Samsung GM60A-40S Radiographic System Recalled for Drive Control Defect
NeuroLogica Corporation is recalling seven Samsung GM60A-40S radiographic systems because drive behavior may not be properly controlled when moving over conductive floors.
What this means for you
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall is classified as FDA Class II, and the hazard involves a risk of harm (loss of control) where injury has not yet been reported, fitting the 'High' severity criteria.
Plain-English summary
NeuroLogica Corporation is recalling seven Samsung GM60A-40S radiographic systems. These devices are intended for use in generating radiographic images of human anatomy. The recall involves specific serial numbers: 50EAM9BG200004X, 50EAM9BG200001A, 50EAM9BG200002P, 50EAM9BG200003D, 50EAM9BG200005N, 50EAM9BG200006E, and 50EAM9BG200007F.
The recall is due to a defect where the system's drive behavior may not be properly controlled when the system moves from a non-conductive floor to a conductive, grounded part. The system can be stopped by releasing the drive handle or pressing the emergency switch, and proper behavior is re-enabled when the system is rebooted.
The affected devices were distributed to the states of North Carolina, Ohio, Massachusetts, and Missouri. Users should contact NeuroLogica Corporation for instructions on how to address the issue.
The recalled product
- Product
- Samsung GM60A-40S Intended for use in generating radiographic images of human anatomy.
- Manufacturer
- NeuroLogica Corporation
- Hazard
- Affected units
- 7
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (6)
- Serial Numbers: 50EAM9BG200004X
- 50EAM9BG200001A 50EAM9BG200002P
- 50EAM9BG200003D
- 50EAM9BG200005N
- 50EAM9BG200006E
- 50EAM9BG200007F
Distribution
Related recalls
Same manufacturer · NeuroLogica Corporation
See all →- HighMobile X-ray systems recalled over bolts that may loosen or fracture
FDA (Devices) · 2025-11-19
- HighNeuroLogica GM60A Mobile X-Ray System Anti-Fall Arm Safety Failure Risk
FDA (Devices) · 2024-08-28
- HighSamsung GM85 Mobile X-ray System Welding Frame Defect Arm Fall Risk
FDA (Devices) · 2024-02-21
- HighOmniTom CT imaging device recalled for wheel loosening and battery contamination
FDA (Devices) · 2023-07-12
- HighSamsung GM85 Digital X-ray Imaging System Arm Latch Failure
FDA (Devices) · 2023-04-26
Same hazard · loss of control
See all →- SevereTesla Recalls 2026 Model Y Vehicles for Front Suspension Fastener Defect
NHTSA · 2026-08-31
- SevereFord Recalls 2026 Maverick, F-150, and Bronco Sport for Joint Issues
NHTSA · 2026-08-25
- SevereHolley Recalls Replacement Carburetors and Throttle Body Kits
NHTSA · 2026-08-17
- SevereLand Rover Recalls 2016-2017 Range Rovers for Suspension Knuckle Defect
NHTSA · 2026-08-13
- SevereBMW X3 Vehicles Recalled Due to Potential Loss of Steering Control
NHTSA · 2026-08-12