemistry ALB-TP Slides because they shed dust and debris during normal operation, potentially contaminating sodium test results and causing misdiagnosis.

    Product
    The VITROS XT Chemistry Products ALB-TP Slides is a single device that contains both an albumin test and a total protein test side by side separated by a plastic barrier sealed within a single slide frame. In this format, individual reactions occur, and test results are…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-2580-2024·2024-08-21

    VITROS Folate Reagent Pack Recalled for Delayed Test Results

    Ortho-Clinical Diagnostics is recalling VITROS Folate Reagent Pack due to fibrinogen precipitation causing delayed test results. The recall affects 18,467 units distributed worldwide.

    Product
    VITROS Immunodiagnostic VITROS Folate Reagent Pack 1/2
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2266-2024·2024-07-10

    VITROS Chemistry FS Diluent Pack 3 Recalled for Negatively Biased Test Results

    Ortho-Clinical Diagnostics is recalling VITROS Chemistry FS Diluent Pack 3, Lot 01-1645, because it may produce negatively biased results when used with CRP slides, potentially affecting patient diagnosis and care.

    Product
    VITROS Chemistry Products FS Diluent Pack 3, Model/Catalog Number: 6801754. Product Description: VITROS Chemistry Products FS Diluent Pack 3 comes with 3 packs per sales unit.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-2245-2024·2024-07-10

    Medical Device Calibrator Kit Assigned Extended Incorrect Expiration Date

    Ortho-Clinical Diagnostics is recalling a calibrator kit with an incorrect expiration date that could bias lab results if used past its true expiration date.

    Product
    VITROS Chemistry Products Calibrator Kit 29- In-Vitor used to calibrate the VITROS 5,1 FS/4600 Chemistry Systems and the VITROS 5600/XT 7600 Integrated Systems for the quantitative measurement of total iron-binding capacity (TIBC) using VITROS Chemistry Products dTIBC Reagent…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2142-2024·2024-06-19

    VITROS Immunodiagnostic Calibrators Recalled for Potential Accuracy Bias

    Ortho-Clinical Diagnostics is recalling VITROS Anti HBs Calibrators due to potential for positively biased results caused by signal reduction over shelf life. Approximately 481 units were distributed worldwide.

    Product
    VITROS Immunodiagnostic Products Anti HBs Calibrators. The VITROS Anti-HBs Calibrator Pack contains 3 X labelled Calibrator tube (1 per level: C1 - green, C2 - yellow, C3 - red).
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-2019-2024·2024-06-12

    Quality Control Tool Vitamin B12 Range Verifier Recalled by FDA

    Ortho-Clinical Diagnostics is recalling VITROS Vitamin B12 Range Verifier Lot 1200 because the product may fail to verify required calibration range. The failure is detectable during use and does not affect actual patient test results.

    Product
    VITROS Immunodiagnostic Products Vitamin B12 Range Verifier - In vitro diagnostic use ly. For use in verifying the calibration range of the VITROS ECi/ECiQ/3600 Immunodiagnostic Systems and the VITROS 5600/XT 7600 Integrated Systems when used for the quantitative measurement of…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1225-2024·2024-03-13

    Laboratory analyzer software defect may cause erroneous patient test results

    A software defect in VITROS 4600 Chemistry Systems causes quality control baselines to not update when users change them, potentially causing erroneous patient results.

    Product
    VITROS 4600 Chemistry System Catalog Nos. 6802445 (regular) and 6900440 (refurbished)
    Category
    Medical Device
    Distribution
    Distributed nationwide

About this company

Community-sourced company facts from Wikidata, not from any recall agency. This is background on the corporate entity — it says nothing about the recall(s) listed on this page, which come from the agency source notice above.

Ortho-Clinical Diagnostics

in vitro diagnostics company for blood testing

Country
United States
Founded
1994

Official website

Source: Wikidata (Q7104389), licensed CC0. Wikidata is community-edited and may not reflect the current corporate structure.