Change Healthcare Cardiology Hemodynamic Software Calculation Error Recall
Change Healthcare Cardiology Hemo software versions 14.1.1–15.1 may produce inaccurate hemodynamic calculations due to unit-of-measurement configuration errors, potentially leading to misdiagnosis and inappropriate patient treatment.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: FDA Class II medical device recall involving software that calculates critical clinical parameters. Configuration and data-entry errors can lead to inaccurate hemodynamic calculations, potentially resulting in misdiagnosis and inappropriate treatment. This risk-of-harm scenario warrants a High severity rating under the FDA Class II classification.
Plain-English summary
Change Healthcare Cardiology Hemo is a cardiology hemodynamic analysis system used for patient assessment. The recall affects versions 14.1.1, 14.2, 14.3, 14.3.2, 15.0.1, and 15.1. Approximately 221 systems have been distributed worldwide, including in the United States, Australia, Canada, Germany, Ireland, Israel, and the United Kingdom.
The issue occurs when the unit of measurement (UOM) for hemoglobin is configured in a way that differs from the user's understanding. This configuration mismatch can lead to incorrect data entry and subsequent inaccurate hemodynamic calculations.
Inaccurate hemodynamic calculations may result in misdiagnosis and inappropriate treatment.
The recalled product
- Product
- Change Healthcare Cardiology Hemo. Versions 14.2, 14.1.1, 14.2, 14.3, 14.3.2, 14.3, 15.0.1, 15.1.
- Manufacturer
- CHANGE HEALTHCARE CANADA COMPANY
- Affected units
- 221
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (14)
- UDI Numbers: (01)17540262100013(10)140100
- (01)17540262100020(10)140101
- (01)17540262100037(10)140200
- (01)17540262100044(10)140300
- (01)17540262100051(10)140302
- (01)17540262100099(10)150000
- (01)17540262100105(10)150001
- (01)17540262100129(10)150100. Versions 14.2
- 14.1.1
- 14.2
- 14.3
- 14.3.2
- 15.0.1
- 15.1.
Distribution
Distributed nationwide across the United States.
Part of a larger recall action
This product is one of 3 recalled by CHANGE HEALTHCARE CANADA COMPANY under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 3 products in this recall →Related recalls
Same manufacturer · CHANGE HEALTHCARE CANADA COMPANY
See all →- HighRadiology software recalled because reports may not display fully when viewed
FDA (Devices) · 2025-10-08
- ModerateChange Healthcare Cardiology Hemo Software Unexpected Shutdown Risk
FDA (Devices) · 2025-07-09
- HighCardiology hemodynamics software update may cause unexpected system shutdowns
FDA (Devices) · 2025-07-09
- HighChange Healthcare Cardiology Hemo Software Recalled for Calculation Errors
FDA (Devices) · 2024-10-23
- HighHorizon Cardiology Hemodynamic System Recalled for Configuration Error
FDA (Devices) · 2024-10-23
Same hazard · data entry error
See all →- HighOmnipod 5 app recalled because magnifying a field duplicates typed digits
FDA (Devices) · 2025-10-22
Similar recalls
- HighChange Healthcare Cardiology Hemo software autosave defect recall
FDA (Devices) · 2025-01-22
- ModerateChange Healthcare Cardiology Hemo Software Unexpected Shutdown Risk
FDA (Devices) · 2025-07-09
- HighChange Healthcare Cardiology Software Defect May Cause Incorrect Medication Administration
FDA (Devices) · 2025-01-22
- HighCardiology hemodynamics software update may cause unexpected system shutdowns
FDA (Devices) · 2025-07-09
- HighMcKesson Cardiology Hemo Software Recall for Hemoglobin Calculation Error
FDA (Devices) · 2014-10-15
- HighChange Healthcare Recalls Horizon Cardiology Hemo Software Due to Medication Discrepancies
FDA (Devices) · 2019-06-05