Stryker sagittal saw attachment cover screws may loosen and detach
Howmedica Osteonics is recalling the MICS3 Angled Sagittal Saw Attachment because the external screws securing the cover to the saw housing may loosen and detach during use.
What this means for you
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: Rubric level 3 covers risk-of-harm products where injury has not yet been reported: screws may detach from a surgical saw during use, and no injuries are stated in the source.
Plain-English summary
Howmedica Osteonics Corp. is recalling the MICS3 Angled Sagittal Saw Attachment, part number 210490. About 1,098 units were distributed nationwide in the United States and to Australia, Germany, Belgium, France, Spain, the United Kingdom and New Zealand. The FDA has classified this recall as Class II.
A potential issue was identified with the torque strength on the external screws of the attachment. These screws, which secure the cover to the Angled Saw Housing, may loosen during use and detach from the attachment.
All lot numbers are affected. The device carries UDI 07613327523874.
Facilities can identify affected attachments by the part number and UDI above, and may contact Howmedica Osteonics Corp. for further information.
The recalled product
- Product
- MICS3 Angled Sagittal Saw Attachment; Part Number: 210490
- Manufacturer
- Howmedica Osteonics Corp.
- Affected units
- 1,098
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (3)
- Part Number: 210490
- UDI: (01)07613327523874(10)TU6563205RP
- Lot numbers: ALL
Distribution
Distributed nationwide across the United States.
Related recalls
Same manufacturer · Howmedica Osteonics Corp.
See all →- HighStryker Triathlon size 3 baseplate packaging contained a size 4 implant
FDA (Devices) · 2026-06-24
- HighStryker Triathlon size 4 baseplate packaging contained a size 3 implant
FDA (Devices) · 2026-06-24
- ModerateCranial tracker devices recalled over an error message preventing use
FDA (Devices) · 2025-12-31
- ModerateHip stems recalled over potential mix of two catalog numbers in packaging
FDA (Devices) · 2025-10-08
- HighKnee Replacement Implant Recalled for Out-of-Spec Locking Tabs
FDA (Devices) · 2024-10-16
Same hazard · component detachment
See all →- ModerateGalt Medical Recalls Boston Scientific Transcarotid Access Kits
FDA (Devices) · 2026-08-19
- ModerateHarmonyAIR A-Series surgical light intensity control buttons may detach
FDA (Devices) · 2026-06-24
- ModerateALYON surgical light head intensity control buttons may detach
FDA (Devices) · 2026-06-24
- SevereForest River trailer shock bolts may break and detach onto the road
NHTSA · 2026-05-26
- SevereFord Bronco hardtop roof panel sections may separate and detach
NHTSA · 2026-05-12