The Recall Desk

FDA (Devices) recall action 82214

LW Scientific, Inc. recalled 3 products on 2019-04-03

The FDA (Devices) publishes a recall like this one as 3 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 3 pages.

Products
3
Announced
2019-04-03
Critical
0
Units
384

Why these products were recalled

The clear, plastic, polycarbonate hematocrit rotor may break apart during use and be ejected through the plastic housing, posing a safety hazard to the technician and potentially damaging the surrounding area.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–3 of 3

  • HighFDA (Devices)··2019-04-03

    LW Scientific ZIP IQ Combo Centrifuge Rotor Recall for Breakage Risk

    LW Scientific is recalling ZIP IQ Combo Centrifuges because the hematocrit rotor may break and eject through the housing, posing a safety hazard.

    Product
    LW Scientific ZIP IQ Combo Centrifuge, ZIC-CMBO-2RT3 Rev B Product Usage: used as a general purpose centrifuge to spin down blood in hematocrit tubes with additional rotor to spin other fluids in test tubes (15ml)
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2019-04-03

    LW Scientific ZIP IQ PCV Centrifuge Recalled for Rotor Breakage Risk

    LW Scientific is recalling 142 ZIP IQ PCV Centrifuges because the plastic rotor may break and be ejected, posing a safety hazard to technicians.

    Product
    LW Scientific ZIP IQ PCV Centrifuge, ZIC-24HD-75T3 Rev C Product Usage: used as a general purpose centrifuge to spin down blood in hematocrit tubes
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2019-04-03

    LW Scientific ZIP IQ Combo Centrifuge Rotor Recall for Breakage Hazard

    LW Scientific is recalling 16 ZIP IQ Combo Centrifuges because the hematocrit rotor may break and eject through the housing, posing a safety hazard to technicians.

    Product
    LW Scientific ZIP IQ Combo Centrifuge, ZIC-CMBO-4RT3 Rev B Product Usage: used as a general purpose centrifuge to spin down blood in hematocrit tubes with additional rotors to spin other fluids in test tubes (15ml and 50ml), microtubes, and PCR strips
    Category
    Distribution
    0 states