Traverse Rail Carrier E-System components with defective short screws
Hill-Rom Traverse Rail Carrier E-System components contain non-conforming screws that are too short to secure the rail properly. The screws could loosen over time, causing the rail system to fall and potentially injure caregivers or patients.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: This is a Class II FDA medical device recall involving a structural defect with potential risk of fall-related injury. The source text reports no actual illnesses or injuries; the harm is theoretical. Per the rubric, risk-of-harm products without reported injury score at most 3.
Plain-English summary
Hill-Rom, Inc. is recalling Traverse Rail Carrier E-System and Traverse Rail Carrier components, which are parts of Liko overhead lift systems used in healthcare settings for patient handling and mobility assistance.
The recall is due to non-conforming screws supplied with the devices that are too short to ensure proper installation. The screws may be unable to secure the traverse rail appropriately and could loosen over time, potentially causing the rail system to fall. This poses a risk of injury to caregivers and patients.
The affected components were distributed worldwide between January 4, 2021 and March 24, 2021, including to facilities throughout the United States and to Austria, Canada, Czech Republic, Denmark, Spain, Finland, France, Great Britain, Hong Kong, Ireland, Israel, Italy, Japan, Netherlands, Norway, New Zealand, Russia, Sweden, Singapore, and Slovakia.
The recalled product
- Product
- (1)Traverse Rail Carrier E-System, wide 31017XX component to Liko overhead lift, (2)Traverse Rail Carrier 31025XX component to Liko overhead lift. Intended for to make safe lifting of a patient possible. Affected Devices: 3101704 TRAV RAIL CARR E-SYST STD 3102514 TR…
- Manufacturer
- Hill-Rom, Inc.
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (18)
- natural 3122007NA Likorall 242 ES R2R
- natural 3122008 Likorall 242 ES R2R
- white 3122008NA Likorall 242 ES R2R
- white 3122011 Likorall 242 S R2R
- natural 3122011NA Likorall 242 S R2R
- natural 3122012 Likorall 242 S R2R
- white 3122012NA Likorall 242 S R2R
- white 3123001 Likorall 243 ES
- natural 3123001NA Likorall 243 ES
- natural 3123002 Likorall 243 ES
- white 3123002NA Likorall 243 ES
- white 3122501 Likorall 250 ES
- natural 3122501NA Likorall 250 ES
- natural 3122502 Likorall 250 ES
- white 3122502NA Likorall 250 ES
- white 3124050 Likorall 250 S
- IRC
- natural
Distribution
Related recalls
Same manufacturer · Hill-Rom, Inc.
See all →- HighLiko M220 mobile lifts recalled for missing bushing, fall risk
FDA (Devices) · 2022-02-23
- HighLiko Mobile Patient Lifts Recalled for Missing Bushings and Wear Risk
FDA (Devices) · 2022-02-23
- HighNurse Call Communication System Components May Fail to Reboot
FDA (Devices) · 2021-12-29
- HighPatient Lift Device Stretch Leveller May Lose Horizontal Position Control
FDA (Devices) · 2021-10-06
- HighPatient lift devices may lose ability to maintain horizontal positioning during transfers
FDA (Devices) · 2021-10-06
Same hazard · loose fasteners
See all →- SevereLucid Air half-shaft may disconnect from the drive unit
NHTSA · 2026-03-26
- HighStryker sagittal saw attachment cover screws may loosen and detach
FDA (Devices) · 2026-03-25
- Severe
- SevereHonda motorcycle handlebar lock screw may loosen and detach
NHTSA · 2026-03-12
- HighHyundai IONIQ battery bus bar may be improperly tightened
NHTSA · 2026-02-05
Similar recalls
- SevereHill-Rom Liko Multirall 200 Overhead Lifts Recalled for Detachment Risk
FDA (Devices) · 2021-02-10
- SevereHill-Rom Recalls Liko Overhead Rail Systems Due to Corrosion
FDA (Devices) · 2013-02-27
- HighPatient lift devices may lose ability to maintain horizontal positioning during transfers
FDA (Devices) · 2021-10-06
- SevereHill-Rom Recalls Liko Universal SlingBars Due to Center Bolt Failure Risk
FDA (Devices) · 2016-06-15
- SevereHillrom LikoScale patient lift components may falsely latch and detach
FDA (Devices) · 2025-07-16
- HighLiko Multirall 200Liko Strap Gripper Overhead Lift Recall
FDA (Devices) · 2018-10-24