The Recall Desk

Hazard

Misdiagnosis Risk recalls

51 recalls in our database tag this hazard, drawn from FDA, USDA FSIS, CPSC, and NHTSA source notices.

What the numbers show

Total
51
Critical
0
Severe
2
Most recent
2026-08-12

Of the 51 misdiagnosis risk recalls we have scored against our rubric, 0 (0%) are Critical — deaths or serious injuries reported, or the agency’s own highest classification — and 2 (4%) are Severe.

About this hazard tag

Hazard tags are short, machine-readable labels we attach to every recall to make cross-cutting queries possible — “all misdiagnosis risk recalls regardless of agency”, “all listeria recalls in California”, “all airbag recalls in 2025”. The tags are extracted from each recall’s source text during the LLM rewrite step and validated against a fixed schema.

A recall can carry multiple tags — a Listeria recall might also be tagged ready-to-eat and undeclared-allergen. Use the full-text search if you want to combine hazard tags with brand or category filters.

1–25 of 51

  • HighFDA (Devices)·Z-2874-2026·2026-08-12

    Merge PACS Medical Imaging Workstation Recalled Due to Patient Data Display Error

    Merge Healthcare is recalling 53 Merge PACS medical imaging workstations due to a software issue where thumbnails from one patient's study may incorrectly display when viewing a different patient's study, potentially compromising patient confidentiality.

    Product
    Merge PACS, medical imaging diagnostic workstation
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2670-2026·2026-07-08

    Radius VSM ECG Pre-connected Set Recalled for Incorrect Heart Rate Detection

    Masimo Corporation is recalling Radius VSM ECG pre-connected sets because the ECG electrode assembly may incorrectly detect extreme bradycardia or tachycardia and trigger false alarms on patients with normal heart rates, potentially distracting clinicians from true positive alarms and delaying treatment.

    Product
    Radius VSM ECG pre-connected Set, REF: 4842, 4695
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2553-2026·2026-07-01

    GE Healthcare AW Server Radiological Image Processing System Software Defect

    GE Healthcare is recalling the AW Server 3.2 ext. 6.5 radiological image processing system due to a software synchronization issue that may cause the wrong patient's images to be displayed without warning, potentially leading to misdiagnosis.

    Product
    GE Healthcare AW Server 3.2 ext.6.5; Radiological Image Processing System
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2005-2026·2026-05-06

    epoc BGEM BUN Test Card for epoc Blood Analysis System

    Siemens Healthcare Diagnostics is recalling epoc BGEM BUN Test Cards due to risk of inaccurate pH and carbon dioxide readings that could lead to inappropriate or missed treatment of acid-base disorders.

    Product
    epoc BGEM BUN Test Card (25 pk) for the epoc Blood Analysis System; Siemens Material Number (SMN): 10736515;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1009-2026·2026-01-14

    Thyroid receptor immunoassay recalled over variable results across instrument platforms

    Roche Diagnostics is recalling 1,300 Elecsys Anti-TSHR cassettes because results vary by instrument platform and calibrator lots vary, which may lead to incorrect diagnosis or treatment.

    Product
    The Elecsys Anti-TSHR immunoassay is a three-step competitive immunoassay with streptavidin-coated microparticles and electrochemiluminescence detection. Patient serum sample is incubated with pretreatment reagent PT1 and the reconstituted PT2 containing a pre-formed…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0941-2026·2025-12-24

    Imaging viewer software recalled over inaccurate length measurement values

    DICOM Grid is recalling 924 units of Intelerad InteleShare software because a bug can produce inaccurate length measurement tool values that could lead to misdiagnosis.

    Product
    Intelerad InteleShare software, with ProViewer component
    Category
    Medical Device
    Distribution
    45 states
  • SevereFDA (Devices)·Z-2023-2025·2025-07-02

    BD BACTEC MGIT 960 PZA Kit tuberculosis susceptibility testing recall

    Becton Dickinson is recalling BD BACTEC MGIT 960 PZA Kit due to reports of intermittent false resistance results during pyrazinamide susceptibility testing of Mycobacterium tuberculosis. The issue affects units distributed worldwide.

    Product
    BD BACTEC MGIT 960 PZA Kit. IVD qualitative procedure for susceptibility testing of Mycobacterium tuberculosis, from culture to pyrazinamide (PZA).
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1925-2025·2025-06-18

    Digital Acquisition Box Electrode Placement Labels Applied Backwards

    Natus Neurology has recalled the OBM00002 OBM DAB (Digital Acquisition Box) because electrode placement labels were applied backwards. This may lead to misdiagnosis of seizure location in neonates monitored with the device.

