Radiation therapy planning software displays wrong images during animation playback
RayStation/RayPlan versions 6.0.0.24 through 9.2.0.483 incorrectly display selected images as both primary and secondary when animation playback is running, affecting treatment planning visualization.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: FDA Class II medical device with a display defect affecting radiation therapy treatment planning visualization. No illnesses or injuries reported; this is a preventable risk-of-harm scenario addressed through software updates.
Plain-English summary
RayStation/RayPlan is FDA-cleared medical device software designed for radiation therapy treatment planning and analysis. Multiple software versions (6.0.0.24 through 9.2.0.483) contain a defect in the image display system.
When a user selects a new primary image set while the cine loop (animation playback) is running, the software incorrectly displays the selected image set as both the primary and secondary image in all side-by-side viewing modes. The same display error occurs when a new patient or case is opened while the cine loop is running.
The defect affects radiation therapy centers across the United States where 75 licenses of the software are deployed. Users should contact RaySearch Laboratories AB for available software updates and guidance to prevent treatment planning errors.
The recalled product
- Product
- RayStation/RayPlan- designed for treatment planning and analysis of radiation therapy. Software Version: 6.0.0.24 , 6.1.0.26 , 6.1.1.2 , 6.2.0.7 , 6.3.0.6 , 7.0.0.19 , 8.0.0.61 , 8.0.1.10 , 8.1.0.47 , 8.1.1.8 , 8.1.2.5 , 9.0.0.113 , 9.1.0.933 , 9.2.0.483 .…
- Manufacturer
- RAYSEARCH LABORATORIES AB
- Affected units
- 75
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (13)
- Software: 6.0.0.24
- 6.1.0.26
- 6.1.1.2
- 6.2.0.7
- 6.3.0.6
- 7.0.0.19
- 8.0.0.61
- 8.0.1.10
- 8.1.0.47
- 8.1.1.8
- 8.1.2.5
- 9.0.0.113
- 9.1.0.933
Distribution
Distributed nationwide across the United States.
Related recalls
Same manufacturer · RAYSEARCH LABORATORIES AB
See all →- HighRayCare 2024A SP1 Oncology Information System Software Recall
FDA (Devices) · 2026-07-08
- HighRayCare 2024A SP4 Oncology System Beam Set Delivery Note Data Loss
FDA (Devices) · 2026-07-08
- HighRayStation 2024A treatment planning recalled over dose that fails to invalidate
FDA (Devices) · 2026-01-28
- HighRayStation 2023B and 2024A recalled over dose that fails to invalidate
FDA (Devices) · 2026-01-28
- HighRayStation v2025 treatment planning recalled over dose that fails to invalidate
FDA (Devices) · 2026-01-28
Same hazard · display error
See all →- ModerateSpectral Instruments Recalls 35 Ami HTX X-Ray Units
FDA (Devices) · 2026-08-26
- ModerateSpectral Instruments Lago X Medical Device Recalled for Display Glitch
FDA (Devices) · 2026-08-26
- ModerateKico Knee Innovation Recalls ARVIS Surgical Navigation System Due to Sensor Issue
FDA (Devices) · 2026-08-26
- ModerateLinkBio CORE Workstation may display an incorrect planning date
FDA (Devices) · 2026-04-22
- HighHexagon Purus Recalls Tern RC8 Vehicles Due to Brake Display Error
NHTSA · 2026-04-20
Similar recalls
- HighMedical Device Software Defects in RayStation Treatment Planning System
FDA (Devices) · 2021-06-16
- HighRayStation 5.0 Radiation Therapy Software Recalled for Dose Display Error
FDA (Devices) · 2016-12-14
- ModerateRayStation Radiation Therapy Software Recalled for Display Errors
FDA (Devices) · 2015-09-23
- HighRayStation Radiation Therapy Software Recalled for Dose Calculation Accuracy Issues
FDA (Devices) · 2018-07-04
- HighRayStation radiation therapy planning software density uncertainty computational defect
FDA (Devices) · 2025-02-26
- HighRayStation Therapy Treatment Planning System Recalled for Image Offset Errors
FDA (Devices) · 2013-06-05