The Recall Desk
Agency classClass IIOur severity:HighFDA (Devices)·Z-0591-2013·Announced 2012-12-26

DePuy Solution Calcar Stem Recalled for Potential Impingement

DePuy Orthopaedics is recalling Solution Calcar Stems because an engineering analysis found they may impinge with certain heads, preventing proper locking.

The agency’s classification

FDA (Devices): Class II

Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.

This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work

What this means for you

Our assessment, not the agency’s: severity 3/5 (High).

Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.

Our severity reasoning: The recall involves a risk-of-harm medical device where the hazard is mechanical failure (impingement) that has not yet been reported to cause injury, fitting the 'High' severity criteria for risk-of-harm products without reported illness.

Plain-English summary

DePuy Orthopaedics, Inc. is recalling 1,732 units of the Solution System Calcar Stem (Product Code 157226210). The recall covers specific lot codes including X43EX1000, X43EXA000, X43EY1000, X49FH1000, X4GDG1000, X57HX1000, and YH1ENA000. These medical devices are used in total hip arthroplasty to replace damaged hip joints.

The recall was initiated after an engineering analysis determined that the Solution Calcar Stem has the potential to result in impingement with a variety of femoral heads. This impingement may prevent the stem and head from locking as intended. The amount of possible impingement varies depending on the specific femoral head used, with a worst-case condition involving a -2mm offset articuleze head resulting in a maximum of 0.937mm radial impingement.

The affected products were distributed nationwide across numerous states, including Alabama, Arizona, California, Florida, New York, and Texas, as well as Guaynabo, Puerto Rico. Healthcare providers and patients should contact DePuy Orthopaedics for further instructions regarding the return or replacement of these devices.

The recalled product

Product
SOL SYS R 9IN CALC 2.25/21 Packaging: Product is packed within a poly protector, with protection added as needed for fragile points. The entire stem is then placed within a foam sleeve for added protection. The assembly is then placed and sealed in an inner and outer tray…
Affected units
1,732

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (7)

  • Product Code: 157226210 and Lot codes: X43EX1000
  • X43EXA000
  • X43EY1000
  • X49FH1000
  • X4GDG1000
  • X57HX1000
  • and YH1ENA000.

Distribution

Distributed nationwide across the United States.

The notice also names 47 states specifically:

Part of a larger recall action

This product is one of 20 recalled by DePuy Orthopaedics, Inc. under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.

See all 20 products in this recall →

Similar recalls