LFS flat screen arm system surgical lamp bracket stopper block defect
The LFS flat screen arm system with surgical lamp has a broken or missing stopper block that could allow the screen to rotate excessively, potentially causing the spring shaft to break and shear off.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: This is a risk-of-harm medical device (surgical lamp system) where the potential hazard is spring shaft failure, but no injuries have been reported. FDA Class II classification without reported incidents supports Score 3 under the rubric.
Plain-English summary
The LFS flat screen arm system with surgical lamp, including multiple model numbers across the Aurora series (Aurora, Aurora 2, Aurora 3, Aurora 4, Aurora Astro) and Stellar series (Stellar XL, Stellar), is being recalled due to a defect in the bracket stopper block.
The stopper block on the LFS bracket in affected units is broken or missing. Without this stopper block, the flat screen can rotate beyond its original angle limit. This excessive rotation applies dangerous force to the spring shaft, which could break and shear off.
256 units with specific serial numbers have been identified as affected. These units were distributed in Michigan. Facilities using these systems should check their equipment against the provided serial number list to determine if their units are involved.
If you have an affected unit, contact the manufacturer DKK Dai-Ichi Shomei Co., Ltd. for repair or replacement. Do not use affected units until the stopper block defect has been corrected.
The recalled product
- Product
- LFS Flat screen arm system with surgical lamp for Models Aurora series, Aurora 2 series, Aurora 3 series, Aurora 4 series, Aurora Astro series, Stellar XL, and Stellar series, LFS Flat Screen Arm System Model Numbers LFSST2323, LFSLFSST29TV23, LFSST29TV23, LFSST23TV, LFSST23…
- Manufacturer
- DKK Dai-Ichi Shomei Co., Ltd.
- Affected units
- 256
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (20)
- Serial Numbers: 06087323 - 06087352
- 06087353 - 06087354
- 06081292 - 06081293
- 06087448 - 06087449
- 06087450 - 06087469
- 06097595 - 06097596
- 06107716 - 06107718
- 06107719 - 06107728
- 06107730
- 06107731 - 06107732
- 06111348 - 06111349
- 06117855 - 06117856
- 06117857 - 06117866
- 06117868 - 06117868
- 06111350 - 06111351
- 06117949 - 06117949
- 06117954 - 06117973
- 06118031 - 06118032
- 06128116 - 06128131
- 06128172 - 06128176
Distribution
Distributed in 1 state:
Part of a larger recall action
This product is one of 2 recalled by DKK Dai-Ichi Shomei Co., Ltd. under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 2 products in this recall →Related recalls
Same manufacturer · DKK Dai-Ichi Shomei Co., Ltd.
See all →- HighAurora 7 Surgical Light Heads Recalled for Cracked Weld Frames
FDA (Devices) · 2024-08-21
- HighSurgical light heads may detach due to weld defects
FDA (Devices) · 2024-08-21
- HighSurgical Lights Recalled for Possible Weld Defect and Frame Detachment
FDA (Devices) · 2024-08-21
- HighSurgical Lighting Equipment Recalled for Cracked Welds and Frame Detachment Risk
FDA (Devices) · 2024-08-21
- HighMedical Device Recall: LFS Flat Screen Support Arm Bracket Failure
FDA (Devices) · 2023-10-11
Same hazard · structural defect
See all →- SevereThomas Built School Buses Recalled for Weak Body Joints
NHTSA · 2026-10-22
- SevereGM Recalls 2026 Chevrolet Traverse and GMC Acadia for Seat Weld Defect
NHTSA · 2026-10-21
- SevereToyota Recalls Sienna Hybrid Vehicles Due to Second-Row Seat Back Defect
NHTSA · 2026-10-20
- SevereSubaru Recalls 2026 Forester and Crosstrek for Ball Joint Defect
NHTSA · 2026-10-19
- SevereVolvo Recalls 2025-2026 VNL Trucks for Loose Structural Bolts
NHTSA · 2026-10-15
Similar recalls
- HighSurgical Lighting Equipment Recalled for Cracked Welds and Frame Detachment Risk
FDA (Devices) · 2024-08-21
- HighSurgical light heads may detach due to weld defects
FDA (Devices) · 2024-08-21
- HighSteris Surgical Lighting Monitor Spring Arms Recalled for Welding Defect
FDA (Devices) · 2014-07-02
- HighSurgical Lights Recalled for Possible Weld Defect and Frame Detachment
FDA (Devices) · 2024-08-21
- HighAurora 7 Surgical Light Heads Recalled for Cracked Weld Frames
FDA (Devices) · 2024-08-21
- HighLuminos dRF Max X-ray Display Support Arm May Unintentionally Lower
FDA (Devices) · 2024-08-14