Hospira Recalls Plum A+ Infusion Systems Due to Broken Sensor Pin Risk
Hospira is recalling 71 Plum A+ Infusion Systems because a pressure sensor pin may break, potentially preventing the detection of line occlusions.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall involves a medical device with a potential failure mode (broken sensor pin) that could prevent the detection of line occlusions, posing a risk of harm to patients, although no specific injuries or illnesses are reported in the source text.
Plain-English summary
Hospira Inc. is recalling 71 units of the Plum A+ Infusion System. This cassette-based multifunction infusion system is designed to deliver parenteral, enteral, or epidural infusions and is indicated for use in various therapies, including the administration of whole blood and blood products. The affected units were distributed nationwide in the United States, including Puerto Rico and the U.S. Virgin Islands, as well as in Costa Rica and the Dominican Republic.
The recall is being issued because the distal (occlusion) pressure sensor pin has the potential to break. This pin is a critical component of the subsystem that measures pressure within the distal line of the administration set to indicate the presence of a full or partial distal occlusion. A broken pin can only be detected through visual inspection of the distal pressure pin.
Healthcare facilities and users should check their inventory for the specific serial numbers listed in the recall notice. If a unit is affected, it should be removed from service and inspected visually for a broken distal pressure pin. Users should contact Hospira for further instructions on the recall process.
The recalled product
- Product
- The Plum A+ Infusion System is a cassette based multifunction infusion system that allows two lines in and one line out and can be used for standard, piggyback, or concurrent delivery. The Plum A+ Infusion System is designed to deliver parenteral, enteral, or epidural infusions…
- Manufacturer
- Hospira Inc.
- Category
- Medical Device — Infusion Pumps
- Affected units
- 71
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (20)
- 1) List Number 11973
- Serial Numbers: 12570148
- 12570150
- 12570154
- 12570163
- 12570170
- 12570177
- 12570204
- 12570211
- 12570212
- 12570255
- 12570270
- 12570319
- 12570359
- 12570385
- 12570413
- 12570425
- 12570455
- 12570487
- 12570662
Distribution
Part of a larger recall action
This product is one of 8 recalled by Hospira Inc. under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 8 products in this recall →Related recalls
Same manufacturer · Hospira Inc.
See all →- CriticalHospira Recalls Dextrose Injection Syringes Due to Human Hair Particulate
FDA (Drugs) · 2017-05-31
- ModerateHospira Recalls Minibore Extension Sets Due to Solvent Occlusion
FDA (Devices) · 2017-01-18
- LowHospira Recalls Lidocaine and Dextrose Injection for Stability Failures
FDA (Drugs) · 2016-12-14
- SevereHospira Recalls Fentanyl Citrate Injection Ampules Due to Broken Tips
FDA (Drugs) · 2016-11-30
- HighHospira Symbiq Infusion Pump Recall for White Screen Safety Shutdown
FDA (Devices) · 2016-11-16
Same hazard · sensor failure
See all →- SevereAbiomed Recalls Impella Optical Controllers Due to Purge Sensor Defect
FDA (Devices) · 2026-09-23
- SevereAbiomed Recalls Automated Impella Controllers Due to Purge Sensor Defect
FDA (Devices) · 2026-09-23
- High
- Severe2025–2026 Ford Bronco Sport and Maverick Occupant Sensor Recall
NHTSA · 2026-03-31
- ModerateBubble sensors recalled over connecting cables that can be damaged
FDA (Devices) · 2026-02-11
Similar recalls
- SevereHospira Recalls Plum A+ Infusion Pumps With Silent Alarm Defect
FDA (Devices) · 2015-03-04
- SevereHospira Recalls Plum A+ Infusion Pumps With Failing Alarm Assemblies
FDA (Devices) · 2015-03-04
- SevereHospira Recalls Plum A+ Infusion Pumps Due to Alarm Failure
FDA (Devices) · 2015-03-04
- SevereHospira Plum A+ Infusion Pump Alarm Assembly Recall
FDA (Devices) · 2015-03-04
- HighHospira Recalls 2,567 Plum A+ Infusion Pumps Due to Calibration Error
FDA (Devices) · 2016-06-15
- SevereHospira Recalls Plum A+ Infusion Pumps Due to Pressure Sensor Drift
FDA (Devices) · 2013-11-20