Philips MX 16-Slice CT Scanner Recall Due to Loose Table Screws
Philips Medical Systems is recalling 412 MX 16-Slice CT Scanner Systems because loose screws may cause the patient table top to detach.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall involves a structural defect in a medical device that could lead to patient injury (falling table), but the source text does not report any actual injuries or illnesses, fitting the 'High' criteria for risk-of-harm products where injury has not yet been reported.
Plain-English summary
Philips Medical Systems (Cleveland) Inc is recalling 412 units of the Philips MX 16-Slice CT Scanner System, a whole-body computed tomography X-ray system. The recall is due to a defect where the four screws that secure the patient table top to the carrier became loose, which caused the table top to detach.
The affected units were distributed worldwide, including in the United States in the states of Alabama, Arkansas, California, Florida, Georgia, Illinois, Indiana, Kentucky, Michigan, Mississippi, North Carolina, New Jersey, Nevada, New York, Ohio, Pennsylvania, Puerto Rico, and Virginia. The recall covers specific serial numbers listed in the official FDA record Z-0899-2013.
Healthcare facilities should contact Philips Medical Systems for instructions on how to address the defect. The source text does not specify if any injuries or illnesses have been reported as a result of this defect.
The recalled product
- Product
- Philips MX 16-Slice CT Scanner System, a whole-body computed tomography X-ray system.
- Manufacturer
- Philips Medical Systems (Cleveland) Inc
- Affected units
- 412
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (20)
- System Code #
- 728131
- Serial #: EP16E090001
- EP16E090002
- EP16E090003
- EP16E090004
- EP16E090006
- EP16E090007
- EP16E090008
- EP16E090009
- EP16E090010
- EP16E090011
- EP16E090012
- EP16E090013
- EP16E090015
- EP16E100001
- EP16E100004
- EP16E100006
- EP16E100008
- EP16E100010
Distribution
Related recalls
Same manufacturer · Philips Medical Systems (Cleveland) Inc
See all →- HighRadiation planning software recalled over image errors in three patient orientations
FDA (Devices) · 2025-09-24
- HighPinnacle 3 radiation planning software may miscalculate stopping power ratio
FDA (Devices) · 2025-08-13
- HighPinnacle3 Radiation Therapy Planning System dose-calculation software error
FDA (Devices) · 2022-09-21
- SeverePhilips Forte Gamma Camera System Recalled for Detector Fall Hazard
FDA (Devices) · 2019-11-13
- SeverePhilips Forte Gamma Camera System Recalled Due to Detector Falling Hazard
FDA (Devices) · 2019-11-13
Same hazard · structural defect
See all →- SevereSubaru Recalls 2026 Forester and Crosstrek for Ball Joint Defect
NHTSA · 2026-10-19
- Severe
- SevereVolvo Recalls 2025-2026 VNL Trucks for Loose Structural Bolts
NHTSA · 2026-10-15
- Severe
- Severe
Similar recalls
- HighPhilips MX 16-Slice CT Scanner Recalled for Pinch Point Injury Risk
FDA (Devices) · 2013-10-23
- ModeratePhilips MX 16-slice CT System Software Recall for Data Integrity Issues
FDA (Devices) · 2016-08-17
- HighPhilips Recalls MX 16-slice CT Scanners Due to Bolt Failure Risk
FDA (Devices) · 2012-09-26
- HighPhilips MX 16-slice CT Systems Recalled for Software Defects
FDA (Devices) · 2016-08-17
- HighPhilips and Neusoft Recall CT Scanners Due to Detaching Patient Table
FDA (Devices) · 2012-12-12
- ModeratePhilips MX 16-slice CT Scanner Recalled for Incorrect Image Orientation
FDA (Devices) · 2014-11-19