LW Scientific ZIP IQ Combo Centrifuge Rotor Recall for Breakage Risk
LW Scientific is recalling ZIP IQ Combo Centrifuges because the hematocrit rotor may break and eject through the housing, posing a safety hazard.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall involves a risk-of-harm product where the rotor may break and eject, posing a safety hazard, but no injuries or illnesses have been reported in the source text.
Plain-English summary
LW Scientific, Inc. is recalling the ZIP IQ Combo Centrifuge, model ZIC-CMBO-2RT3 Rev B. The recall involves 226 units distributed worldwide, including the United States, Canada, Australia, and several other countries.
The recall is due to a defect in the clear, plastic, polycarbonate hematocrit rotor. This component may break apart during use and be ejected through the plastic housing. This failure poses a safety hazard to the technician and may damage the surrounding area.
The centrifuge is used as a general-purpose device to spin down blood in hematocrit tubes and other fluids in test tubes. Users should stop using the affected rotors immediately and contact LW Scientific for instructions on the recall.
The recalled product
- Product
- LW Scientific ZIP IQ Combo Centrifuge, ZIC-CMBO-2RT3 Rev B Product Usage: used as a general purpose centrifuge to spin down blood in hematocrit tubes with additional rotor to spin other fluids in test tubes (15ml)
- Manufacturer
- LW Scientific, Inc.
- Affected units
- 226
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (20)
- Serial Numbers: 6C1712002
- 6C1712008
- 6C1803021
- 6C1803022
- 6C1803053
- 6C1803054
- 6C1803055
- 6C1803057
- 6C1803058
- 6C1803059
- 6C1803060
- 6C1803097
- 6C1803124
- 6C1803128
- 6C1803129
- 6C1803134
- 6C1803135
- 6C1803136
- 6C1803140
- 6C1805146
Distribution
Distributed nationwide across the United States.
Part of a larger recall action
This product is one of 3 recalled by LW Scientific, Inc. under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 3 products in this recall →Related recalls
Same manufacturer · LW Scientific, Inc.
See all →- HighLW Scientific ZIP IQ PCV Centrifuge Recalled for Rotor Breakage Risk
FDA (Devices) · 2019-04-03
- HighLW Scientific ZIP IQ Combo Centrifuge Rotor Recall for Breakage Hazard
FDA (Devices) · 2019-04-03
Same hazard · projectile
See all →- HighHonda Recalls Hybrid and Fuel Cell Vehicles for Tire Repair Kit Defect
NHTSA · 2026-10-08
- HighPhilips BGM 4 Blood Glucose Meter Battery Expansion Risk Recall
FDA (Devices) · 2025-03-12
- HighJoyson Recalls Side Curtain Air Bag Inflator Due to Detachment Risk
NHTSA · 2025-02-24
- High
- High
Similar recalls
- HighIris Diagnostics Recalls RTH8 Rotor for StatSpin Express 4 Centrifuge
FDA (Devices) · 2014-12-31
- HighIntelliCuff Portable Automatic Cuff Pressure Controller Recall
FDA (Devices) · 2026-07-08
- HighZoll QUATTRO Intravascular Heat Exchange Catheter Kit Recalled for Potential Tubing Leaks
FDA (Devices) · 2021-10-13
- HighZoll COOLINE Intravascular Catheter Kit recalled for potential leak hazard
FDA (Devices) · 2021-10-13
- HighZoll ICY intravascular heat exchange catheter kit recalled for potential leaks
FDA (Devices) · 2021-10-13
- HighBaxter Spectrum IQ Infusion Pump Recalled for Potential Front Panel Cracks
FDA (Devices) · 2024-05-22