    Product
    Brand Name: OBM00002 OBM DAB (Digital Acquisition Box) Product Name: OBM00002 OBM DAB (Digital Acquisition Box) Model/Catalog Number: OBM00002 OBM DAB (Digital Acquisition Box) Software Version: N/A Product Description: OBM00002 OBM DAB (Digital Acquisition Box) Component…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1812-2025·2025-05-28

    PhenoMATRIX bacterial culture plate interpretation software recalled nationwide

    Copan WASP is recalling PhenoMATRIX, an AI-powered bacterial culture plate interpretation software, because it lacks required FDA premarket approval, creating potential risk of inaccurate colony counts and morphology data that could affect patient diagnosis.

    Product
    PhenoMATRIX
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1785-2025·2025-05-28

    True 3D Viewer Model EP-00003 Software Orientation Display Issue Recall

    Echopixel's True 3D Viewer (Model EP-00003) with certain software versions can display ultrasound image orientation incorrectly, potentially leading to misdiagnosis or incorrect surgical planning.

    Product
    True 3D Viewer, Model: EP-00003
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0735-2025·2025-01-01

    Blood analyzer system recalled for patient name assignment software error

    Siemens recalled its epoc NXS Host blood analyzer due to a software malfunction that can assign incorrect patient names to test results, potentially leading to misdiagnosis or mistreatment.

    Product
    epoc NXS Host; Siemens Material Number (SMN): 11413497 (EU); Software Version Number: 4.14.9. Intended for use by trained medical professionals as an in vitro diagnostic device for the quantitative testing of samples of heparinized or un-anticoagulated arterial, venous or…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0740-2025·2025-01-01

    Siemens Blood Analysis Device Recalled Over Patient Name Mix-up Software Defect

    Siemens is recalling the epoc NXS blood analysis system due to a software flaw that can assign incorrect patient names to test results, risking misdiagnosis or mistreatment if the operator doesn't verify the patient name.

    Product
    epoc NXS Host; Siemens Material Number (SMN): 11413583(CN, KR) ; Software Version Number: 4.14.9. Intended for use by trained medical professionals as an in vitro diagnostic device for the quantitative testing of samples of heparinized or un-anticoagulated arterial, venous or…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0737-2025·2025-01-01

    Blood Analysis System Software Malfunction May Assign Incorrect Patient Names

    Siemens epoc NXS Host blood analysis systems contain a software malfunction that may assign incorrect patient names to blood measurements. This could result in clinicians making treatment decisions based on the wrong patient's test results.

    Product
    epoc NXS Host; Siemens Material Number (SMN): 11413506 (CA); Software Version Number: 4.14.9. Intended for use by trained medical professionals as an in vitro diagnostic device for the quantitative testing of samples of heparinized or un-anticoagulated arterial, venous or…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0741-2025·2025-01-01

    Siemens epoc blood analyzer software assigns incorrect patient names

    Siemens epoc NXS blood analysis system has a software defect that can incorrectly assign patient names to blood test results, potentially causing misdiagnosis or mistreatment.

    Product
    epoc NXS Host; Siemens Material Number (SMN): 11413879(IN) ; Software Version Number: 4.14.9. Intended for use by trained medical professionals as an in vitro diagnostic device for the quantitative testing of samples of heparinized or un-anticoagulated arterial, venous or…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0736-2025·2025-01-01

    Siemens epoc Blood Analysis System Recalled for Patient Identification Software Error

    Siemens is recalling epoc NXS Host blood analysis systems due to a software malfunction that can assign incorrect patient names to test results, potentially leading to misdiagnosis and mistreatment if not verified by the user.

    Product
    epoc NXS Host; Siemens Material Number (SMN): 11413498(JP) ; Software Version Number: 4.14.9. Intended for use by trained medical professionals as an in vitro diagnostic device for the quantitative testing of samples of heparinized or un-anticoagulated arterial, venous or…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0583-2025·2024-12-11

    Spectral CT Devices Pose Collision and Misdiagnosis Software Risks

    Philips Spectral CT and Spectral CT Plus devices may fail to update plan positioning during interventional procedures, creating collision risks to operators and patients. A software defect may also assign incorrect patient IDs, potentially causing misdiagnosis.

    Product
    Spectral CT and Spectral CT Plus. Model Numbers: 728333, 728340, 728344.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0126-2025·2024-10-23

    Change Healthcare Cardiology Hemodynamic Software Calculation Error Recall

    Change Healthcare Cardiology Hemo software versions 14.1.1–15.1 may produce inaccurate hemodynamic calculations due to unit-of-measurement configuration errors, potentially leading to misdiagnosis and inappropriate patient treatment.

