Philips and Neusoft Recall CT Scanners Due to Detaching Patient Table
Philips and Neusoft are recalling 45 CT scanners because the patient table top can detach from the carrier, posing a risk of injury.
What this means for you
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall involves a structural defect in a medical device that poses a risk of physical harm (falling table) to patients, but the source text does not report any injuries or illnesses, fitting the 'High' severity criteria for risk-of-harm products without reported injury.
Plain-English summary
Philips and Neusoft Medical Systems Co., Ltd. are recalling 20 units of the NeuViz 16 Multi-Sliced CT Scanner System and 25 units of the NeuViz Dual series CT Scanner System. The recall was initiated because the four screws that secure the patient table top to the carrier may pull out, causing the table top to detach.
The affected units were distributed nationwide, including in North Carolina, Ohio, Texas, and Puerto Rico. The specific serial numbers for the NeuViz 16 systems are N16E090002, N16E090003, N16E090004, N16E090007, N16E090008, N16E090010, N16E100015, N16E100017, N16E100023, N16E110016, N16E110023, N16E110026, N16E110027, N16E110040, N16E110053, N16E120016, N16E120022, N16E120020, N16E120023, and N16E120024. The serial numbers for the NeuViz Dual systems are 400562, 400563, 400626, NDH009EI, NDH011EI, NDH012EI, NDH013EI, NDH016EI, NDH017EI, NDH026EI, NDH027EI, NDH028EI, NDH030EI, NDH031EI, NDH034EI, NDHR080001, NDHR080002, NDHR080003, NDHR080004, NDHR080007, NDHR080012, NDHR080021, NDHR090003, NDHR090012, and NDHR090013.
Healthcare facilities should contact the recalling firm to arrange for the necessary repairs or replacements to ensure the patient table is securely attached.
The recalled product
- Product
- NeuViz 16 Multi-Sliced CT Scanner System, part number 989605858501. NeuViz Dual series CT Scanner System, part number 989605651321 The NeuViz 16 CT Scanner System can be used as a whole body computed tomography X-ray system featuring a continuously rotating X-ray tube and det
- Manufacturer
- Philips And Neusoft Medical Systems Co., Ltd.
- Category
- Medical Device — CT Scanner
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (20)
- NeuViz 16 system Serial numbers: N16E090002
- N16E090003
- N16E090004
- N16E090007
- N16E090008
- N16E090010
- N16E100015
- N16E100017
- N16E100023
- N16E110016
- N16E110023
- N16E110026
- N16E110027
- N16E110040
- N16E110053
- N16E120016
- N16E120022
- N16E120020
- N16E120023
- N16E120024. Neuviz Dual Serial numbers: 400562
Distribution
Related recalls
Same manufacturer · Philips And Neusoft Medical Systems Co., Ltd.
See all →- HighPhilips NeuViz 16 CT Scanner Recall for Patient Data Error
FDA (Devices) · 2016-05-04
- HighPhilips NeuViz Dual CT Scanner Recalled for Pinch Hazard
FDA (Devices) · 2013-05-01
- HighPhilips NeuViz 16 CT Scanner Recalled Due to Pinch Hazard
FDA (Devices) · 2013-05-01
- HighPhilips and Neusoft Recall NeuViz 16 CT Scanners Due to Detached Oil Accumulator
FDA (Devices) · 2012-08-01
Same hazard · structural defect
See all →- HighCanon Medical Alphenix X-ray Systems Recalled for Loose Ceiling Screws
FDA (Devices) · 2026-09-02
- HighLight Boy Recalls GBX12SK Light Towers Due to Frame Bending Risk
NHTSA · 2026-09-01
- SevereKmaier Infant Walkers Recalled for Serious Fall Hazard
CPSC · 2026-08-27
- Highrestor3d Recalls Veritas Standard Glenoid Baseplates Due to Machining Defect
FDA (Devices) · 2026-08-26
- SevereFord Recalls 2026 Maverick, F-150, and Bronco Sport for Joint Issues
NHTSA · 2026-08-25