CHANGE HEALTHCARE CANADA COMPANY recalled 3 products on 2024-10-23

The FDA (Devices) publishes a recall like this one as 3 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 3 pages.

Products
3
Announced
2024-10-23
Critical
0
Units
259

Why these products were recalled

A potential issue has been identified in Change Healthcare Cardiology Hemo where incorrect data entry can lead to inaccurate hemodynamic calculations, potentially resulting in misdiagnosis and inappropriate treatment. This issue occurs when the unit of measurement (UOM) for hemoglobin is configured in a way that differs from the users understanding.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–3 of 3

  • HighFDA (Devices)··2024-10-23

    Horizon Cardiology Hemodynamic System Recalled for Configuration Error

    Horizon Cardiology Hemo Version 12.2 is recalled for a unit-of-measurement configuration issue that can produce incorrect hemodynamic calculations, potentially resulting in misdiagnosis and inappropriate treatment. The recall affects 3 worldwide systems.

    Product
    Horizon Cardiology Hemo. Version 12.2.
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-10-23

    Change Healthcare Cardiology Hemo Software Recalled for Calculation Errors

    Change Healthcare is recalling Cardiology Hemo software versions 13.0 through 14.0 because incorrect unit configurations can lead to inaccurate hemodynamic calculations and potential misdiagnosis.

    Product
    McKesson Cardiology Hemo. Versions 13.0, 13.1.1, 13.1.2, 13.2, 13.2.1, 14.0.
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-10-23

    Change Healthcare Cardiology Hemodynamic Software Calculation Error Recall

    Change Healthcare Cardiology Hemo software versions 14.1.1–15.1 may produce inaccurate hemodynamic calculations due to unit-of-measurement configuration errors, potentially leading to misdiagnosis and inappropriate patient treatment.

    Product
    Change Healthcare Cardiology Hemo. Versions 14.2, 14.1.1, 14.2, 14.3, 14.3.2, 14.3, 15.0.1, 15.1.