    Product
    Change Healthcare Cardiology Hemo. Versions 14.2, 14.1.1, 14.2, 14.3, 14.3.2, 14.3, 15.0.1, 15.1.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0055-2025·2024-10-16

    Enterprise Imaging XERO Viewer medical software reference line placement error

    Enterprise Imaging XERO Viewer incorrectly displays reference lines on medical images due to a software defect affecting images with non-square pixel spacing. The misplaced reference lines may lead to diagnostic errors.

    Product
    Enterprise Imaging XERO Viewer- software application used for reference and diagnostic viewing of multispecialty medical imaging and non-imaging data . Enterprise Imaging XERO Viewer 8.0.0, 8.0.1, 8.1, 8.1.1, 8.1.2, 8.1.3,8.1.4, 8.2.0.000, 8.2.1.000, 8.2.2.000, 8.3.0.000…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3217-2024·2024-10-02

    Siemens Diagnostic Ultrasound Systems Recalled for Incorrect Unit Conversion

    Siemens ACUSON Maple ultrasound systems may perform incorrect unit conversion in software, producing values 10 times smaller than actual. This could lead to misdiagnosis or inappropriate patient management.

    Product
    ACUSON Maple 1.0 Diagnostic Ultrasound System, REF: 11711750, with software. ACUSON Maple Select 1.0 Diagnostic Ultrasound System, REF: 11711779, with software
    Category
    Medical Device
    Distribution
    42 states
  • HighFDA (Devices)·Z-3216-2024·2024-10-02

    ACUSON Redwood Ultrasound Systems Recall for Software Unit Conversion Error

    Siemens ACUSON Redwood ultrasound systems may display measurements 10 times smaller than actual if operators change the software from factory defaults to milliliters per second or minute. This could lead to misdiagnosis and poor patient management.

    Product
    ACUSON Redwood 1.0 and 2.0 Diagnostic Ultrasound Systems, REF: 11503314, with software
    Category
    Medical Device
    Distribution
    42 states
  • HighFDA (Devices)·Z-3215-2024·2024-10-02

    Siemens ACUSON Juniper ultrasound systems software unit conversion error

    Siemens is recalling ACUSON Juniper ultrasound systems that may perform incorrect unit conversions when configured with non-default settings, potentially producing measurements 10 times smaller than actual values and risking patient misdiagnosis.

    Product
    ACUSON Juniper 1.0, 1.5, 2.0, 2.5 Diagnostic Ultrasound Systems, REF: 11335791, with software. ACUSON Juniper Select 2.0, 2.5 Diagnostic Ultrasound Systems, REF: 11653093, with software.
    Category
    Medical Device
    Distribution
    42 states
  • HighFDA (Devices)·Z-2034-2024·2024-06-19

    Streptococcal A Rapid Test Kits Recalled for Unauthorized At-Home Distribution

    Wondfo USA has recalled Streptococcal A rapid test kits that are being sold for unauthorized at-home and direct-to-consumer use. The kits should not be used outside professional healthcare settings.

    Product
    Streptococcal A rapid test kits, labeled as: Wondfo Biotech Co., Ltd, WondView Strep A Test Kit, Guangzhou Wondfo Biotech Co., Ltd, Wondfo One Step Strep A Swab Test, Dealmed Medical Supplies LLC, Strep A Rapid Test Kit, Jant Pharmacal Corporation, Accustrip VALUPAK Strep A…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1756-2024·2024-05-08

    Philips Spectral CT Oncology Imaging System Recall: Laser Reflection Hazard

    Philips recalls 12 units of its Spectral CT Oncology Essentials system because a marking laser could reflect off the gantry cover, potentially causing misdiagnosis and incorrect patient treatment.

    Product
    The Oncology Essentials Package of the Spectral CT imaging system.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1600-2024·2024-05-01

    ACUSON Redwood 1.0 ultrasound displays incorrect cardiac measurement values

    The ACUSON Redwood 1.0 Diagnostic Ultrasound System may display the last measured cardiac value instead of the minimum or maximum value as configured, potentially leading to misdiagnosis or improper patient management.

    Product
    ACUSON Redwood 1.0 Diagnostic Ultrasound System, REF: 11503314, running software version VA10A to VA10G
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1171-2024·2024-02-28

    Cardiology imaging software may calculate heart chamber size incorrectly

    Synapse Cardiology PACS software versions 7.0–7.3.0 may calculate left ventricle mass incorrectly, potentially leading to misdiagnosis or wrong treatment plans. Thirteen units have been distributed in the US and Israel.

    Product
    Synapse Cardiology PACS V7.3.0, V7.2, V7.1, V7.0. A web-based application as the primary user interface for the processing of medical images.
    Category
    Medical Device
    Distribution
    Distributed nationwide