Mobile X-ray systems recalled over bolts that may loosen or fracture
NeuroLogica is recalling 236 GM85 digital mobile X-ray systems because the four M5 bolts securing the tube head unit to the arm may loosen or fracture from metal fatigue.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: Scored 3 under the risk-of-harm criterion: bolts that can fracture on a mounted tube head create a fall hazard, and no injuries are reported.
Plain-English summary
The GM85 Digital Mobile X-ray Imaging System, model GM85, from NeuroLogica, is being recalled.
The four M5 bolts that secure the Tube Head Unit to the end of the arm may loosen or fracture due to metal fatigue.
The recall covers 236 units carrying UDI-DI 08806088582412, distributed nationwide in the United States and internationally including Australia, Belgium, Canada, Chile, China Mainland and the Czech Republic. The FDA has classified the action as Class II.
The tube head is the heavy component at the end of the arm, so bolt failure creates a risk of it moving or falling. The enforcement record reports no injuries. Sites operating an affected system should follow the firm's instructions.
The recalled product
- Product
- GM85 Digital Mobile X-ray imaging System; Model Number: GM85;
- Manufacturer
- NeuroLogica Corporation
- Category
- Medical Device — X-ray system
- Affected units
- 236
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (20)
- Model Number: GM85
- UDI-DI: 08806088582412
- Serial Numbers: 514GM3IK400001L
- 5143M3IK300001X
- 5143M3IJ100010M
- 5143M3IHC00014H
- 5143M3IK300008R
- 5143M3JJ400007E
- 5143M3HJ900002V
- 5143M3IK400004F
- 513YM3IK500003Z
- 5143M3HK600005V
- 5143M3JJ200010J
- 5143M3HK600008M
- 5143M3HK200004H
- 5143M3JK500003W
- 5143M3IJ200002F
- 5143M3IHC00007A
- 5143M3JJ300001K
- 5143M3IJ600014H
Distribution
Distributed nationwide across the United States.
Related recalls
Same manufacturer · NeuroLogica Corporation
See all →- HighNeuroLogica GM60A Mobile X-Ray System Anti-Fall Arm Safety Failure Risk
FDA (Devices) · 2024-08-28
- HighSamsung GM85 Mobile X-ray System Welding Frame Defect Arm Fall Risk
FDA (Devices) · 2024-02-21
- HighOmniTom CT imaging device recalled for wheel loosening and battery contamination
FDA (Devices) · 2023-07-12
- HighSamsung GM85 Digital X-ray Imaging System Arm Latch Failure
FDA (Devices) · 2023-04-26
- ModerateNeuroLogica OmniTom CT Systems Recalled for Operational Shutdown Issues
FDA (Devices) · 2020-08-19
Same hazard · metal fatigue
See all →- High
Similar recalls
- HighNeuroLogica Recalls Samsung GM85 Mobile Digital X-ray Imaging Systems
FDA (Devices) · 2019-06-05
- HighSamsung GM85 Mobile X-ray System Welding Frame Defect Arm Fall Risk
FDA (Devices) · 2024-02-21
- HighSamsung GM85 Digital X-ray Imaging System Arm Latch Failure
FDA (Devices) · 2023-04-26
- HighNeuroLogica GM60A Mobile X-Ray System Anti-Fall Arm Safety Failure Risk
FDA (Devices) · 2024-08-28
- HighMobile Radiographic System Arm May Drop Due to Missing Safety Bolts
FDA (Devices) · 2025-01-15
- HighMedical radiographic system arm drop hazard due to missing safety bolts
FDA (Devices) · 2025-01-15