Dexcom G6 CGM App iOS Software Recall for Low Alarm Defect
Dexcom Inc is recalling the Dexcom G6 CGM App for iOS due to a software defect that may prevent low blood glucose alarms from triggering correctly.
What this means for you
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall involves a medical device software defect that could prevent critical low blood glucose alarms from triggering, posing a risk of harm to users, which aligns with the 'High' severity criteria for risk-of-harm products.
Plain-English summary
Dexcom Inc is recalling the Dexcom G6 CGM App for iOS, software number SW11677, versions 1.4.0 through 1.4.4. The recall involves 237,888 registered users in the United States and several international markets, including Austria, Belgium, Canada, and others.
The recall was initiated because the application's low alarm feature may not properly alert users when the Alert Schedule feature is enabled more than 30 days after the initial installation of the app. This defect could result in users not receiving critical low blood glucose warnings.
Consumers using the affected software versions should update their application to the corrected version 1.4.4 or later to ensure the low alarm feature functions as intended.
The recalled product
- Product
- Brand Name: Dexcom G6 CGM App for iOS Software Number: SW11677 Software Version: Version 1.4.0 up to corrected Version 1.4.4 GUDID Registered GTIN: 00386270000804
- Manufacturer
- Dexcom Inc
- Affected units
- 237,888
Distribution
Distribution scope not specified by the agency.
Related recalls
Same manufacturer · Dexcom Inc
See all →- SevereDexcom G7 Continuous Glucose Monitoring System sensors recalled
FDA (Devices) · 2026-07-01
- HighDexcom ONE+ iOS CGM App Software Defect Delayed Glucose Alerts
FDA (Devices) · 2026-05-20
- HighDexcom G7 CGM iOS App Software Defect Delays Glucose Alerts
FDA (Devices) · 2026-05-20
- HighDexcom G7 app defect may delay glucose readings and alerts
FDA (Devices) · 2026-05-20
- CriticalGlucose monitoring app recalled because it can terminate and miss alerts
FDA (Devices) · 2025-11-05
Same hazard · software defect
See all →- HighMidmark IQecg Digital ECG Software Recall for Inaccurate Waveforms
FDA (Devices) · 2026-09-02
- HighMidmark Digital ECG Software Recall for Incorrect Waveform Display
FDA (Devices) · 2026-09-02
- ModerateEndoFLIP System Recall Due to Incorrect Catheter Expiration Flags
FDA (Devices) · 2026-08-26
- HighBeckman Coulter DxC 500i Analyzer Modules Recalled for Software Defect
FDA (Devices) · 2026-08-26
- HighMitsubishi Recalls Outlander Models for Rearview Camera Software Defect
NHTSA · 2026-08-25