Change Healthcare Cardiology Hemodynamic Software Calculation Error Recall
Change Healthcare Cardiology Hemo software versions 14.1.1–15.1 may produce inaccurate hemodynamic calculations due to unit-of-measurement configuration errors, potentially leading to misdiagnosis and inappropriate patient treatment.
What this means for you
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: FDA Class II medical device recall involving software that calculates critical clinical parameters. Configuration and data-entry errors can lead to inaccurate hemodynamic calculations, potentially resulting in misdiagnosis and inappropriate treatment. This risk-of-harm scenario warrants a High severity rating under the FDA Class II classification.
Plain-English summary
Change Healthcare Cardiology Hemo is a cardiology hemodynamic analysis system used for patient assessment. The recall affects versions 14.1.1, 14.2, 14.3, 14.3.2, 15.0.1, and 15.1. Approximately 221 systems have been distributed worldwide, including in the United States, Australia, Canada, Germany, Ireland, Israel, and the United Kingdom.
The issue occurs when the unit of measurement (UOM) for hemoglobin is configured in a way that differs from the user's understanding. This configuration mismatch can lead to incorrect data entry and subsequent inaccurate hemodynamic calculations.
Inaccurate hemodynamic calculations may result in misdiagnosis and inappropriate treatment.
The recalled product
- Product
- Change Healthcare Cardiology Hemo. Versions 14.2, 14.1.1, 14.2, 14.3, 14.3.2, 14.3, 15.0.1, 15.1.
- Manufacturer
- CHANGE HEALTHCARE CANADA COMPANY
- Hazard
- data-entry-error
- calculation-error
- configuration-error
- misdiagnosis-risk
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (14)
- UDI Numbers: (01)17540262100013(10)140100
- (01)17540262100020(10)140101
- (01)17540262100037(10)140200
- (01)17540262100044(10)140300
- (01)17540262100051(10)140302
- (01)17540262100099(10)150000
- (01)17540262100105(10)150001
- (01)17540262100129(10)150100. Versions 14.2
- 14.1.1
- 14.2
- 14.3
- 14.3.2
- 15.0.1
- 15.1.
Distribution
Distributed nationwide across the United States.
Related recalls
Same category
- SevereT-Lok Bone Marrow Tray Kits Recalled for Sterility Testing Failure
FDA (Devices) · 2026-07-22
- SevereAbiomed Impella CP Introducer Kit heart pump accessory recall
FDA (Devices) · 2026-07-22
- HighOlympus Glass Body for 194 Ellik Evacuator Model 190A Sterilization Recall
FDA (Devices) · 2026-07-22
- HighOlympus Toomey Evacuator 50cc with Adaptor sterilization recall
FDA (Devices) · 2026-07-22
- SevereMedtronic Bio-Medicus Adult Cannula Kit Recalled for Foreign Material
FDA (Devices) · 2026-07-22