Why these products were recalled

The stopper block on the LFS bracket could be broken or missing. Without this stopper block, the flat screen attached to the LFS flat screen system can rotate beyond the original angle limit and it could cause excessive force to be given to the spring shaft of the LFS bracket. Due to the excessive force given to the spring shaft, the spring shaft could break and shear off.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–2 of 2

  • HighFDA (Devices)··2023-10-11

    Medical Device Recall: LFS Flat Screen Support Arm Bracket Failure

    The FDA is recalling LFS flat screen support arm systems because the stopper block could be broken or missing. Without this stopper, the screen can rotate excessively and damage the bracket's spring shaft.

    Product
    LFS Flat screen support arm systems without surgical lamp for models Aurora series, Aurora 2 series, Aurora 3 series, Aurora 4 series, Aurora Astro series, Stellar XL, and Stellar series, LFS Flat Screen Support Arm Model Numbers
    Category
    Distribution
    1 state
  • HighFDA (Devices)··2023-10-11

    LFS flat screen arm system surgical lamp bracket stopper block defect

    The LFS flat screen arm system with surgical lamp has a broken or missing stopper block that could allow the screen to rotate excessively, potentially causing the spring shaft to break and shear off.

    Product
    LFS Flat screen arm system with surgical lamp for Models Aurora series, Aurora 2 series, Aurora 3 series, Aurora 4 series, Aurora Astro series, Stellar XL, and Stellar series, LFS Flat Screen Arm System Model Numbers LFSST2323, LFSLFSST29TV23, LFSST29TV23, LFSST23TV, LFSST23…
    Category
    Distribution
    1 state