Medical Device Software Defects in RayStation Treatment Planning System
RayStation radiation therapy treatment planning software versions 6 through 10A contain defects that may cause misleading dose evaluations and contouring errors. Healthcare facilities should verify treatment plans and contact the manufacturer for guidance.
What this means for you
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: This is an FDA Class II medical device recall involving software defects in a critical treatment-planning tool. While no illnesses or injuries have been reported, the defects affect dose calculation and contouring, which are essential functions that directly impact patient treatment accuracy in cancer care.
Plain-English summary
RayStation is a treatment planning system used by cancer centers to design radiation therapy. The recall affects versions 6.0.0.24 through 10.1.0.613 of RayStation and related RayPlan software, with approximately 13,722 systems deployed domestically across the United States and internationally.
The software contains two defects. First, when clinical goals are edited, the evaluation of dose distributions may become misleading. Second, contouring functions may produce errors when the spacing between image slices is 1 millimeter or less. These defects could affect treatment planning accuracy.
Healthcare facilities using affected RayStation versions should verify the accuracy of their treatment plans and contact RaySearch Laboratories for a software update or additional guidance on managing this issue.
The recalled product
- Product
- RayStation (Treatment Planning System) Model Number RayStation is 6.0.0.24 , 6.1.1.2 , 6.2.0.7 , 6.3.0.6 , 7.0.0.19 , 8.0.0.6 , 8.0.1.10 , 8.1.0.47 , 8.1.1.8 , 8.1.2.5 , 9.0.0.113 , 9.1.0.933 , 9.2.0.483 , 10.0.0.1154 , 10.0.1.52 or 10.1.0.613 .
- Manufacturer
- RAYSEARCH LABORATORIES AB
- Affected units
- 13,722
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (9)
- Software version: RayStation 6
- 7
- 8A
- 8B
- 9A
- 9B
- 10A
- 10B
- RayPlan 2 including some service packs.
Distribution
Related recalls
Same manufacturer · RAYSEARCH LABORATORIES AB
See all →- HighRayCare 2024A SP1 Oncology Information System Software Recall
FDA (Devices) · 2026-07-08
- HighRayCare 2024A SP4 Oncology System Beam Set Delivery Note Data Loss
FDA (Devices) · 2026-07-08
- HighRayStation 2024A treatment planning recalled over dose that fails to invalidate
FDA (Devices) · 2026-01-28
- HighRayStation 2023B and 2024A recalled over dose that fails to invalidate
FDA (Devices) · 2026-01-28
- HighRayStation v2025 treatment planning recalled over dose that fails to invalidate
FDA (Devices) · 2026-01-28
Same hazard · software defect
See all →- HighMidmark IQecg Digital ECG Software Recall for Inaccurate Waveforms
FDA (Devices) · 2026-09-02
- HighMidmark Digital ECG Software Recall for Incorrect Waveform Display
FDA (Devices) · 2026-09-02
- ModerateEndoFLIP System Recall Due to Incorrect Catheter Expiration Flags
FDA (Devices) · 2026-08-26
- HighBeckman Coulter DxC 500i Analyzer Modules Recalled for Software Defect
FDA (Devices) · 2026-08-26
- HighMitsubishi Recalls Outlander Models for Rearview Camera Software Defect
NHTSA · 2026-08-25