Philips IntelliSpace Portal Software Defect May Display Incorrect Cardiac Measurements
Philips is recalling 32 IntelliSpace Portal systems due to a software defect that may display incorrect cardiac and aortic anatomy measurements used for sizing Transcatheter heart valves.
What this means for you
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall involves a software defect in a medical device that may lead to incorrect sizing of a heart valve, representing a risk of harm where injury has not yet been reported.
Plain-English summary
Philips Medical Systems (Cleveland) Inc is recalling 32 units of IntelliSpace Portal DX/HX/EX and IX systems due to a software defect. The affected model numbers are 881001, 881011, 881012, and 881030.
The defect may cause the application to display incorrect measurements of cardiac and aortic anatomy in certain circumstances. These measurements are used to select and size the Transcatheter heart valve (THV) to be implanted.
The recall covers units distributed worldwide, including in the United States (Florida, North Carolina, and Vermont) and in Argentina, Austria, Belgium, Canada, Denmark, France, India, Italy, Japan, Netherlands, United Kingdom, and Vietnam. Users should contact Philips Medical Systems for further instructions regarding the software defect.
The recalled product
- Product
- IntelliSpace Portal DX/HX/EX model number: 881001 IntelliSpace Portal DX/HX/EX Upgrade model number: 881011 IntelliSpace Portal DX/HX/EX Demo model number: 881012 IntelliSpace Portal IX model number; 881030 System, x-ray, tomography, computed
- Manufacturer
- Philips Medical Systems (Cleveland) Inc
- Affected units
- 32
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (20)
- IntelliSpace Portal DX/HX/EX: 96027
- 96029
- 96025
- 96030
- 96031
- 85479
- 79935
- 98087
- 85175
- 96024
- 106013
- 450003
- 950002
- 950003
- 35009
- 35030
- 870056
- 550012
- 85081
- 97069
Distribution
Related recalls
Same manufacturer · Philips Medical Systems (Cleveland) Inc
See all →- HighRadiation planning software recalled over image errors in three patient orientations
FDA (Devices) · 2025-09-24
- HighPinnacle 3 radiation planning software may miscalculate stopping power ratio
FDA (Devices) · 2025-08-13
- HighPinnacle3 Radiation Therapy Planning System dose-calculation software error
FDA (Devices) · 2022-09-21
- SeverePhilips Forte Gamma Camera System Recalled for Detector Fall Hazard
FDA (Devices) · 2019-11-13
- ModeratePhilips Forte Gamma Camera System Recalled for Detector Fall Hazard
FDA (Devices) · 2019-11-13
Same hazard · software defect
See all →- HighMidmark IQecg Digital ECG Software Recall for Inaccurate Waveforms
FDA (Devices) · 2026-09-02
- HighMidmark Digital ECG Software Recall for Incorrect Waveform Display
FDA (Devices) · 2026-09-02
- ModerateEndoFLIP System Recall Due to Incorrect Catheter Expiration Flags
FDA (Devices) · 2026-08-26
- HighBeckman Coulter DxC 500i Analyzer Modules Recalled for Software Defect
FDA (Devices) · 2026-08-26
- HighMitsubishi Recalls Outlander Models for Rearview Camera Software Defect
NHTSA · 2026-08-25