The FDA (Devices) publishes a recall like this one as 3 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 3 pages.

Products
3
Announced
2013-04-24
Critical
0
Units

Why these products were recalled

Ortho-Clinical Diagnostics, Inc. (OCD) determined that extension springs on some VITROS¿ Systems may be out of its manufacturing specifications. Extension springs are part of the slide alignment guide assembly. During an internal evaluation in its manufacturing facility, it was discovered that an extension spring had insufficient force. If the extension spring does not provide sufficient force, MicroSlides may not be properly positioned within the MicroSlide incubator which has the potential to cause imprecise results. Extension springs are just one component in the slide alignment assem…

Ortho-Clinical Diagnostics, Inc. (OCD) determined that extension springs on some VITROS¿ Systems may be out of its manufacturing specifications. Extension springs are part of the slide alignment guide assembly. During an internal evaluation in its manufacturing facility, it was discovered that an extension spring had insufficient force. If the extension spring does not provide sufficient force, MicroSlides may not be properly positioned within the MicroSlide incubator which has the potential to cause imprecise results. Extension springs are just one component in the slide alignment asse…

The agency’s own wording, quoted from the enforcement report.

States named across these recalls

1–3 of 3

  • ModerateFDA (Devices)··2013-04-24

    Ortho-Clinical Diagnostics Recalls VITROS 4600 Chemistry Systems

    Ortho-Clinical Diagnostics is recalling VITROS 4600 Chemistry Systems because extension springs may be out of specification, potentially causing imprecise test results.

    Product
    VITROS 4600 Chemistry System (VITROS 5,1 FS System family member), Product Code 6802445. Intended for use in the measurement of a variety of chemistry analytes.
    Category
    Distribution
    1 state
  • ModerateFDA (Devices)··2013-04-24

    Ortho-Clinical Diagnostics Recalls VITROS 5600 Integrated Systems

    Ortho-Clinical Diagnostics is recalling VITROS 5600 Integrated Systems because extension springs may be out of specification, potentially causing imprecise chemistry test results.

    Product
    VITROS 5600 Integrated System, Product Code 6802413. Intended for use in the measurement of a variety of chemistry analytes.
    Category
    Distribution
    1 state
  • ModerateFDA (Devices)··2013-04-24

    Ortho-Clinical Diagnostics Recalls VITROS 5,1 FS Chemistry Systems

    Ortho-Clinical Diagnostics is recalling VITROS 5,1 FS Chemistry Systems because extension springs may be out of specification, potentially causing imprecise test results.

    Product
    VITROS 5,1 FS Chemistry System, Product Code 6801375 and VITROS 5,1 FS Chemistry System Refurbished, Product Code 6801890. Intended for use in the measurement of a variety of chemistry analytes.
    Category
    Distribution
    1 state