The Recall Desk
Agency classClass IIOur severity:HighFDA (Devices)·Z-0466-2013·Announced 2012-12-12

Philips and Neusoft Recall CT Scanners Due to Detaching Patient Table

Philips and Neusoft are recalling 45 CT scanners because the patient table top can detach from the carrier, posing a risk of injury.

The agency’s classification

FDA (Devices): Class II

Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.

This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work

What this means for you

Our assessment, not the agency’s: severity 3/5 (High).

Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.

Our severity reasoning: The recall involves a structural defect in a medical device that poses a risk of physical harm (falling table) to patients, but the source text does not report any injuries or illnesses, fitting the 'High' severity criteria for risk-of-harm products without reported injury.

Plain-English summary

Philips and Neusoft Medical Systems Co., Ltd. are recalling 20 units of the NeuViz 16 Multi-Sliced CT Scanner System and 25 units of the NeuViz Dual series CT Scanner System. The recall was initiated because the four screws that secure the patient table top to the carrier may pull out, causing the table top to detach.

The affected units were distributed nationwide, including in North Carolina, Ohio, Texas, and Puerto Rico. The specific serial numbers for the NeuViz 16 systems are N16E090002, N16E090003, N16E090004, N16E090007, N16E090008, N16E090010, N16E100015, N16E100017, N16E100023, N16E110016, N16E110023, N16E110026, N16E110027, N16E110040, N16E110053, N16E120016, N16E120022, N16E120020, N16E120023, and N16E120024. The serial numbers for the NeuViz Dual systems are 400562, 400563, 400626, NDH009EI, NDH011EI, NDH012EI, NDH013EI, NDH016EI, NDH017EI, NDH026EI, NDH027EI, NDH028EI, NDH030EI, NDH031EI, NDH034EI, NDHR080001, NDHR080002, NDHR080003, NDHR080004, NDHR080007, NDHR080012, NDHR080021, NDHR090003, NDHR090012, and NDHR090013.

Healthcare facilities should contact the recalling firm to arrange for the necessary repairs or replacements to ensure the patient table is securely attached.

The recalled product

Product
NeuViz 16 Multi-Sliced CT Scanner System, part number 989605858501. NeuViz Dual series CT Scanner System, part number 989605651321 The NeuViz 16 CT Scanner System can be used as a whole body computed tomography X-ray system featuring a continuously rotating X-ray tube and…

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (20)

  • NeuViz 16 system Serial numbers: N16E090002
  • N16E090003
  • N16E090004
  • N16E090007
  • N16E090008
  • N16E090010
  • N16E100015
  • N16E100017
  • N16E100023
  • N16E110016
  • N16E110023
  • N16E110026
  • N16E110027
  • N16E110040
  • N16E110053
  • N16E120016
  • N16E120022
  • N16E120020
  • N16E120023
  • N16E120024. Neuviz Dual Serial numbers: 400562

Distribution

Distributed nationwide across the United States.

The notice also names 4 states specifically:

Same manufacturer · Philips And Neusoft Medical Systems Co., Ltd.

See all →

Similar recalls