The Recall Desk

Manufacturer

DKK Dai-Ichi Shomei Co., Ltd.

6 recalls in our database name DKK Dai-Ichi Shomei Co., Ltd. as the manufacturing or recalling firm.

What the numbers show

Total
6
Critical
0
Severe
0
Most recent
2024-08-21

Of the 6 recalls from DKK Dai-Ichi Shomei Co., Ltd. we have scored against our rubric, 0 (0%) are Critical — deaths or serious injuries reported, or the agency’s own highest classification — and 0 (0%) are Severe.

About this listing

The manufacturer string comes from the agency’s own source notice — for FDA recalls, the “recalling firm” field; for NHTSA recalls, the vehicle manufacturer; for CPSC recalls, the firm-of-record on the joint press release. A single corporate family can appear under multiple manufacturer strings as agency notices aren’t standardized on a canonical name. If you don’t see recalls you expect, try a parent-company search or a broader full-text search.

Severity scores below are computed from the agency’s classification and the recall text, never from manufacturer behaviour or history. Many manufacturers issue routine corrective actions (software updates, labelling corrections) that are correctly scored low — having recalls on this list isn’t inherently a quality signal.

1–6 of 6

  • HighFDA (Devices)·Z-2618-2024·2024-08-21

    Aurora 7 Surgical Light Heads Recalled for Cracked Weld Frames

    DKK Dai-Ichi Shomei is recalling 555 Aurora 7 surgical light heads because frame welds can crack, causing the frame to sag, the housing to crack, or the light head to detach.

    Product
    Aurora Series, Aurora 7 lighthead; Models: LED7 (including TV models). The Aurora Series is intended to illuminate the surgical field and ensure visibility of the surgical site.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2620-2024·2024-08-21

    Surgical light heads may detach due to weld defects

    DKK Dai-Ichi Shomei is recalling 293 Aurora surgical light heads because the frame welds can crack, potentially causing the light head to detach during use. No injuries have been reported.

    Product
    Aurora 3 Series, Aurora 7 lighthead; Models: AUT7 (including TV models). The Aurora Series is intended to illuminate the surgical field and ensure visibility of the surgical site.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2619-2024·2024-08-21

    Surgical Lights Recalled for Possible Weld Defect and Frame Detachment

    DKK Dai-Ichi Shomei Co., Ltd. is recalling 822 Aurora surgical light heads due to a weld defect that could cause the light head to sag, crack, or detach during use.

    Product
    Aurora 2 Series, Aurora 7 lighthead; Models: AUR7 (including TV models). The Aurora Series is intended to illuminate the surgical field and ensure visibility of the surgical site.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2621-2024·2024-08-21

    Surgical Lighting Equipment Recalled for Cracked Welds and Frame Detachment Risk

    Aurora Series surgical light heads are recalled due to cracked welds on the frame. The defects can cause the frame to sag, housing to crack, or the light head to detach.

    Product
    Aurora 4 Series, Aurora 7 lighthead; Models: AUA7 (including TV models). The Aurora Series is intended to illuminate the surgical field and ensure visibility of the surgical site.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0072-2024·2023-10-11

    Medical Device Recall: LFS Flat Screen Support Arm Bracket Failure

    The FDA is recalling LFS flat screen support arm systems because the stopper block could be broken or missing. Without this stopper, the screen can rotate excessively and damage the bracket's spring shaft.

    Product
    LFS Flat screen support arm systems without surgical lamp for models Aurora series, Aurora 2 series, Aurora 3 series, Aurora 4 series, Aurora Astro series, Stellar XL, and Stellar series, LFS Flat Screen Support Arm Model Numbers
    Category
    Medical Device
    Distribution
    1 state
  • HighFDA (Devices)·Z-0071-2024·2023-10-11

    LFS flat screen arm system surgical lamp bracket stopper block defect

    The LFS flat screen arm system with surgical lamp has a broken or missing stopper block that could allow the screen to rotate excessively, potentially causing the spring shaft to break and shear off.

    Product
    LFS Flat screen arm system with surgical lamp for Models Aurora series, Aurora 2 series, Aurora 3 series, Aurora 4 series, Aurora Astro series, Stellar XL, and Stellar series, LFS Flat Screen Arm System Model Numbers LFSST2323, LFSLFSST29TV23, LFSST29TV23, LFSST23TV, LFSST23, LFS
    Category
    Medical Device
    Distribution
    1 state