Ortho-Clinical Diagnostics Recalls VITROS 5,1 FS Chemistry Systems
Ortho-Clinical Diagnostics is recalling VITROS 5,1 FS Chemistry Systems because extension springs may be out of specification, potentially causing imprecise test results.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 2/5 (Moderate).
Minor labeling or packaging issue, or a voluntary precautionary recall. Risk of harm is low but real — follow the manufacturer's instructions.
Our severity reasoning: The recall is classified as FDA Class II, but the source explicitly states that the probability of the defect adversely affecting results is unlikely and no injuries or illnesses are reported, fitting the criteria for a moderate severity precautionary recall.
Plain-English summary
Ortho-Clinical Diagnostics, Inc. (OCD) is recalling VITROS 5,1 FS Chemistry Systems and VITROS 5,1 FS Chemistry System Refurbished units. The recall involves specific serial numbers for both product codes 6801375 and 6801890. These devices are intended for use in the measurement of a variety of chemistry analytes.
The recall was initiated after an internal evaluation at OCD's manufacturing facility discovered that extension springs on some systems may be out of manufacturing specifications. Specifically, an extension spring was found to have insufficient force. If the spring does not provide sufficient force, MicroSlides may not be properly positioned within the MicroSlide incubator, which has the potential to cause imprecise results. OCD's evaluation determined that the probability of an out-of-specification extension spring alone adversely affecting results is unlikely.
The affected units were distributed worldwide, including nationwide in the United States (including Puerto Rico) and in countries such as Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, the Republic of Panama, Singapore, Spain, and Venezuela. Users should check their device serial numbers against the list provided in the official recall notice to determine if their unit is affected.
The recalled product
- Product
- VITROS 5,1 FS Chemistry System, Product Code 6801375 and VITROS 5,1 FS Chemistry System Refurbished, Product Code 6801890. Intended for use in the measurement of a variety of chemistry analytes.
- Manufacturer
- Ortho-Clinical Diagnostics
- Hazard
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (20)
- VITROS 5
- 1 FS Chemistry System Serial Numbers: 34000119 through 34002273
- 1 FS Chemistry System Refurbished Serial Numbers: 34000133
- 34001240
- 34000193
- 34001243
- 34000212
- 34001248
- 34000316
- 34001251
- 34000318
- 34001274
- 34000337
- 34001297
- 34000360
- 34001311
- 34000420
- 34001317
- 34000451
- 34001345
Distribution
Part of a larger recall action
This product is one of 3 recalled by Ortho-Clinical Diagnostics under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 3 products in this recall →Related recalls
Same manufacturer · Ortho-Clinical Diagnostics
See all →- HighClinical Laboratory Quality Control Device Recalled for Diluent Assembly Error
FDA (Devices) · 2024-04-03
- HighLaboratory performance verifier recalled for mismatched diluent assembly
FDA (Devices) · 2024-04-03
- HighVITROS Troponin I ES Immunodiagnostic Reagent Pack Recalled for Incorrect Wells
FDA (Devices) · 2023-08-16
- HighFDA Recalls Troponin Cardiac Test Calibrators for Potential False Negatives
FDA (Devices) · 2023-08-16
- HighOrtho Recalls VITROS Chemistry Calibrator Kit 20 for Analytical Bias
FDA (Devices) · 2023-07-19
Same hazard · imprecise results
See all →- ModerateRandox Lipase Assay Buffer Recalled for Imprecision in Quality Control
FDA (Devices) · 2020-08-12
- ModerateOrtho-Clinical Diagnostics Recalls VITROS Chemistry Products PHYT Slides
FDA (Devices) · 2018-07-11
- ModerateBeckman Coulter PrepPlus 2 Recalled for Sample Preparation Imprecision
FDA (Devices) · 2017-07-12
- HighSiemens Healthcare Recalls Emit 2000 Phenobarbital Assay Due to Imprecision
FDA (Devices) · 2013-11-13
- ModerateOrtho-Clinical Diagnostics Recalls VITROS 4600 Chemistry Systems
FDA (Devices) · 2013-04-24
Similar recalls
- HighOrtho-Clinical VITROS 5,1 FS Chemistry System Software Recall
FDA (Devices) · 2016-01-20
- ModerateOrtho-Clinical VITROS 5,1 FS Chemistry System Recalled for Biased Results
FDA (Devices) · 2014-06-18
- ModerateOrtho-Clinical Diagnostics Recalls VITROS 5,1 FS Chemistry System Software
FDA (Devices) · 2018-12-26
- ModerateOrtho-Clinical VITROS 5,1 FS Chemistry System Recall for Software Defect
FDA (Devices) · 2016-03-16
- ModerateOrtho VITROS 5,1 FS Chemistry System Software Recall
FDA (Devices) · 2015-09-09
- SevereOrtho-Clinical Diagnostics Recalls VITROS 5,1 FS Chemistry Systems Due to Electrical Short
FDA (Devices) · 2013-07-10