Ortho-Clinical VITROS 5,1 FS Chemistry System Software Recall
Ortho-Clinical Diagnostics is recalling VITROS 5,1 FS Chemistry Systems because a software anomaly may cause the device to process assays at incorrect temperatures, potentially producing erroneous results.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall is classified as FDA Class II, and the hazard involves a risk of harm (erroneous clinical results) where injury has not yet been reported, fitting the 'High' severity criteria.
Plain-English summary
Ortho-Clinical Diagnostics is recalling the VITROS 5,1 FS Chemistry System and VITROS 5,1 FS Chemistry System Refurbished, specifically units running Software Version 2.8 and below. The recall affects specific serial numbers for both the standard and refurbished models.
The recall is due to a software anomaly that may allow the systems to sample and process assays while fluids and reagent temperatures are not within the required range for optimal processing. When this intermittent issue occurs, the systems continue to operate and process results without notifying the user, which may result in erroneous test outcomes.
The affected products were distributed worldwide, including the United States, Puerto Rico, and numerous foreign countries. The source text does not specify a corrective action for consumers or healthcare providers, nor does it report any illnesses or injuries associated with this recall.
The recalled product
- Product
- VITROS 5,1 FS Chemistry System, Catalog Number 6801375, Unique Device Identifier Number 10758750001132, and VITROS 5,1 FS Chemistry System Refurbished, Catalog Number 6801890, Unique Device Identifier Number 10758750001644, Software Version 2.8 and below, IVD. Product Usage…
- Manufacturer
- Ortho-Clinical Diagnostics
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (20)
- Software Version 2.8 and below
- VITROS 5
- 1 FS Chemistry System Serial Numbers J34000016 - J34002323
- 1 FS Chemistry System Refurbished Serial Numbers 34000133
- 34000193
- 34000197
- 34000212
- 34000285
- 34000316
- 34000318
- 34000337
- 34000360
- 34000420
- 34000451
- 34000478
- 34000495
- 34000497
- 34000529
- 34000538
- 34000552
Distribution
Distributed nationwide across the United States.
Part of a larger recall action
This product is one of 4 recalled by Ortho-Clinical Diagnostics under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 4 products in this recall →Related recalls
Same manufacturer · Ortho-Clinical Diagnostics
See all →- HighClinical Laboratory Quality Control Device Recalled for Diluent Assembly Error
FDA (Devices) · 2024-04-03
- HighLaboratory performance verifier recalled for mismatched diluent assembly
FDA (Devices) · 2024-04-03
- HighVITROS Troponin I ES Immunodiagnostic Reagent Pack Recalled for Incorrect Wells
FDA (Devices) · 2023-08-16
- HighFDA Recalls Troponin Cardiac Test Calibrators for Potential False Negatives
FDA (Devices) · 2023-08-16
- HighOrtho Recalls VITROS Chemistry Calibrator Kit 20 for Analytical Bias
FDA (Devices) · 2023-07-19
Same hazard · erroneous results
See all →- HighBeckman Coulter DxC 500i Analyzer Modules Recalled for Software Defect
FDA (Devices) · 2026-08-26
- HighChemistry analyzer control modules recalled over software that may give erroneous results
FDA (Devices) · 2025-12-24
- HighLaboratory automation recalled over unlabelled or mismatched secondary sample tubes
FDA (Devices) · 2025-12-03
- HighBlood lead test kits recalled over mislabeled calibration buttons
FDA (Devices) · 2025-12-03
- HighLaboratory automation recalled over sample tubes with mismatched labels
FDA (Devices) · 2025-12-03
Similar recalls
- HighOrtho-Clinical VITROS 5,1 FS Chemistry System Software Recall
FDA (Devices) · 2015-05-06
- ModerateOrtho-Clinical VITROS 5,1 FS Chemistry System Recall for Software Defect
FDA (Devices) · 2016-03-16
- ModerateOrtho VITROS 5,1 FS Chemistry System Software Recall
FDA (Devices) · 2015-09-09
- ModerateOrtho-Clinical Diagnostics Recalls VITROS 5,1 FS Chemistry System Software
FDA (Devices) · 2018-12-26
- ModerateOrtho VITROS 5,1 FS Chemistry System Recalled for Software Calibration Anomaly
FDA (Devices) · 2019-08-14
- ModerateOrtho VITROS 5,1 FS Chemistry System Recalled for Software Calibration Issue
FDA (Devices) · 2019-08-14