TS Femoral Prep Tray, Triathlon. Catalog Number: 6543-6-507.
Reprocessing instructions provided in LSTPI-B specify a sterilization dry time of 30 minutes for the trays sterilized in blue wrap.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
Severity not yet assessed
We have not scored this recall against our rubric yet, so we are not showing a severity rating or our usual guidance. The agency’s own notice below is the authoritative source — read it first.
Plain-English summary
Reprocessing instructions provided in LSTPI-B specify a sterilization dry time of 30 minutes for the trays sterilized in blue wrap. When blue wrap is used, the 30-minute dry time may lead to a small amount of residual moisture remaining after sterilization.
The recalled product
- Product
- TS Femoral Prep Tray, Triathlon. Catalog Number: 6543-6-507.
- Manufacturer
- Howmedica Osteonics Corp.
- Category
- Medical Device — Devices
- Affected units
- 582
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (1)
- 2026 are impacted.
Distribution
Distribution scope not specified by the agency.
Related recalls
Same manufacturer · Howmedica Osteonics Corp.
See all →- SevereHowmedica Recalls Scorpio Patella Implants Due to Polyethylene Wear Risk
FDA (Devices) · 2026-09-23
- HighHowmedica Recalls Orthopaedic Implants Due to Potential UHMWPE Oxidation
FDA (Devices) · 2026-09-23
- SevereHowmedica Recalls Orthopaedic Implants Due to Potential UHMWPE Oxidation
FDA (Devices) · 2026-09-23
- HighHowmedica Recalls Orthopaedic Implant Components Due to Material Aging
FDA (Devices) · 2026-09-23
- HighHowmedica Recalls Scorpio Receded Metal Backed Patella Implants
FDA (Devices) · 2026-09-23